Abstract
As a result of a recent change in the National Institute of Standards and Technology (NIST) activity standard for 18F, we have determined new relative response ratios for a 68Ge solid epoxy mock syringe source used in activity calibrators as a long-lived substitute for 18F. New standardized solutions of each radionuclide were used to determine the response ratios while maintaining traceability to national standards. This work updates our previously published data from 2010. Methods: Following our previously published methodology, solution-filled mock syringe sources, identical in geometry to the solid 68Ge epoxy calibration source currently on the market, were prepared using NIST-calibrated solutions of 68GeCl4 and 18F-FDG and directly compared in several models of activity calibrators to determine empirically the relative response ratios for these 2 radionuclides. Results: The new relative response ratios measured in this study reflect the change in 18F activity measurements that arise from the recent −4% change in the NIST activity standard. The results allow the 68Ge activity of the mock syringe source to be expressed in terms of equivalent 18F activity, with a relative combined standard uncertainty of about 0.8% for the activity calibrators used in this study. Conclusion: This work revises our previously derived relative response ratios for 18F and 68Ge by −3.7%, allowing users of the commercial mock syringe surrogate source to calibrate their activity calibrators in a way that is consistent with the recent change in the NIST 18F standard.
In 2 recent publications (1,2), the National Institute of Standards and Technology (NIST) reported on a change of −4% in its radioactivity standard for the positron emitter 18F. In addition to the obvious change that this brings in directly measuring this radionuclide in activity calibrators, it has an influence on the relative response ratio that allows 68Ge-based solid mock syringe sources to be used as long-lived calibration artifacts for 18F in those instruments.
In 2010, we published a procedure for calibrating a specific commercially available mock syringe source for 68Ge content (3). At the same time, we provided data on the relative response between that source and the equivalent activity of 18F when measured in several models of activity calibrators. Although it would have been possible to merely take the magnitude of the published change in the 18F standard and recalculate a new relative response ratio, a rigorous approach involving new direct measurements of sources containing calibrated 68Ge and 18F solutions gives more confidence in the results. Direct measurement of the ratios has 2 distinct advantages: direct traceability back to the NIST primary standards for the 2 radionuclides is maintained, and a proper uncertainty assessment for the new ratios is possible. We have undertaken a new series of measurements to experimentally determine the new 18F-to-68Ge relative response ratios for the 6-mL (filling volume nominally 3 mL) epoxy-based 68Ge mock syringe source to reflect the change in the NIST 18F standard.
MATERIALS AND METHODS
Source Preparation
The procedure for preparing and measuring the sources used in this study was similar to that used in the experiments reported by Zimmerman and Cessna (3). Because of the length of time that had passed since those experiments were performed, the activity concentrations of the original solutions were too low and it was necessary to prepare a new set of 68Ge sources from a new standardized solution.
All gravimetric transfers described in this paper were performed using polyethylene aspiration–type pycnometers. The amount of solution transferred during the preparation of any source was determined by mass difference using a calibrated microbalance.
A stock solution containing 150 MBq of 68GeCl4 in 2 mL of 0.5 mol·L−1 HCl (International Isotopes Idaho, Inc.) was diluted to a total volume of about 18 mL with a carrier solution containing nominally 45 μg each of nonradioactive Ge+4 and Ga+3 per gram of solution in 0.5 mol·L−1 HCl. Five grams of the solution were then gravimetrically transferred to each of 3 NIST-standard 5-mL glass ampoules.
The solution from one of the ampoules was then used to make a second, diluted solution that was used to make up the 68Ge mock syringe solution masters. The entire 5 g of 68Ge solution were gravimetrically transferred to a 22-mL glass liquid scintillation vial containing 10 mL of a freshly prepared carrier solution containing nominally 65 μg each of nonradioactive Ge+4 and Ga+3 per gram of solution in 0.5 mol·L−1 HCl. The difference in ion concentrations between the carrier solutions used in the first and second dilutions is a result of variability in the preparation procedure and is not expected to affect the stability of the 68Ge solutions.
After agitating to ensure complete mixing, 3 g of the solution were gravimetrically transferred to each of 3 empty mock syringe cylinders (RadQual, LLC). Between dispensing the first and second mock syringes, 5 mL of the solution was transferred into a NIST standard 5-mL glass ampoule to calibrate the activity concentration of the solution. The ampoule was measured in NIST “4π”γ ionization chamber “A” (IC “A”) 40 times, in 4 groups of 10 measurements, alternating with 5 groups of 10 measurements of 226Ra reference source RRS1000. The total activity of the 68Ge solution was determined using the calibration coefficient (K value) for IC “A” that was measured during the original 68Ge standardization in 2008 (4). The 68Ge activity concentration of the solution was calculated by dividing the total 68Ge activity in the ampoule by the transferred solution mass. The resulting activity concentration value was used along with the solution masses contained in each mock syringe to calculate the total 68Ge activity in each syringe.
The 18F mock syringes were prepared in the same manner as for 68Ge, except without the initial dilution step. Approximately 3 mL of 18F-FDG with a nominal activity of 560 MBq was obtained from a commercial radiopharmacy (IBA Molecular) and transferred to a 22-mL glass liquid scintillation vial containing 10 mL of distilled water. After agitation, 3 g of the solution were gravimetrically transferred to each of 3 empty mock syringe cylinders. Between dispensing the first and second mock syringes, 5 mL of the solution were transferred into a NIST standard 5-mL glass ampoule so that the activity concentration of the solution could be determined.
The ampoule was measured in the IC “A” using the same procedure as for the 68Ge ampoule, except that the total activity of the 18F solution was determined using the K value for IC “A” that was determined in 2013 and which is based on the revised 18F standard (1). The 18F activity concentration in each ampoule was calculated by dividing the total activity in the ampoule by the transferred solution mass.
Activity Calibrator Measurements
All the sources described above were measured in the activity calibrators maintained at NIST: a NIST-maintained Vinten 671 ionization chamber (hereafter denoted VIC), Capintec CRC-12, Capintec CRC-15R, Capintec CRC-25PET, Capintec 1.8-ATM (the 1.8-atm chamber from Capintec is the basis of their CRC-PC dose calibrator), and AtomLab AL-500. The primary purpose for measuring the ampoules was to provide a consistency check for both the IC “A” measurements and the activity calibrator setting data that were reported previously (1,2). For each VIC measurement, the ampoule was placed in the ampoule holder, and the current was directly read from a Keithley 6517 electrometer using a LabVIEW interface; 100 readings were acquired at 2-s intervals. The currents were corrected for background and decay corrected to the reference time before being averaged. The sources were remeasured 10 d later using the same counting protocol.
For the Capintec and AtomLab chambers, the calibration curve method (5) was used with the ampoule in the bottom of the appropriate dipper from each chamber; the mock syringes were measured hanging in the syringe holder position of the dipper. For the VIC, the mock syringes were measured using the vial holder and the ampoules were measured using the ampoule holder.
Monte Carlo Calculations
To investigate possible measurement effects due to differences in composition and density between the epoxy and the 68Ge solution, as well as to provide a check on the magnitude of the 18F-to-68Ge relative response ratio, simulations of the VIC with the liquid- and 68Ge epoxy–filled mock syringes were performed using the PENCYL user code from the 2011 version of the PENELOPE (6) Monte Carlo package. The dimensions and materials of the VIC were obtained from the manufacturer’s technical drawings, and a CT image of the chamber was acquired on a Philips Gemini TF 16 PET/CT scanner. Cross-section data for all of the materials, with the exception of the pressurized nitrogen in the VIC, came directly from each code package’s materials database. The cross sections for the pressurized nitrogen were calculated using the PENELOPE PENDBASE materials program, assuming molecular nitrogen at standard temperature and a pressure of 1.01 MPa (10 atm). Similarly, the cross sections for the epoxy were calculated using the composition provided by the manufacturer (John Gilbert, Emerson & Cuming, oral communication, 2008) and PENDBASE. The positron spectra for 68Ga and 18F were calculated using SPEBETA (7) and the output reformatted so as to be compatible with the PENCYL input data file format.
Each simulation was run for a minimum of 5 × 106 primary (positron or photon) events. The cutoff energies for photons and electrons (positrons) were 1 keV and 100 eV, respectively.
All nuclear and atomic data used as input for the simulations, as well as for the analysis of experimental data, were taken from the Decay Data Evaluation Project evaluations (8–10).
RESULTS
Unless otherwise specified, uncertainties are given as combined standard (k = 1) uncertainties and are calculated as the quadratic sum of individual uncertainty components. Standard uncertainties on numerical values are expressed in parentheses as the uncertainty on the final digits.
The IC “A” measurements of the 68Ge activity concentration for the solution used to make up the mock syringes gave a value of 2.824(13) × 106 Bq·g−1 at the measurement reference time. The uncertainty is based on components due to: measurement repeatability (0.01% for 40 measurements), the IC “A” calibration coefficient (0.31%, due mostly to the uncertainty on the NIST 68Ge standard), 226Ra reference source ratios (0.05%), positioning of the 226Ra reference source (0.10%), decay correction (0.01%), and a sample holder position correction (0.31%). Likewise, the activity concentration of the 18F solution used to make up the mock syringes was found to be 7.948(29) × 106 Bq·g−1 at the measurement reference time, with the following uncertainty components: measurement repeatability (0.01% for 40 measurements), the IC “A” calibration figure (0.35%, due mostly to the uncertainty on the NIST 18F standard), positioning of the 226Ra reference source (0.10%), and decay correction (4 × 10-4%).
In addition to deriving the calibration figures for the VIC (KVIC,x,y, where x is either F18 or Ge68 for the 2 radionuclides, and y is either amp or MS for the ampoule or mock syringe geometries, respectively) for the 18F and 68Ge ampoules and the 68Ge mock syringe to confirm our previously published data (2,3), a new KVIC,F18,MS was also determined for the 18F mock syringe to reflect the change in the NIST 18F activity standard. A comparison of the newly measured KVIC,x,y values with those previously published is given in Table 1.
On the basis of the measured activity concentration of the 68Ge solution and the contained masses, the average KVIC,Ge68,amp from the 4 68Ge ampoules was 10.04(5) pA·MBq−1, where the uncertainty was calculated from the following components: measurement repeatability (average SD of the mean of 0.04% for 100 measurements), the uncertainty on the 68Ge activity (0.45%, including decay correction to VIC measurement time), between-source reproducibility (0.07% for 4 sources), and background (1.1 × 10-3%). The KVIC,Ge68,amp value determined in 2008 was 10.121(34) pA·MBq−1 but was measured in the VIC vial holder instead of the ampoule holder used in the present experiments. We chose to measure the KVIC,Ge68,amp in the ampoule holder in the present study because the ampoule holder provides a much more reproducible geometry than the vial holder. The previous choice of geometry was a matter of convenience because it allowed both the ampoules and the mock syringes to be measured sequentially without switching holders, albeit conceding a small additional uncertainty in the ampoule measurement due to source positioning. A separate set of measurements performed in the present study revealed the magnitude of the effect of changing holders to be 0.79%. Using this value to make a correction to the previously measured KVIC,Ge68,amp, we obtained a calibration coefficient of 10.041(34) pA·MBq−1, which is in agreement with the result from the present study.
For the 68Ge mock syringes, the average KVIC,Ge68,MS value was measured to be 10.00(5) pA·MBq−1, where the uncertainty was calculated from components due to: measurement repeatability (average SD of the mean of 0.02% for 100 measurements), the uncertainty on the 68Ge activity (0.45%, including decay correction to VIC measurement time), between-source reproducibility (0.10% for 4 sources), and background (5.9 × 10-3%). As seen in Table 1, this new measurement is identical to our previously reported value.
The measured KVIC,F18,amp for the 18F ampoule was measured in this study to be 10.31(4) pA·MBq−1, where the uncertainty was a standard uncertainty calculated from the quadratic combination of components due to: measurement repeatability (average of 0.02% for 100 measurements), the uncertainty on the 18F activity (0.36%), 18F half-life and decay correction (1.1 × 10-3%), within-measurement repeatability (SD of the mean of 0.03% for 100 measurements), and background (9 × 10-4%). This value is in agreement with the value reported by Fitzgerald et al. (1) to within their respective standard uncertainties.
For the mock syringe, KVIC,F18,MS was calculated from the present data to be 10.24(4) pA·MBq−1, where the uncertainty components were calculated the same way as for the ampoule, with nearly identical magnitudes for each of the different components. This new KVIC,F18,MS is 5.3% higher than the (unpublished) value determined during the experiments performed by Zimmerman and Cessna (3). This observation is primarily a result of the −4.0% change in the NIST 18F standard and improvements in the measurement protocols for the VIC since those original measurements were made.
Taking the previous KVIC,F18,MS value of 10.72(9) pA·MBq−1 that was determined during the experiments conducted in the 2010 study (3), but not included in the publication, and correcting for the change in the NIST standard, we obtained a value of 10.29(9) pA·MBq−1, which is in agreement with the present determination. The ratios between the old and new KVIC,F18 values for both the ampoule and the mock syringe geometries are also in accord within experimental uncertainties, providing additional internal consistency.
The dial settings for the 18F and 68Ge mock syringes determined in our activity calibrators are given in Table 2. The uncertainties on the dial settings are dominated by the uncertainty in the activity calibration for the radionuclide (0.36% and 0.45% for 18F and 68Ge, respectively), although much smaller components due to decay correction, the goodness of fit of the relationship between dial setting and apparent activity, and measurement repeatability were also considered. When compared with the data for the 68Ge mock syringe presented in Table 3 of Zimmerman and Cessna (3), we found that the dial settings for the CRC-12 and CRC-35R were in agreement between the 2 studies. The data reported for the CRC-15R are in agreement within their respective expanded (k = 2) uncertainties.
The AtomLab 100, which was used in the previous studies of Zimmerman and Cessna (3), has been retired from use in our laboratory. The AL-500 chamber used in this work is relatively new for our laboratory, and we have no previous data against which to compare.
In the VIC, these new results give a 18F-to-68Ge response ratio of 1.024(6) for the mock syringe geometry. For the Capintec and AtomLab chambers, the 18F-to-68Ge relative response ratios were measured in the same way as done by Zimmerman and Cessna (3)—that is, by measuring the 18F mock syringes at the appropriate dial setting that had been determined for the 68Ge mock syringes (Table 2). The results of those measurements are also given in Table 3 and supplant the data presented in Table 4 of Zimmerman and Cessna (3).
DISCUSSION
The data presented in this study revise our results previously published in 2010 (3). This revision was prompted by the recent change in the NIST standard for 18F (1). To ensure consistency with previous data presented for 68Ge mock syringes, we have repeated the measurements for the determination of dial settings and calibration coefficients for these sources and checked against previously determined dial settings for the NIST ampoule. In all cases, agreement between the 2010 data and the present experiments was achieved within the experimental uncertainties.
For 18F, the same approach was used with the NIST ampoules to give confidence in the correctness of the 18F mock syringe dial setting determinations. In 2010, using the previous NIST 18F standard, we measured the relative response ratio of the mock syringes for the Capintec chambers to be 1.054(10). If the 2010 18F activities are adjusted by −4.0%, we obtain an average ratio of 1.012(10), which is in accord with the ratios recovered for all of the activity calibrators studied. This means that a mock syringe containing a certain amount of 18F would give an activity value (at the 68Ge setting) that now reads 1.2% high, as opposed to the previous value of 5.4%.
The measured relative response ratio of 1.024(6) obtained for the VIC for the mock syringes in this study is in agreement with the previously measured (and adjusted for the change in the 18F standard) value of 1.029(10) for that chamber. With our improved model of the chamber and a larger number of events, the PENELOPE 2011 calculations predict a ratio of 1.017(8), with the uncertainty estimated from the uncertainty on the dose deposited in the ionization chamber gas as reported by the simulation and the uncertainties in the positron branching ratios and emission rate for the 1077 keV γ-ray. Within the respective uncertainties, agreement can be seen between the Monte Carlo prediction for the relative response, the previous (adjusted) measured value, and the ratio obtained from the present study, thereby providing a degree of confidence in the newly measured result.
The Monte Carlo results also provided verification that no attenuation correction is required when relating the activity in the liquid-filled mock syringe sources to the activity in the solid epoxy mock syringes. The ratio of calculated doses deposited in the ionization chamber gas for the solid mock syringe to the solution-filled mock syringe was 0.995(4) and is consistent with the results given in our previous work (3).
Information regarding the change in the NIST 18F standards and the updated relative response ratios for the mock syringes in the Capintec and AtomLab chambers has already been transmitted to the calibration source and instrument manufacturers.
For most clinical applications, the change in the NIST 18F standard will have a relatively minor impact on routine practice. The NIST 68Ge standard has not changed, thus the calibrated value for 68Ge activity in a traceable source is still valid. If the solid epoxy 68Ge mock syringe source is used to calibrate an activity calibrator for 68Ge, the user should expect to obtain the same calibration factors (dial settings) as they always have, but now the equivalent activity of 18F is only 1.015–1.02 times the 68Ge activity, instead of 1.054. Thus, the 18F calibrated activity will shift to a lower value and should give dial settings close to those given in Table 2. If a clinical site obtains its 18F from a commercial radiopharmacy that has not yet changed its calibration to match the new NIST standard, a difference between the measured value and activity on the label could be evident and should be investigated.
Because most calibrations for PET scanners are made relative to the activity reported by an activity calibrator, scanners should be recalibrated once the updated 18F equivalent activity value is adopted in the radiopharmacy. Otherwise, a systematic 4% difference between the scanner-reported activities and the true injected activity will occur. This will also affect the calculation of standardized uptake values by 4%, since the relationship between the activity of 18F used to calibrate the scanner and the calibration of new patient dosages is no longer the same.
CONCLUSION
In this work, we have performed a series of experiments to redetermine the relative response ratio between 68Ge and 18F for a commercially available solid 68Ge mock syringe source. This −3.7% change was prompted by a recent revision of the NIST activity standard for 18F. These new results allow the calibration of activity calibrators to once again be made using that source in a way that can be related back to NIST standards of activity.
DISCLOSURE
The costs of publication of this article were defrayed in part by the payment of page charges. Therefore, and solely to indicate this fact, this article is hereby marked “advertisement” in accordance with 18 USC section 1734.
Certain commercial equipment, instruments, or materials are identified in this paper to foster understanding. Such identification does not imply recommendation by the National Institute of Standards and Technology, nor does it imply that the materials or equipment identified are necessarily the best available for the purpose.
No other potential conflict of interest relevant to this article was reported.
Acknowledgments
We thank Keith Allberg (RadQual, LLC) for providing the 68Ge solution and empty mock syringes used in this work.
Footnotes
Published online Jul. 16, 2015.
- © 2015 by the Society of Nuclear Medicine and Molecular Imaging, Inc.
REFERENCES
- Received for publication April 15, 2015.
- Accepted for publication July 7, 2015.