Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • View or Listen to JNM Podcast
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
Meeting Report

Deployment of a dose-reduction algorithm in Molecular breast Imaging - impact of reader experience on its optimization

Ashlee Stanke and Michael O'Connor
Journal of Nuclear Medicine May 2020, 61 (supplement 1) 3115;
Ashlee Stanke
1Radiology Mayo Foundation Rochester MN United States
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Michael O'Connor
1Radiology Mayo Foundation Rochester MN United States
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
Loading

Abstract

3115

Objectives: Molecular Breast Imaging, performed with Tc-99m sestamibi and small CZT-based detectors is a valuable technique for detection of breast cancers occult on mammography. Previous work has shown that a noise reduction algorithm (ClearMBI) allows for a 50% reduction in image acquisition time, while yielding improved image quality relative to a standard acquisition. Optimization of the filter settings for ClearMBI in our lab was accomplished by experienced breast radiologists and nuclear medicine technologists. Deployment of this algorithm in sites with little or no experience with MBI may not achieve these optimal results. Hence, the objective of this study was to determine if this optimization was dependent on user experience with MBI

Methods: Two groups of readers were selection. Group 1 comprised a total of 4 breast imagers / scientists and 4 nuclear medicine technologists with each more than 2 year’s experience with MBI (>2000 patient studies/reader). Group 2 comprised 2 nuclear medicine physicians and 4 nuclear medicine technologists with no experience with MBI. Each reviewer was presented with a test set of 20 MBI exams (each containing a lesion) representing a range of count densities. For each exam, only the original half-dose image was displayed, and the reviewer was allowed to freely adjust the ClearMBI filter setting until their preferred lesion conspicuity was achieved. The average filter setting as a function of counts per pixel in the MBI images was obtained for each group and compared to the previously derived equation that was used to validate the algorithm.

Results: Table 1 below shows the previously reported filter settings, those from experienced reviewers, and those from nuclear medicine physicians and technologists (filter value % = slope * mean cts/pixel in image + intercept). Figure 1 shows an example of an image filtered using the previous published filter values and the corresponding values from Table 1 above. By comparison to experienced reviewers, the nuclear medicine physicians tended to over denoised the MBI images, whereas the technologists tended to under denoise the images. Lack of consistency in use of the algorithm may reduce acceptance by physicians. Implementation of this algorithm into a clinical practice that is initiating an MBI service should provide optimized values of the filter settings to ensure consistent results.

Conclusions: Implementation of the ClearMBI algorithm is strongly dependent on user experience. For new installations with little or no experience in MBI, it is recommended that staff use the previously published filter settings, rather than determining their own.

View this table:
  • View inline
  • View popup

Table 1

Figure
  • Download figure
  • Open in new tab
  • Download powerpoint
Previous
Back to top

In this issue

Journal of Nuclear Medicine
Vol. 61, Issue supplement 1
May 1, 2020
  • Table of Contents
  • Index by author
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Deployment of a dose-reduction algorithm in Molecular breast Imaging - impact of reader experience on its optimization
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
Deployment of a dose-reduction algorithm in Molecular breast Imaging - impact of reader experience on its optimization
Ashlee Stanke, Michael O'Connor
Journal of Nuclear Medicine May 2020, 61 (supplement 1) 3115;

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Deployment of a dose-reduction algorithm in Molecular breast Imaging - impact of reader experience on its optimization
Ashlee Stanke, Michael O'Connor
Journal of Nuclear Medicine May 2020, 61 (supplement 1) 3115;
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
  • Figures & Data
  • Info & Metrics

Related Articles

  • No related articles found.
  • Google Scholar

Cited By...

  • No citing articles found.
  • Google Scholar

More in this TOC Section

  • Rest Dose Spillover Correction of Stress Blood Flow Measurements in Digital Rb-82 Myocardial Perfusion PET/CT Imaging
  • Radiolabeled hyaluronic acid (HA) fragments for lymphatic imaging
  • Simplifying and Confirming Ketosis: A Quality Improvement Initiative in Cardiac Sarcoid Assessment
Show more

Technologist Poster Session

  • Development of Radiolabeling Methods for 5 Fluorouracil (5FU) with Technetium-99mfor Preclinical Nuclear Imaging (PNI)
  • TRANQUILTY SCORING TO OPTIMIZE PEDIATRIC IMAGING AND REDUCE RADIATION EXPOSURE ON TOTAL-BODY PET SCANNERS
  • Evaluation of the dedicated breast PET imaging technique including axillary region
Show more Technologist Poster Session

Similar Articles

SNMMI

© 2025 SNMMI

Powered by HighWire