Abstract
Performing PET imaging during ongoing radionuclide therapy can be a promising method to follow tumor response in vivo. However, the high therapeutic activity can interfere with the PET camera performance and degrade both image quality and quantitative capabilities. As a solution, low-energy photon emissions from the therapeutic radionuclide can be highly attenuated, still allowing sufficient detection of annihilation photons in coincidence. Methods: Hollow Rose metal cylinders with walls 2–4 mm thick were used to shield a 22Na point source and a uniform phantom filled with 18F as they were imaged on a preclinical PET camera with increasing activities of 177Lu. A mouse with a subcutaneous tumor was injected with 18F-FDG and imaged with an additional 120 MBq of 177Lu and repeated with shields surrounding the animal. Results: The addition of 177Lu to the volume imaged continuously degraded the image quality with increasing activity. The image quality was improved when shielding was introduced. The shields showed a high ability to produce stable and reproducible results for both spatial resolution and quantification of up to 120 MBq of 177Lu activity (maximum activity tested). Conclusion: Without shielding, the activity quantification will be inaccurate for time points at which therapeutic activities are high. The suggested method shows that the shields reduce the noise induced by the 177Lu and therefore enable longitudinal quantitative intratherapeutic imaging studies.
To improve cancer therapy, there is a need for sensitive and quantitative molecular imaging techniques to follow therapeutic efficacy and progress of disease (1,2).
In preclinical research, molecular imaging techniques such as PET help decipher the response to treatment in animal tumor models. Changes in proliferation, hypoxia, angiogenesis, and other factors important to tumor growth and survival can be monitored in vivo with PET tracers (2–5). PET adds physiologic information that can have better prognostic value than measuring tumor size and volume with calipers or with morphologic imaging such as CT or MRI (6). PET tracers enable longitudinal studies, allowing measurements of both early and late treatment response (7).
Radionuclide therapy (RNT) is an expanding field of systemic therapies in which tumor-targeting molecules are labeled with radionuclides to deliver localized radiotherapy to tumor cells with high specificity. 177Lu-DOTATATE is a recent effective RNT for the treatment of neuroendocrine tumors, and several other promising treatments are currently under development (8,9), such as 177Lu-prostate-specific membrane antigen treatment for metastatic prostate cancer (10–12). Radionuclides used for RNT are chosen for their mode of decay and the linear energy transfer of the emitted particles. To follow response markers in vivo with PET can be both a step toward more personalized medicine and a tool to broaden the radiobiologic understanding of RNT.
However, photon emissions from RNT radionuclides can interfere during intratherapeutic PET. 177Lu has γ-emissions of 113 and 208 keV, with emission yields of 6.2% and 10.4%, respectively (13). This excess photon flux risks the addition of noise and distortions. Although the low-energy photons can be rejected by energy discrimination, the detector electronics must still process these events. The result may be dead-time losses and loss in sensitivity. Accepted single events from the photon background are also likely to generate a significant amount of random coincidences that will add noise to the image once subtracted. A large photon flux is also likely to result in pulse pileup, which might interfere with the detection of the annihilation photons and will affect the spatial resolution (14,15). We examined this problem on several preclinical PET scanners (16) and found that some systems have limited abilities in handling a high background photon flux from a therapeutic source and display a reduced spatial resolution and image artifacts. In a longitudinal study of an animal undergoing an RNT, where the therapeutic activity will decay from high to low over time, the quantification of the therapeutic effect measured with PET will be uncertain.
We propose a metal shield that surrounds the animal during intratherapeutic PET imaging. The attenuation differs such that lower-energy photons will be more attenuated than the 511-keV photons. Transmission through a Rose metal shield as a function of shield thickness calculated from mass attenuation factors for Rose metal (17) are presented in Figure 1.
The number of transmitted photons I through Rose metal shield in terms of the number without a shield I0 for relevant photon energies representing γ-emissions from 177Lu and 511-keV annihilation photons.
Absorbing shields have previously been used in PET imaging. Muehllehner (18) used a graded absorber to attenuate γ-rays scattered in the patient. This was followed up by Spinks and Shah (19), who evaluated the effect of lead filters on a neuro-PET system. Goertzen et al. investigated the opportunity to perform simultaneous PET and CT by shielding the PET detector with lead (20).
In this paper, image quality is evaluated in a manner similar to that performed by Goertzen et al. evaluating small-animal PET systems (21). They followed the National Electrical Manufacturers Association 4-2008 protocol that standardizes how to report preclinical PET imaging parameters (22). We report similar measurements with a uniform volume of 18F and a 22Na point source. We also show in vivo proof of concept in a xenografted mouse. Additionally, Monte Carlo simulations are performed to evaluate the count rate effects of shields and energy window discrimination.
MATERIALS AND METHODS
Shields
The attenuation of low-energy photons was achieved with cylindric shields of Rose metal. Rose metal is an alloy consisting of 50% bismuth, 25%–28% lead, and 22%–25% tin. It has a low melting point and does not contract after cooling, making it suitable for custom-made radiation shields. It is also less toxic than lead, favoring its use (23). Three cylinders of Rose metal were constructed with an inner diameter of 35 mm and a wall thickness of 2, 3, or 4 mm and were all 9 cm long (Fig. 2).
Shield of Rose metal for intratherapeutic PET imaging of mice during RNT. Shields used were 2, 3, and 4 mm thick (here shown with 4-mm thickness). All had inner diameter of 35 mm.
PET Imaging and Image Evaluation
The preclinical G8 PET/CT system (Sofie BioSciences, PerkinElmer) (24), the successor of the preclinical PET/CT system Genisys 4 (Sofie BioSciences) (25), was used in this study. The G8 applies a dead-time correction to the data; however, it is accurate only up to a photon flux produced by a mouse injected with 3 MBq of 18F-FDG.
To run imaging procedures with the shield surrounding the phantom or mouse, a plastic sheet was constructed and used in place of the system’s ordinary animal bed. The system’s default energy window of 150–650 keV was used in all measurements. All acquisitions were reconstructed as static frames with the system’s maximum-likelihood expectation-maximization (60 iterations) algorithm.
VivoQuant (version 3.0; inviCRO) was used to select representative images and to draw volumes of interest (VOIs) over regions in the phantom studies of the reconstructed DICOM images. MATLAB (R2017b; MathWorks) was used to select line profiles through the 22Na point source in the images and to calculate the full width at half maximum (FWHM) for the axial, coronal, and sagittal projections. The volume FWHM was calculated as the product of the axial, sagittal, and coronal FWHMs.
Count Rate Characteristics
A 70-kBq 22Na point source (half-life, 2.6 y) was imaged together with 0.1–50 MBq of 99mTc-sodiumpertechnetate (Triad Isotopes Inc.) dissolved in water in a 1-mL syringe. The point source was placed at the center of the FOV on a stand; the syringe was centered along the transversal axis, just below the point source. The acquisition was made as a dynamic scan, acquiring data for 5 min every hour until the initial 50 MBq of 99mTc activity had decayed to less than 0.1 MBq. The count rates of coincidence events (prompt, delayed, and true events) for the 5-min acquisitions were retrieved from the histogram file.
The low-energy γ-photons emitted from 99mTc (141 keV) adds a background photon flux that will interact in the PET detectors, simulating the presence of a therapeutic radionuclide in the PET camera. The short half-life of 6.0 h of 99mTc, compared with 6.7 d for 177Lu, allowed a broad photon flux range to be evaluated within a reasonable time frame. Emission branching leads to about a factor of 5 times higher photon flux per decay from 99mTc than from 177Lu.
Point Source
The 22Na point source was placed on a stand at the center of the field of view (FOV) together with a spheric hollow phantom (Hollow Sphere Sets, 0.5 mL; Data Spectrum) filled with 0–120 MBq of 177Lu (77Lu-DOTATATE; Advanced Accelerator Applications) dissolved in 0.5 mL of water. The 177Lu sphere phantom (outer diameter, 1.2 cm) was placed at the foot of the stand, approximately 1 cm below the point source. This setup was imaged in 5-min static acquisitions without and with increasing thickness of shields (2–4 mm) surrounding both point source and sphere phantom. The volume FWHM of the reconstructed point source was evaluated. The prompt coincidence count rates were extracted from the histogram files to evaluate the effect of the 177Lu background on the coincidence count rate. A spheric VOI with 7.3-mm diameter (200 mm3) was placed over the 22Na point source in the reconstructed images. Image counts in the VOIs were summed and normalized by the summed VOI image counts from the point source when measured without any shield in place or 177Lu activity present in the FOV.
Uniform 18F Phantom
A 20-mL syringe (inner diameter, 21.2 mm) filled with 790 kBq of 18F-FDG (produced at the Biomedical Cyclotron Facility at UCLA) was centered along the transversal axis and imaged alone and together with 0–120 MBq of 177Lu in the 0.5-mL sphere phantom placed on top of the syringe at its middle point. Static 10-min acquisitions were made without the shields, and then with the phantoms placed inside shields of increasing thickness (2–4 mm). A cylindric VOI (710 mm3) was centered inside the 18F syringe phantom in the reconstructed images. The relative SD (RSD) (coefficient of variation) of the image counts in the VOI was calculated as previously reported (21,22). The sum of image counts detected in the VOI was normalized by the events detected from the 18F phantom measured without any shield or 177Lu activity present in the FOV.
Mouse 18F-FDG Imaging
The animal experiment was approved by the UCLA Animal Research Committee and was performed according to the guidelines of the Division of Laboratory Animal Medicine at UCLA.
One C57BL/6 mouse bearing a RM1-PGLS–based subcutaneous tumor in the shoulder region was injected with 790 kBq of 18F-FDG. After 1 h, the mouse was euthanized and repeatedly imaged in 10-min static acquisitions: first by itself with no 177Lu present and then with the spheric phantom containing 120 MBq of 177Lu taped to the right side of its back. Lastly, the animal and the 177Lu source were imaged with each of the shields (2–4 mm).
Monte Carlo Simulations of the PET System and Shielding
To get a better understanding of the origin of the image distortions during intratherapeutic PET imaging, spectra of the energy depositions in the PET camera’s bismuth germanate detectors were simulated with GATE 8.0, a software for Monte Carlo simulation of medical imaging and radiotherapy (26). Simulating the shields also clarifies the underlying attenuation effects of our method’s performance. A model of the Genisys4 preclinical PET system, whose detectors have similar properties and geometry to the G8 PET/CT system, and a cylindric phantom filled with 10 MBq of 177Lu in water (radius, 1 cm; height, 5 cm) and a point source of 100 kBq of 18F placed at the center of the cylindric phantom were simulated. Simulations were repeated with a 4-mm-thick Rose metal shield surrounding the cylinder phantom and point source. The energy depositions of the interactions in the detector volumes were recorded. At least 107 initial emissions were generated in both simulations.
A digitizer model describing the detection of singles and coincidences among the energy depositions in the detectors was optimized to reproduce measured single and coincidence rates from 20 kBq to 17 MBq of 18F dissolved in water in a 20-mL cylindric phantom (Supplemental Figs. 1A and 1B; supplemental materials are available at http://jnm.snmjournals.org). The dead time was estimated to 1,550 ns from the singles rate, and the model assumed a paralyzable dead-time behavior. The digitizer model was then used to simulate the coincidence rate when 100 kBq of 18F and 0–30 MBq of 177Lu homogeneously distributed within a cylindric phantom was imaged with wide (150–650 keV) or narrow (350–650 keV) energy windows, with and without a 3-mm Rose metal shield surrounding the phantom.
RESULTS
Count Rate Characteristics
The detected count rates from the 22Na point source as a function of 99mTc activity are shown in Figure 3. The true coincidence count rate dropped off almost exponentially as a function of increasing 99mTc activity because of count losses due to the system’s dead time, produced by the increasing flux of low-energy photons. At about 3.4 MBq of 99mTc, the true coincidence count rate was reduced to half that generated by the point source itself. At 6.2 MBq, the true coincidence count rate was 32% (1,321/4,251 counts/s) and the numbers of true and delayed coincidences were equal, a point often used as a cutoff for when count rates are too high for accurate PET imaging.
Coincidence counts detected in 5-min time frames at beginning of every hour as 99mTc-filled syringe was imaged together with 70-kBq 22Na point source. As 99mTc decayed, broad interval (>0.1–50 MBq) of low-energy photon flux was examined to evaluate effect on coincidence count rate. Dead-time losses to true coincidence count rate increased as 99mTc activity increased.
Point Source
The volume spatial resolution of the 22Na point source as a function of increasing activities of 177Lu in the FOV is presented in Figure 4A and shows an increasing volume FWHM with increasing 177Lu activity. The use of shields recovered the volume FWHM to approximately the same spatial resolution as measured for the 22Na point source alone.
Results from imaging of 22Na point source with increasing 177Lu activity present in FOV. Imaging was performed with 0-, 2-, 3-, and 4-mm Rose shields surrounding activity. (A) Image resolution evaluated as volume FWHM of point source with and without increased shield width. (B) Prompt coincidence count rate with increasing thickness of shields. Points are normalized to data point where no 177Lu activity or shield surrounded 22Na point source. (C) Normalized sum of image counts detected in spheric VOI placed around point source.
The prompt count rates show the attenuating differences between the shields in Figure 4B. The prompt coincidence count rate increased slightly for all shields for increased 177Lu activity, but with a reduced effect for increasing thickness. Without shielding, the prompt coincidences reached a maximum count rate at 40 MBq of 177Lu. As the 177Lu activity was further increased, dead-time losses reduced the coincidence count rate.
The image counts detected in the VOI around the point source are presented in Figure 4C. The signal fell off almost exponentially as the 177Lu activity was increased. The data in Figure 4C show the reduction in number of coincidences detected because of the attenuation of the shields, which results in an effective sensitivity reduction of the system. The mean and SD of the normalized sum of image counts over the 177Lu interval of 0–120 MBq for the 4-mm shield was 25% ± 1%. This large reduction in image counts results in an increase in image noise or requires increased imaging time. The reduction will require a new calibration factor for quantitative measurement of the PET tracer. However, the low variation in counts over the interval of 0–120 MBq shows that reliable quantitative intratherapeutic studies are possible to perform.
Uniform 18F Phantom
The image quality of a cylindric phantom with a uniform distribution of 790 kBq ± 40 kBq (mean ± SD) of 18F activity was clearly reduced when an increasing amount of 177Lu activity was introduced to the FOV, as demonstrated by the axial images in Figure 5A. The use of shields restored the expected volume distribution of 18F inside the uniform phantom (Fig. 5A, image collected using a 2-mm shield).
Results from 790-kBq 18F in uniform phantom imaged with 0- to 120-MBq 177Lu in spheric phantom and 0-, 2-, 3-, and 4-mm Rose metal shields surrounding activity. (A) Axial projections at centrum of uniform phantom filled with 790 kBq of 18F together with 0, 40, 80, and 100 MBq of 177Lu. In final image, 2-mm Rose shield was added. (B) RSD of image counts measured with cylindric VOI inside 18F-filled phantom. (C) Normalized sum of image counts detected in cylinder VOI in 18F phantom. Mean and SD for 0- to 120-MBq measurements with shields are presented.
The RSD of image counts inside the uniform volume with 18F as a function of 177Lu activity in the FOV is shown in Figure 5B. Without a shield, the RSD increases with increasing 177Lu activity. In the absence of 177Lu, RSD increases from 9% without shielding to 13%–15% with the shields in place. This small increase in noise is expected because of the attenuation of counts in the shields and results in a slight reduction in image quality and increasing variance of a quantitative measurement. However, the presence of the shields results in an RSD that is relatively constant within the range of 177Lu activities investigated (11%–18%).
The normalized sum of 18F image counts in the cylindric VOI is presented in Figure 5C. Without shields and with increasing 177Lu activity present, the system lost the ability to reproduce a constant number of image counts in the VOI, but once the shielding was added, repeated measurements over the range of 177Lu activity tested were able to reproduce the image counts, thereby again demonstrating the possibility to perform quantitative studies over a range of 177Lu activities present in the FOV.
Mouse 18F-FDG imaging
Figure 6 shows maximum-intensity projections from a mouse injected with 790 kBq of 18F-FDG and repeatedly imaged 1 h after injection. Figure 6A with no shielding or 177Lu activity added shows only 18F-FDG uptake. Expected uptake in heart, kidney, and subcutaneous tumor in the right shoulder area was observed. In Figure 6B, 120 MBq of 177Lu activity in a sphere phantom was taped to the animal’s right flank. As can be seen in this image, the 18F-FDG distribution is completely distorted in the presence of 177Lu. In Figure 6C, 4 mm of Rose metal shielding was placed around the mouse and the 177Lu phantom. The 18F-FDG distribution in the mouse when imaged with the Rose metal shields is now restored and comparable to the image in 6A. With the shielding in place, the 18F-FDG uptake in the subcutaneous tumor on the shoulder and the heart and kidneys was again clearly distinguishable.
Maximum-intensity projections of mouse injected with 790 kBq of 18F-FDG and imaged in regular setup (A), with 120 MBq of 177Lu in spheric phantom taped to flank (B), and with 4-mm Rose metal shield surrounding mouse and 177Lu phantom (C).
Monte Carlo Simulation of PET System and Shielding
The energy deposited in the bismuth germanate detectors by a 18F point source inside a 177Lu-filled water phantom was simulated with the GATE 8.0 Monte Carlo simulation toolkit. Without the shield, the rate of energy deposition interactions from the 10-MBq 177Lu phantom and 100-kBq 18F point source (including secondary scattered events) was 3.87 × 107 per second. Of these, 3.73 × 107 per second originated from the 177Lu and 0.14 × 107 per second from 18F. Adding a 4-mm Rose metal shield lowered the rate of interactions to 4.22 × 106 per second, and the interactions originating from 177Lu and 18F were reduced to 3.42 × 106 (9.2%) and 0.81 × 106 (57.9%), respectively. Figure 7A is a histogram of these photon interactions of energy deposition in the PET detectors. The low-energy interactions (<511 keV) dominate, explaining the dead-time losses seen in Figures 3 and 4C, where radionuclides emitting low-energy photons contributed high photon flux interacting in the detectors. In Figure 7B, interactions from the simulation run with a 4-mm Rose metal shield surrounding the phantom are shown. The low-energy interactions from 177Lu are reduced to 9.2%, but the shield still allows 57.9% of the annihilation photons from the 18F source to be detected.
Energy deposited in detectors, as simulated by GATE (version 8.0) Monte Carlo simulation toolkit. A 100-kBq 18F point source centered inside water-filled cylindric phantom with 10 MBq of 177Lu homogeneously distributed inside it were set up inside PET camera FOV. (A) Simulation without shield. (B) Simulation with 4-mm Rose metal shield surrounding phantom and sources.
The digitizer model showed good agreement with measured single and coincidence count rates (Supplemental Figs. 1A and 1B). Simulating increasing 177Lu activity present in the phantom volume continuously lowered the prompt coincidence rates for both the wide (150–650 keV) and the narrow (350–650 keV) energy windows (Supplemental Fig. 1C). The random coincidence rate was significantly reduced using the narrow window. The decrease in prompt rate with increased 177Lu activity is still observed, as this is due to dead-time losses at high count rates. Simulating a 3-mm Rose shield around the phantom showed that the prompt coincidence rate was less affected by increasing 177Lu activity for both energy windows.
DISCUSSION
In recent years, new models of small preclinical PET cameras have been introduced on the market. These systems can be influential in the development of new RNTs, offering an in vivo opportunity to follow treatment effect in tumor models. The sensitivity to overall system dead-time losses will be a problem in small preclinical systems, which often excel in high sensitivity (21). Fortunately, with shielding, the system’s high sensitivity is also a helpful factor when the signal from the PET tracer, inevitably reduced by the shield, still can be successfully detected as shown in this study. Also, injected PET tracer activity could potentially be increased to compensate for the signal loss. The attenuation of annihilation photons in the shields will also result in an increase in scattered photons, which may affect quantification. A narrow energy window could compensate.
The addition of 177Lu activities relevant in preclinical RNT (27) reduced the spatial resolution as seen in Figure 4A, increased the noise, and obscured the images beyond the point at which they could contribute useful and quantifiable information, as shown in Figures 5C and 6B. The shields produced a consistent image resolution that was unaffected by the 177Lu activity interval tested. The detected activity in uniform volumes of 18F remained constant within the investigated activity range of 177Lu activities in the FOV (Figs. 4C and 5B). The use of the shields would therefore allow quantitative imaging and longitudinal studies. The reduction in sensitivity could be compensated by increasing the acquisition time.
The simulated effects of the shields are presented in Figure 7. Rose metal attenuates the photons from 177Lu to a higher degree than the 511-keV annihilation photons. Reducing detected events from 177Lu γ-emissions lowers the total detector interaction count rate and reduces dead-time losses of true annihilation photon events. Simulations also showed how count losses due to high count rates could not be compensated for using a narrow energy window (Supplemental Fig. 1C). However, for systems tolerating higher count rates or with lower sensitivity, the use of a narrow energy window can be a component in optimization for intratherapeutic PET.
Despite the slight loss of image quality, PET imaging during RNT with shielding can provide clear images of a PET tracer in vivo, as shown in Figure 6C, and makes quantitative measures possible. Our proposed solution makes RNT intratherapeutic PET imaging a promising, easy, and inexpensive method to overcome the deteriorating effect.
CONCLUSION
Performing preclinical PET imaging during RNT can lead to loss of image quality and distorted quantitative measurements. Shielding makes intratherapeutic PET imaging possible, even with highly sensitive preclinical block detector PET systems. If the effects of a therapeutic radionuclide present in the animal are not accounted for, quantitative studies can underestimate the uptake of a PET tracer because of count losses at high detector dead time. For longitudinal studies of animal models undergoing RNT, this will distort data of activity uptake as the therapeutic radionuclide decays over time.
DISCLOSURE
Financial support was provided by the Swedish Cancer Society and Mrs. Berta Kamprad’s Foundation. Susan Evans-Axelsson was supported by a Swedish Research Council International postdoc grant (grant 2015-00452). No other potential conflict of interest relevant to this article was reported.
Acknowledgments
Thanks to Zheng Gu (Crump Institute, UCLA) for providing the GATE model. Thanks to Lars Ljus and the staff at the Radiation Therapy Unit workshop at Skane University Hospital Lund for help constructing the shields.
Footnotes
Published online Nov. 2, 2018.
- © 2019 by the Society of Nuclear Medicine and Molecular Imaging.
REFERENCES
- Received for publication July 3, 2018.
- Accepted for publication October 5, 2018.