|
|
||||||||
Basic Science Investigations |
1 Department of Radiology, Washington University School of Medicine, St. Louis, Missouri
2 Concorde Microsystems, Inc., Knoxville, Tennessee
3 Department of Molecular and Medical Pharmacology, Crump Institute for Molecular Imaging, UCLA, Los Angeles, California
| ABSTRACT |
|---|
|
|
|---|
Key Words: microPET animal PET scanner instrumentation performance evaluation
| INTRODUCTION |
|---|
|
|
|---|
Advancements in molecular biology have made genetically modified mice the animal of choice to study models of human disease. By inserting or knocking out specific genes, these transgenic mice exhibit phenotypes that are susceptible or immune to human disease. Individual risk factors can be studied to establish disease correlation as well as to develop new treatments. To better study laboratory mice using PET, researchers continue to develop new techniques to improve the performance of animal scanners, with particular emphasis on the image resolution and system sensitivity (415).
The microPET Focus is the latest generation of commercial microPET scanner from Concorde Microsystems, Inc. This system incorporates several changes in design that significantly improve its performance compared with previous models (16,17). This work presents an investigation into the improvement in technology and performance of the Focus system compared with the previous microPET systems.
| MATERIALS AND METHODS |
|---|
|
|
|---|
The fiber optic bundle consists of 8 x 8 elements of square, multiclad plastic fibers each measuring 2.2 x 2.2 x 100.0 mm3. The same reflective material is also placed between individual fibers to provide optical isolation and to improve light collection efficiency. The use of optical fibers of cross-section greater than that of the LSO elements improves the light collection efficiency. These design changes were critical for good energy resolution as the LSO crystals have a relatively small cross-section that limits the light collection efficiency (18,19).
The position-sensitive photomultiplier tube continues to be the Hamamatsu R5900-C12. A simple resistor network was used to convert the 12 anode outputs to 4 position-encoded signals (20). These 4 signals are fed into the pulse processing circuits and subsequently sent to the coincidence processor for coincidence determination. To improve the linearity of the analog-to-digital converters, each event is digitized twice and averaged for event positioning and energy determination. The new electronics doubles the data transfer speed and effectively doubles the counting capability of the system under high counting rate situations.
The system acquires data in list mode to permit maximum flexibility in the postprocessing and reconstruction. Coincidence events can be sorted into 3-dimensional (3D) sinograms with different combinations of span and ring differences (21) or directly into 2-dimensional (2D) sinograms by single-slice rebinning (SSRB) (22). Images can be reconstructed using filtered backprojection (FBP) or ordered-subsets expectation maximization (23) (OSEM) algorithms.
The system is equipped with a point source rotating mechanism that is used to acquire a normalization scan for system calibration and transmission scans for attenuation correction. The horizontal and vertical motion of the animal bed is computer controlled, and there is a laser-positioning marker. The animal port has an opening of 22 cm to accommodate laboratory primates. The system occupies a small footprint (130 x 140 cm2) and comes on casters so that it can be easily moved between laboratories. Table 1 summarizes the system parameters of the Focus scanner as well as the previous microPET systems for comparison.
|
20,000 per crystal element) that were sorted by pulse height and crystal identification to provide energy spectra of individual crystals. The energy resolution, defined as the full width at half maximum (FWHM) of the photopeak divided by the energy corresponding to the center of photopeak, was determined for each crystal in the system and the average energy resolution was calculated.
Spatial Resolution
All spatial resolution measurements were acquired with an energy window of 250750 keV and a timing window of 6 ns using a 22Na point source with a nominal size of 0.5 mm, embedded in a Lucite disk (North American Scientific). 22Na has emission energy (Eaveß+ = 250 keV) and positron range (
0.23 mm) similar to 18F, the most widely used positron-emitting isotope. Results of the spatial resolution measurements were not corrected for source dimension, positron range, or acolinearity of positron annihilation.
The point source was attached to the animal bed, positioned at the CFOV, and moved radially toward the edge of the gantry opening in steps. List mode data at each location were sorted into 3D sinograms with the delayed events subtracted from the prompt events to correct for the random coincidences, followed by normalization to compensate for the differences in detection efficiency. The corrected 3D sinograms were Fourier rebinned (FORE) (24) to form 2D sinograms and subsequently reconstructed using 2D FBP with a ramp filter cutoff at the Nyquist frequency and a 10x image zoom. Tangential and radial resolutions were estimated from the horizontal and vertical profiles through the pixel with maximum intensity. Axial resolution was estimated from oversampled axial profiles by interleaving 4 measurements with the source stepped along the axial direction with a step size of 0.2 mm. The FWHM and full width at tenth maximum (FWTM) were measured from the extracted profiles by linear interpolation between 2 nearest pixels bracketing the half-values and one-tenth values, respectively.
Sensitivity
The absolute sensitivity of the camera was measured using a 68Ge point source encapsulated in a stainless steel sleeve. All positrons emanating from the source were assumed to annihilate in the surrounding metallic sleeve. The activity of the point source was 1.52 MBq (41 µCi) directly measured in a dose calibrator (Capintec CRC-10). The system was set to coincidence mode with an energy window of 250750 keV and a timing window of 6 ns. The source was stepped across the entire axial FOV in 4-mm steps. At each axial location, the source was first positioned at the center of the transverse plane and then stepped toward the edge of the gantry opening in 4-mm steps, data were acquired for 60 s, and true coincidence events were sorted into 2D sinograms. The number of true coincidences was normalized to the scan duration (60 s), divided by the source activity, and corrected for the branching ratio (0.89) of 68Ge. Attenuation of the stainless steel casing was not compensated for as the activity was referenced to the dose calibrator measurement. The results were plotted against the source locations to obtain the system sensitivity profile as a function of the axial and transverse locations. The same experiment was repeated under 2 other system conditions, 250750 keV with 10 ns and 350650 keV with 6 ns.
Counting Rate Performance
The counting rate performance of a PET system depends on the source distribution as well as the object dimension. Since the size of animals can vary drastically in different types of studies, counting rate performance was evaluated using phantoms that simulate the geometries of a mouse, a rat, and a monkey, the most commonly studied animals in our laboratory. The design of the phantoms was based on the National Electrical Manufacturers Association (NEMA) NU2-2001 Performance Measurements of Positron Emission Tomographs (25), with the phantom dimension scaled down to reflect the relative difference in object size between human and animal. Although these phantoms are an oversimplified representation of real source distributions in animals, they do cover the extreme distributions where most activity is within the imaging FOV (in a mouse study) and most activity is outside the imaging FOV (as in a monkey study).
All 3 phantoms were solid cylinders made of high-density (0.96 g/cm3) polyethylene. The mouse-like phantom was 3 cm in diameter and 7-cm long with a 3-mm hole drilled along the entire length at a radial offset of 0.75 cm from its central axis. The rat-like phantom was similarly constructed but 6 cm in diameter and 15 cm in length, and the hole was drilled at a 1.5-cm offset from the central axis. The monkey-like phantom was 10.1 cm in diameter and 40 cm in length and had a 4.8-mm-diameter hole drilled at 2.5-cm offset from the central axis. The activity was enclosed in glass tubing to have the exact same length as the respective phantoms.
For each phantom experiment, the corresponding line source was filled with >296 MBq (8 mCi) of 11C before insertion into the phantom. The phantoms were centered in the tomograph and data were collected every 23 min for 200 min (10 half-lives) at 2 energy windows (250750 and 350650 keV) and 2 coincidence timing windows (6 and 10 ns). The SSRB algorithm was used to sort the list mode data into 2D sinograms. The random coincidences were measured by the delayed window technique.
The noise equivalent count (NEC) was calculated using the following formula (26):
![]() | (Eq. 1) |
is the fraction of the object in the FOV, and Sf is the fraction of events included in the trues due to scatter.
Imaging Studies
All emission studies were collected in coincidence mode using 250- to 750-keV energy and 6-ns timing window. Random coincidences were subtracted and the detector efficiencies were normalized. List mode data were sorted into 3D sinograms with a span of 3 and a maximum ring difference of 47, followed by the FORE algorithm before the reconstruction using 2D FBP. The filter used was a ramp filter cutoff at the Nyquist frequency, unless otherwise specified.
Phantom Studies.
A custom-made, micro-Derenzo hot rod phantom was scanned in the Focus and the R4 (Rodent model) to visualize the improvement in image resolution. The phantom has an inner diameter of 32 mm and contains fillable hot rods of different sizes (0.80, 1.00, 1.25, 1.50, 2.00, and 2.50 mm) arranged into 6 segments. The spacing between adjacent rods in each segment is twice the rod diameter. The length of the rods is 25 mm. The phantom was filled with 18.5 MBq (0.5 mCi) of 18F-FDG and scanned for 20 min in each scanner.
Animal Studies.
An 18.7-g mouse was positioned in the gantry with its head centered in the FOV. A catheter was inserted into the jugular vein through which 17.8 MBq (0.48 mCi) of 2ß-11C-carbomethoxy-3ß-(4-fluorophenyl)tropane (11C-CFT) (27), in a specific activity >37 MBq/nmol (28), were injected. List mode data were acquired for 120 min and sorted into 35 dynamic frames before the images were reconstructed.
A 363-g SpragueDawley rat was injected with 35.7 MBq (0.97 mCi) of 18F-FDG via tail vein and 110 min was allowed for uptake before the rat was positioned in the scanner. Pediatric pads of an electrocardiogram-monitoring device, AccuSync 71 (AccuSync Medical Research Corp.), were connected to the paws of the rat. The AccuSyncss standard transistortransistor logic pulses were fed into the "Gating Input" of the Focus so that the trigger time would be inserted into the list mode data stream. The data were subsequently sorted into eight 3D sinograms representing 8 gates of the cardiac cycle. The images were reconstructed with a zoom of 3 using the OSEM algorithm. The same list mode data were also sorted into a single 3D sinogram without the gating option and reconstructed with the same parameters.
A 5.4-kg macaque monkey was positioned in the scanner with its brain centered in the FOV. The scanner was set up in singles mode using a 460- to 560-keV energy and 6-ns coincidence timing window. A 52-min transmission scan was acquired using a 68Ge point source. Immediately after the start of a 120-min dynamic scan, 166 MBq (4.5 mCi) of 6-18F-fluoro-L-dopa (18F-FDOPA) (29,30) were injected intravenously. List mode data of the emission scan were sorted into 31 dynamic frames. The transmission data were sorted using the SSRB algorithm and reconstructed with a Hanning filter cutoff at 50% of the Nyquist frequency. The choice of the Hanning filter over a ramp filter is 2-fold: to avoid amplification of statistical noise typically seen in transmission studies, and the resolution requirement for transmission images is not as high as that for emission images. Regions of interests were drawn on the transmission images to calculate the calibration factor that will scale down the linear attenuation coefficient of soft tissues to 0.095 cm1. The calibrated transmission images were forward projected to form 2D sinograms followed by inverse Fourier rebinning to generate a 3D attenuation correction file. The emission sinograms were normalized, corrected for attenuation and scatter (31), and reconstructed using FORE and 2D FBP.
All animal experiments were conducted in compliance with the Guidelines for the Care and Use of Research Animals established by the Animal Studies Committee at Washington University School of Medicine in St. Louis.
| RESULTS |
|---|
|
|
|---|
Spatial Resolution
Figure 1A shows the radial, tangential, and axial components of the reconstructed image resolution. At the CFOV, the image resolution is 1.30-mm FWHM for both radial and tangential directions and 1.46-mm FWHM for the axial direction. All 3 components remain well under 1.80-mm FWHM within 1.0-cm radial offset (2-cm FOV). The tangential component remains below 1.86-mm FWHM throughout the central imaging FOV up to a radial offset of 60 mm. The axial component remains below 2.06-mm FWHM up to a radial offset of 60 mm. The radial component increases at a higher rate, due to depth of interaction effects, to 1.90-, 2.53-, and 3.21-mm FWHM at 20-, 40-, and 60-mm radial offsets, respectively. Volumetric resolution, defined as the product of all 3 image resolution components of a system, is shown in Figure 1B and compared with that of the R4. The improvement of the Focus over the R4 ranges from 2.6-fold at the CFOV (2.5 µL vs. 6.6 µL) to 3-fold at 2-cm radial offset (6.3 µL vs. 19.2 µL). At 60-mm radial offset, the Focus still has a volumetric resolution of 12.3 µL, which is better than the volumetric resolution of the R4 everywhere except at the very center.
|
Counting Rate Performance
The projection fraction (
) of the mouse, rat, and monkey phantoms is 0.128, 0.256, and 0.430, respectively. The average scatter fraction of all 12 experiments calculated from the sinogram slices within the central 6-cm section of the axial FOV is listed in Table 2. The scatter fraction of the mouse, rat, and monkey phantoms ranges from 12.7% to 19.0%, from 27.4% to 36.5%, and from 40.9% to 58.1%, respectively.
|
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
The option of adjusting the energy window according to the animal size permits the user to optimize the acquisition protocol for different types of studies. For a rodent study that has a small amount of intrasubject scatter, the energy window can be set wider to accept events that undergo a single Compton interaction in the LSO crystals but still describe the correct line of response. When the subject is a primate, the amount of intrasubject scatter may be high enough to warrant a narrower energy window to reduce the scatter contribution to the final image. This is reflected in the system counting rate performance study using a monkey phantom where the highest NEC rate is achieved with a tighter energy window (350650 keV) and a tighter timing window (6 ns).
Due to the attenuation of the source housing, the effective activity of the 68Ge point source used in the calculation of system sensitivity was measured directly in a dose calibrator instead of decay corrected from the assayed value. The
-rays scattered in the source housing have a good chance of being recorded in the dose calibrator but may escape when imaged by the PET scanner. The activity presented to the PET camera may be less than that reported by the dose calibrator; therefore, the reported system sensitivity may be slightly underestimated.
For small animals such as mice, the amount of attenuation is fairly small and does not change significantly due to intersubject variability. Transmission scans are not typically performed for mouse studies at our institution. Instead, the amount of attenuation is assumed to be a constant for mouse-size objects and is factored into the system calibration to scale up the measured activity concentration. For studies involving large animals, such as primates, where the attenuation correction is critical for quantitative accuracy, we perform a transmission scan in singles mode using a smaller energy window (460560 keV). For the medium-size animals such as rats, we are still evaluating the trade-offs between the use of measured attenuation correction and the use of a predetermined amount of attenuation. If the accuracy of the attenuation correction using a short transmission scan can be further improved, the measured attenuation correction may be extended to the studies using medium-size animals.
Data management is an important issue for practical use of the Focus system. A typical 30-frame dynamic study may require 13 GB of disk storage for sinograms, whereas the list mode data may only take up 12 GB. This has forced us to change our data archival strategy from sinograms to list mode data for all dynamic studies. To further reduce the volume of data, 2 strategies can be applied: (a) reduce the transverse imaging FOV and (b) use a larger span (21) when sorting the data into 3D sinograms. The transverse imaging FOV can be reduced by software trimming when the list mode data are sorted into sinograms. The use of larger span is known to create axial blurring toward the edge of the transverse FOV. However, if the object is a mouse or a rat centered in the gantry, the loss in resolution may be negligible.
On measuring the axial resolution as a function of radial offset using a span of 7 when sorting the 3D sinograms, the axial resolution degrades by <5% within the central 50-mm-diameter FOV compared with a span of 3. This compromise reduces the size of a 3D sinogram to less than one half that of a file rebinned using a span of 3. If we limit the transverse FOV to 5 cm and use a span of 7, then the file size drops to one tenth of the original with little degradation. This is of benefit to users who have a large number of dynamic studies using mice.
The 3D reprojection (33), 3D OSEM, and maximum a posteriori (MAP) (34) reconstruction algorithms became available from the manufacturer after this study was completed. It is expected that the more sophisticated statistical reconstruction algorithms, along with improved modeling of the physical characteristics of the system, will further improve the resolution and noise properties of the reconstructed images. Evaluation of the improvement in image resolution using these algorithms is not included in this work and will be studied in the future.
When the system was released, the calculation of the normalization file was based on a direct inversion of 3D sinograms acquired from a rotating 68Ge point source. For a typical 3D sinogram that has >100 million lines of response, it is difficult to obtain high-quality normalization files with good statistics. Even with 100 GB of list mode data, the average number of counts per line of response is only slightly more than 100. Poisson statistics dictate that the variance may then be as high as 10%. A new component-based normalization procedure has been released by the manufacturer to improve the accuracy of the normalization. An initial evaluation indicates improved uniformity using this new method. The quantitative accuracy of both normalization methods, along with the accuracy of attenuation, scatter, and dead time corrections, will require further investigation.
| CONCLUSION |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
For correspondence contact: Yuan-Chuan Tai, PhD, Washington University School of Medicine, Campus Box 8225, 510 S. Kingshighway Blvd., St. Louis, MO 63110.
E-mail: taiy{at}wustl.edu
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
S. R. Bayly, R. C. King, D. J. Honess, P. J. Barnard, H. M. Betts, J. P. Holland, R. Hueting, P. D. Bonnitcha, J. R. Dilworth, F. I. Aigbirhio, et al. In Vitro and In Vivo Evaluations of a Hydrophilic 64Cu-Bis(Thiosemicarbazonato)-Glucose Conjugate for Hypoxia Imaging J. Nucl. Med., November 1, 2008; 49(11): 1862 - 1868. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. H. Wei, T. Olafsen, C. Radu, I. J. Hildebrandt, M. R. McCoy, M. E. Phelps, C. Meares, A. M. Wu, J. Czernin, and W. A. Weber Engineered Antibody Fragments with Infinite Affinity as Reporter Genes for PET Imaging J. Nucl. Med., November 1, 2008; 49(11): 1828 - 1835. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Vandervoort and V. Sossi Impact of Contamination from Scattered Photons in Singles-Mode Transmission Data on Quantitative Small-Animal PET Imaging J. Nucl. Med., November 1, 2008; 49(11): 1852 - 1861. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Wu, D. Pal, T. Y. Song, J. A. O'Sullivan, and Y.-C. Tai Micro Insert: A Prototype Full-Ring PET Device for Improving the Image Resolution of a Small-Animal PET Scanner J. Nucl. Med., October 1, 2008; 49(10): 1668 - 1676. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. I. Shoghi, R. J. Gropler, T. Sharp, P. Herrero, N. Fettig, Y. Su, M. S. Mitra, A. Kovacs, B. N. Finck, and M. J. Welch Time Course of Alterations in Myocardial Glucose Utilization in the Zucker Diabetic Fatty Rat with Correlation to Gene Expression of Glucose Transporters: A Small-Animal PET Investigation J. Nucl. Med., August 1, 2008; 49(8): 1320 - 1327. [Abstract] [Full Text] [PDF] |
||||
![]() |
H.-Y. Lee, Z. Li, K. Chen, A. R. Hsu, C. Xu, J. Xie, S. Sun, and X. Chen PET/MRI Dual-Modality Tumor Imaging Using Arginine-Glycine-Aspartic (RGD)-Conjugated Radiolabeled Iron Oxide Nanoparticles J. Nucl. Med., August 1, 2008; 49(8): 1371 - 1379. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Yang, Y. Wu, J. Qi, S. St. James, H. Du, P. A. Dokhale, K. S. Shah, R. Farrell, and S. R. Cherry A Prototype PET Scanner with DOI-Encoding Detectors J. Nucl. Med., July 1, 2008; 49(7): 1132 - 1140. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. H. Wei, H. Su, I. J. Hildebrandt, M. E. Phelps, J. Czernin, and W. A. Weber Changes in Tumor Metabolism as Readout for Mammalian Target of Rapamycin Kinase Inhibition by Rapamycin in Glioblastoma Clin. Cancer Res., June 1, 2008; 14(11): 3416 - 3426. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. C. Huisman, T. Higuchi, S. Reder, S. G. Nekolla, T. Poethko, H.-J. Wester, S. I. Ziegler, D. S. Casebier, S. P. Robinson, and M. Schwaiger Initial Characterization of an 18F-Labeled Myocardial Perfusion Tracer J. Nucl. Med., April 1, 2008; 49(4): 630 - 636. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y.-C. Tai, H. Wu, D. Pal, and J. A. O'Sullivan Virtual-Pinhole PET J. Nucl. Med., March 1, 2008; 49(3): 471 - 479. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. L. Vavere, S. J. Kridel, F. B. Wheeler, and J. S. Lewis 1-11C-Acetate as a PET Radiopharmaceutical for Imaging Fatty Acid Synthase Expression in Prostate Cancer J. Nucl. Med., February 1, 2008; 49(2): 327 - 334. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. M. Sharkey, H. Karacay, S. Vallabhajosula, W. J. McBride, E. A. Rossi, C.-H. Chang, S. J. Goldsmith, and D. M. Goldenberg Metastatic Human Colonic Carcinoma: Molecular Imaging with Pretargeted SPECT and PET in a Mouse Model Radiology, February 1, 2008; 246(2): 497 - 507. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Fonge, K. Vunckx, H. Wang, Y. Feng, L. Mortelmans, J. Nuyts, G. Bormans, A. Verbruggen, and Y. Ni Non-invasive detection and quantification of acute myocardial infarction in rabbits using mono-[123I]iodohypericin {micro}SPECT Eur. Heart J., January 2, 2008; 29(2): 260 - 269. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Wu, D. Pal, J. A. O'Sullivan, and Y.-C. Tai A Feasibility Study of a Prototype PET Insert Device to Convert a General-Purpose Animal PET Scanner to Higher Resolution J. Nucl. Med., January 1, 2008; 49(1): 79 - 87. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. B. Chin, S. D. Metzler, A. Lemaire, A. Curcio, S. Vemulapalli, K. L. Greer, N. A. Petry, T. G. Turkington, R. E. Coleman, H. Rockman, et al. Left Ventricular Functional Assessment in Mice: Feasibility of High Spatial and Temporal Resolution ECG-gated Blood Pool SPECT Radiology, November 1, 2007; 245(2): 440 - 448. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. Maeda, B. Ji, T. Irie, T. Tomiyama, M. Maruyama, T. Okauchi, M. Staufenbiel, N. Iwata, M. Ono, T. C. Saido, et al. Longitudinal, Quantitative Assessment of Amyloid, Neuroinflammation, and Anti-Amyloid Treatment in a Living Mouse Model of Alzheimer's Disease Enabled by Positron Emission Tomography J. Neurosci., October 10, 2007; 27(41): 10957 - 10968. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. S. Kim, J. S. Lee, K. C. Im, S. J. Kim, S.-Y. Kim, D. S. Lee, and D. H. Moon Performance Measurement of the microPET Focus 120 Scanner J. Nucl. Med., September 1, 2007; 48(9): 1527 - 1535. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Wei, C. Butcher, Y. Miao, F. Gallazzi, T. P. Quinn, M. J. Welch, and J. S. Lewis Synthesis and Biologic Evaluation of 64Cu-Labeled Rhenium-Cyclized {alpha}-MSH Peptide Analog Using a Cross-Bridged Cyclam Chelator J. Nucl. Med., January 1, 2007; 48(1): 64 - 72. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. R. Stickel, J. Qi, and S. R. Cherry Fabrication and Characterization of a 0.5-mm Lutetium Oxyorthosilicate Detector Array for High-Resolution PET Applications J. Nucl. Med., January 1, 2007; 48(1): 115 - 121. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Wei, J. Easmon, R. K. Nagi, B. D. Muegge, L. A. Meyer, and J. S. Lewis 64Cu-Azabicyclo[3.2.2]Nonane Thiosemicarbazone Complexes: Radiopharmaceuticals for PET of Topoisomerase II Expression in Tumors J. Nucl. Med., December 1, 2006; 47(12): 2034 - 2041. [Abstract] [Full Text] [PDF] |
||||
![]() |
Y. Wang, J. Seidel, B. M.W. Tsui, J. J. Vaquero, and M. G. Pomper Performance Evaluation of the GE Healthcare eXplore VISTA Dual-Ring Small-Animal PET Scanner J. Nucl. Med., November 1, 2006; 47(11): 1891 - 1900. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Su, C. Bodenstein, R. A. Dumont, Y. Seimbille, S. Dubinett, M. E. Phelps, H. Herschman, J. Czernin, and W. Weber Monitoring Tumor Glucose Utilization by Positron Emission Tomography for the Prediction of Treatment Response to Epidermal Growth Factor Receptor Kinase Inhibitors. Clin. Cancer Res., October 1, 2006; 12(19): 5659 - 5667. [Abstract] [Full Text] [PDF] |
||||