Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • View or Listen to JNM Podcast
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
Research ArticleBasic Science Investigation
Open Access

Evaluation of Data-Driven Rigid Motion Correction in Clinical Brain PET Imaging

Matthew G. Spangler-Bickell, Samuel A. Hurley, Ali Pirasteh, Scott B. Perlman, Timothy Deller and Alan B. McMillan
Journal of Nuclear Medicine October 2022, 63 (10) 1604-1610; DOI: https://doi.org/10.2967/jnumed.121.263309
Matthew G. Spangler-Bickell
1PET/MR Engineering, GE Healthcare, Waukesha, Wisconsin;
2Radiology, University of Wisconsin–Madison, Madison, Wisconsin; and
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Samuel A. Hurley
2Radiology, University of Wisconsin–Madison, Madison, Wisconsin; and
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Ali Pirasteh
2Radiology, University of Wisconsin–Madison, Madison, Wisconsin; and
3Medical Physics, University of Wisconsin–Madison, Madison, Wisconsin
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Scott B. Perlman
2Radiology, University of Wisconsin–Madison, Madison, Wisconsin; and
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Timothy Deller
1PET/MR Engineering, GE Healthcare, Waukesha, Wisconsin;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Alan B. McMillan
2Radiology, University of Wisconsin–Madison, Madison, Wisconsin; and
3Medical Physics, University of Wisconsin–Madison, Madison, Wisconsin
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
  • PDF
Loading

Visual Abstract

Figure
  • Download figure
  • Open in new tab
  • Download powerpoint

Abstract

Head motion during brain PET imaging can significantly degrade the quality of the reconstructed image, leading to reduced diagnostic value and inaccurate quantitation. A fully data-driven motion correction approach was recently demonstrated to produce highly accurate motion estimates (<1 mm) with high temporal resolution (≥1 Hz), which can then be used for a motion-corrected reconstruction. This can be applied retrospectively with no impact on the clinical image acquisition protocol. We present a reader-based evaluation and an atlas-based quantitative analysis of this motion correction approach within a clinical cohort. Methods: Clinical patient data were collected over 2019–2020 and processed retrospectively. Motion was estimated using image-based registration on reconstructions of ultrashort frames (0.6–1.8 s), after which list-mode reconstructions that were fully motion-corrected were performed. Two readers graded the motion-corrected and uncorrected reconstructions. An atlas-based quantitative analysis was performed. Paired Wilcoxon tests were used to test for significant differences in reader scores and SUVs between reconstructions. The Levene test was used to determine whether motion correction had a greater impact on quantitation in the presence of motion than when motion was low. Results: Fifty standard clinical 18F-FDG brain PET datasets (age range, 13–83 y; mean ± SD, 59 ± 20 y; 27 women) from 3 scanners were collected. The reader study showed a significantly different, diagnostically relevant improvement by motion correction when motion was present (P = 0.02) and no impact in low-motion cases. Eight percent of all datasets improved from diagnostically unacceptable to acceptable. The atlas-based analysis demonstrated a significant difference between the motion-corrected and uncorrected reconstructions in cases of high motion for 7 of 8 regions of interest (P < 0.05). Conclusion: The proposed approach to data-driven motion estimation and correction demonstrated a clinically significant impact on brain PET image reconstruction.

  • PET
  • image reconstruction
  • data-driven motion correction
  • brain imaging

As the spatial resolution of modern whole-body PET scanners reaches 2–4 mm in full width at half maximum, together with improved sensitivity and time-of-flight resolution, it is becoming increasingly likely that even slight head motion may substantially degrade the reconstructed image. Although patient motion with translations of up to 15 mm and rotations of up to 4 degrees have been reported (1,2), less motion is quite common. Various motion-tracking and correction techniques have been presented for head motion (2–7). These usually use an external tracking device (such as a camera) to track a marker attached to the head (5), or they directly track the head (6). The motion estimates can then be used to perform frame-based reconstructions (8) or a full event-by-event motion-corrected (MoCo) reconstruction (9,10). However, none of these motion correction approaches have been implemented into widespread standard clinical routine, for several reasons. Some patient motion can be partially managed through head restraints and by discarding motion-corrupted portions of the data. To date, most motion-tracking methods rely on external hardware around the scanner (such as cameras) or attached to the patient (such as head markers), which complicate routine clinical protocols. Until recently, there has not been a substantial effort from vendors to incorporate motion correction into their products; efforts have thus remained predominantly within the research setting.

Fully data-driven approaches to motion correction that do not require external hardware have been presented. These usually estimate when motion occurred so that the data can be suitably framed (11), or they may estimate the motion itself to be used in a MoCo reconstruction (12–15). Because of the typically low count rates in PET imaging and long reconstruction times, the temporal resolution used for such motion estimation is usually on the order of tens of seconds or longer. Such low temporal resolutions may lead to residual intraframe motion blurring and inaccurate motion estimates. Alternatively, when higher temporal resolutions are used (on the order of ∼1 s), as described previously (12,13), the motion is estimated using centroid-of-distribution or inertial tensor calculations.

In this work, we evaluated a recently proposed approach to data-driven motion estimation and correction (16,17). The motion was estimated using rigid image registration on reconstructed images of very short frames. The estimated motion was then used in a full event-by-event MoCo list-mode reconstruction of the data, including all PET corrections. The approach is completely data-driven and can be applied retrospectively. An evaluation on 50 standard clinical 18F-FDG brain PET datasets is presented, showing the results of a reader study and an atlas-based quantitation analysis.

MATERIALS AND METHODS

Data

Patient data were acquired at the Wisconsin Institutes for Medical Research at the University of Wisconsin over 2019 and 2020, from a 4-ring Discovery MI PET/CT device (n = 11; 20-cm axial field of view), a Discovery 710 PET/CT device (n = 18), and a SIGNA PET/MRI device (n = 21) (all from GE Healthcare). Fifty consecutive 18F-FDG brain PET datasets were collected retrospectively, and none were rejected. All datasets were routine clinical imaging studies, and the need to obtain consent was waived by the institutional review board. Preliminary results using these datasets were presented at the 2021 annual meeting of the Society of Nuclear Medicine and Molecular Imaging (18); the current work presents a more thorough analytic and statistical evaluation.

Motion Estimation and Image Reconstruction

The data were processed in 2 steps before being analyzed. Figure 1 shows a flowchart of the study. First, ultrafast reconstructions of very short frames (16) (0.6–1.8 s, set automatically and adjusted at each frame to ensure a constant 500 × 103 true and scattered events per frame (17)) were performed for the entire scan duration, and image-based registration was then performed on these frames to estimate the motion. Rigid registration was performed using a least-squares metric and a gradient descent optimizer; further details were given in a previous publication (17). The 6 degrees of freedom of the motion were thus estimated directly at about 1 Hz, with an accuracy of less than 1 mm (measured as the mean error in absolute displacement of a mesh of points moved by the estimated motion) (17). The reference frame for the image registrations was chosen to ensure that the PET reconstruction aligned with the attenuation map. For PET/MRI, the MRI acquisitions for the attenuation map occur concurrently with the PET acquisitions; thus, the PET frames corresponding to data acquired during the MRI attenuation correction pulse sequence were averaged to create the reference frame. For PET/CT, however, the CT acquisitions occur before the PET acquisitions; thus, after estimating motion using the first PET frame as the reference, a single automatic cross-modality registration is performed using a mutual-information metric to set this reference frame to align with the CT image. The mean of all the short PET frames, after being aligned according to the estimated motion, was used for this registration. For 12 of the 29 PET/CT datasets, the automated cross-modality registration between the PET and the CT scans was suboptimal and manual intervention was required to ensure good registration. We will make this registration step more robust in the future to ensure fully automated processing.

FIGURE 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1.

Flowchart of reconstruction process and analysis. 4D = 4-dimensional; MNI = Montreal Neurological Institute; recons = reconstructions.

The data were categorized into 4 motion groups using a metric based on the magnitude of the motion for each dataset. Similarly to previous investigators (19,20), we categorized the motion by moving points in image space and measuring their displacement. Although others have used an average displacement, we were interested in the maximum extent of the motion, and thus 2 points at the extreme extent of the brain were sufficient. Two points located in image space at 70 mm anterior and 70 mm posterior to the brain center were chosen and moved according to the estimated motion parameters. The brain center relative to the scanner center was set at a typically observed value for each scanner: for the PET/MRI device at 0, 40, and 20 mm for left–right, anterior–posterior, and superior–inferior, respectively (the anterior–posterior offset was due to the head coil used), and for both PET/CT devices at 0, 0, and 20 mm, respectively. The median absolute displacement from the reference was calculated for each point. The larger of these 2 medians was used as a metric to classify the datasets into 4 motion groups: low (median displacement < 1 mm), offset (median displacement < 1 mm but initial displacement > 2 mm), medium (median displacement = 1–2 mm), and high (median displacement > 2 mm).

The offset-motion group captures those datasets with little motion during the PET acquisition but with a large offset between the attenuation map acquisition and the PET acquisition (this usually applies only to PET/CT scans). This group classification was chosen empirically on the basis of our experience with many clinical datasets.

After estimation of head motion, a full reconstruction was performed, with each event being corrected according to the estimated motion, as shown in Figure 1. List-mode time-of-flight–based block sequential regularization expectation maximization (21,22) was performed with a β-parameter of 50. Spatially variant point-spread function modeling was performed using a hybrid image-space/projection-space approach (23). For clarity, the list-mode maximum likelihood expectation maximization with motion correction (9,10) is given here; subsets and a regularization term are added for the block sequential regularization expectation maximization: Embedded Image Embedded Image where Embedded Image is the image value at pixel j and iteration n, Embedded Image is the line of response (LOR, the line joining a detecting crystal pair) i associated with list-mode event m, M is the total number of list-mode events, Embedded Image is the MoCo LOR i for event m, Embedded Image is the system matrix, Embedded Image is the motion-aware scatter contribution along LOR i, Embedded Image is the randoms contribution, Embedded Image is the attenuation correction factor through the patient attenuation map along LOR i, Embedded Image is the attenuation correction factor through the attenuating material exterior to the patient along LOR i, and Embedded Image is the scanner sensitivity factor (crystal efficiency and dead time) for LOR i. The time-averaged sensitivity image Embedded Image is calculated by moving the endpoints of each LOR (l, of which there are L in the scanner) by a particular set of motion parameters p, backprojecting the appropriate attenuation and sensitivity factors, and calculating the time-weighted (Embedded Image) average across all the motion data, P. The attenuation factors Embedded Image and Embedded Image are handled separately since the patient is moving whereas the rest of the attenuating material is not; therefore, the MoCo LORs (Embedded Image) are used for the patient attenuation correction factors.

Additionally, a non-MoCo list-mode reconstruction was performed for comparison.

Reader Study

The MoCo and non-MoCo reconstructions were randomized and read by a masked nuclear medicine physician with 36 y of experience and a dual–board-certified nuclear medicine physician and radiologist with fellowship training in nuclear medicine and body MRI with 2 y of experience. Images were evaluated on a 5-point Likert scale for sharpness and diagnostic quality, as specified in Table 1. The readers were not aware of the motion groups of the datasets; these were used only during the analysis.

View this table:
  • View inline
  • View popup
TABLE 1.

Likert-Scale Scoring for Image Evaluation

Quantitative Analysis

An atlas-based analysis of the quantitative accuracy of the reconstructions was performed. The reconstructions were individually and nonrigidly registered to an aggregated 18F-FDG atlas in Montreal Neurosciences Institute–152 image space (24,25) using the Advanced Normalization Tools toolbox (26). Eight regions of interest (ROIs) were extracted and analyzed: frontal lobe, occipital lobe, temporal lobe, parietal lobe, cerebellum, left/right cerebral cortex, and whole cerebral cortex.

Statistical Analysis

For the reader study, interreader variability was tested using 2 metrics: the modified interrater agreement index (Embedded Image) was used to evaluate the 5-point Likert scores (27), and the Cohen κ (28) was used to evaluate agreement on whether an image was diagnostically acceptable (i.e., having a Likert score of ≥3). These tests were conducted for both reader questions and for the non-MoCo and MoCo reconstructions.

Using the average Likert scores for the 2 readers, paired Wilcoxon tests were conducted to test for significant differences between the median scores of the MoCo and non-MoCo reconstructions at a 2-sided significance level of P < 0.05, using the statistical toolbox in MATLAB (MathWorks). Correction was made for the false-discovery rate using the Benjamini–Hochberg procedure with a rate (Q-value) of 10%.

For the atlas-based quantitative analysis, paired Wilcoxon tests for significant differences between the medians of ROI SUVmax in the MoCo and non-MoCo reconstructions within each motion group were conducted at a 2-sided significance level of P < 0.05, with false-discovery-rate correction. Additionally, the relative differences between the SUVmax of the MoCo and non-MoCo reconstructions were calculated, and Levene tests (29) were conducted to determine whether the variance in these differences for each motion group was significantly different from the low-motion group, indicating that motion correction had made a quantitative difference in reconstructions when motion was present. A 2-sided significance level of P < 0.05 was used, and false-discovery-rate correction was applied.

RESULTS

Patient Characteristics and Injected Activity

Patient age ranged from 13–83 y (mean ± SD, 59 ± 20 y). There were 27 women. The injected activity was 390 ± 24 MBq for the Discovery 710 (55 ± 4 min uptake time, 15-min duration, 4.6 ± 1.1 × 108 counts), 400 ± 26 MBq for the Discovery MI (56 ± 5 min uptake time, 15-min duration, 8.9 ± 2.3 × 108 counts), and 450 ± 85 MBq for the SIGNA (63 ± 13 min uptake time, 25-min duration [except for 3 datasets with a 11- to 15-min duration], 17.2 ± 5.3 × 108 counts).

Motion Estimation

The number of datasets in each motion group is shown in Table 2, and the distribution of the displacement metric is shown in Figure 2.

View this table:
  • View inline
  • View popup
TABLE 2.

Distribution of Datasets Among Defined Motion Groups

FIGURE 2.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 2.

Violin plots (30) showing distribution of estimated motion for all datasets within 4 motion groups. Width of violins indicates density of data points, and length indicates range of data. Actual data points are scattered within violins, with white dot being median.

Motion Correction Case Studies

Figure 3 shows examples of reconstructions from the low-, offset-, and high-motion groups. The relative-difference images were calculated as (MoCo − non-MoCo)/MoCo × 100%.

FIGURE 3.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 3.

Example reconstructions for 3 case studies from low-motion (A), offset-motion (B), and high-motion (C) groups. All 3 are PET/CT data. Six degrees of freedom of motion data are plotted at top. Smoothed relative differences between images are shown at bottom overlayed on MoCo image. Low-motion case demonstrates that when there is little motion, motion correction has very small effect on reconstruction. Offset case shows that although no obvious differences are visible between images, relative-difference image shows gradient due to misalignment of Non-MoCo image with attenuation map. In high-motion case, much of blurring due to motion visible in Non-MoCo image has been removed in MoCo image. Rot. = rotation; Trans. = translation.

Reader Study

The results of the interreader variability analysis are shown in Table 3. Agreement between the readers was high according to all tests. Notably, agreement on whether an image was diagnostically acceptable (κ) was higher (no disagreement) for the MoCo reconstructions than for the non-MoCo reconstructions (for which there was 1 disagreement). The Likert scores for the 2 readers are shown in Figure 4 for the 2 questions. In 5 (10%) of the 50 datasets, the diagnostic quality of the reconstruction with motion correction improved by at least 1 on the Likert scale. The non-MoCo reconstructions for 4 (8%) datasets were rated as not diagnostically acceptable, and for all of these the MoCo reconstructions were rated as diagnostically acceptable. The results of the paired Wilcoxon tests on the reader scores are shown in Table 4.

View this table:
  • View inline
  • View popup
TABLE 3.

Results of Interreader Variability Analysis (High Agreement in All Cases)

FIGURE 4.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 4.

Likert scores for MoCo and non-MoCo reconstructions for 2 questions: image sharpness (A) and diagnostic quality (B). In non-MoCo cases, reader scores had more variation among datasets in higher-motion groups, with some images not being diagnostically acceptable. In MoCo cases, scores were consistent across all motion groups.

View this table:
  • View inline
  • View popup
TABLE 4.

P Values of Paired Wilcoxon Tests Between MoCo and Non-MoCo Reconstructions, According to Reader Likert Scores

Quantitative Analysis

The relative differences in ROI SUVmax between the MoCo and non-MoCo reconstructions are shown in Figure 5, calculated as (MoCo SUVmax − non-MoCo SUVmax)/MoCo SUVmax × 100%. Table 5 presents the results of the statistical analysis.

FIGURE 5.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 5.

Relative differences between SUVmax of ROIs extracted from MoCo and non-MoCo reconstructions. Differences are larger in higher-motion groups, as expected. Since MoCo reconstruction ensures better alignment with attenuation map, SUVmax is expected to be more accurate in MoCo reconstructions than in non-MoCo reconstructions, regardless of which is greater or lesser.

View this table:
  • View inline
  • View popup
TABLE 5.

Results from Paired Wilcoxon Tests on ROI SUVmax of MoCo and Non-MoCo Images and from Levene Tests on Variance of ROI SUVmax Relative-Difference Values

Figure 5 shows that the relative differences in SUVmax between the MoCo and non-MoCo reconstructions were larger in the higher-motion groups than in the low-motion group. In the high-motion group, the SUVmax of the parietal lobe in the MoCo reconstructions differed from that in the non-MoCo reconstructions by 1.5% ± 2.7%, with a maximum of 6.6%, and in the temporal lobe the SUVmax differed by 1.8% ± 2.6%, with a maximum of 8.2%. In all cases, the SUVmax in the MoCo reconstruction is assumed to be more accurate, whether it was higher or lower than the non-MoCo reconstruction, since the former ensures better alignment with the attenuation map and has reduced motion blurring. Table 5 demonstrates that in the high-motion group the SUVmax significantly differed between the MoCo and non-MoCo reconstructions in 7 of 8 ROIs. Even in the offset-motion group, in which there was minimal motion during the PET acquisition, motion correction made a significant difference in 5 of 8 ROI SUVmax results because of the improved alignment with the attenuation map. The Levene tests indicated that the variance in relative differences between the MoCo and non-MoCo reconstructions was higher for 7 of the 8 ROIs in the medium- and high-motion groups than in the low-motion group (Table 5). These variances can be seen in the extent of the plots in Figure 5, comparing the higher-motion groups with the low-motion group. Table 5 indicates that when there was high motion, the motion correction significantly changed the reconstruction. The results of the reader study then confirm that the MoCo reconstructions were preferred.

DISCUSSION

An evaluation of a fully data-driven motion estimation and correction technique for reconstruction of brain PET datasets has been presented. Fifty standard clinical 18F-FDG brain PET datasets were processed retrospectively, acquired on 1 PET/MRI and 2 PET/CT scanners. No additional motion-tracking hardware was used during the scan, and there was no impact on the standard clinical routine. The motion estimation used a temporal resolution of about 1 s and detected motion of more than 1 mm in 70% (35/50) of cases and more than 2 mm in 24% (12/50) of cases, the latter of which usually resulted in visually obvious differences between the MoCo and non-MoCo reconstructions. The masked reader study showed that the MoCo reconstructions improved the diagnostic quality in 10% (5/50) of the datasets, and the improvement was significant in the high-motion group (P = 0.02) and when considering all the data (P = 0.003). In 8% (4/50) of the datasets, which is a substantial portion of the cohort, the image was improved from diagnostically unacceptable to acceptable. The atlas-based quantitative analysis found significant differences (P < 0.05) in the SUVmax in 7 of the 8 ROIs in the medium- and high-motion groups and no significant differences in 5 of 8 ROIs in the low-motion group. The reader study confirmed that the image quality of the MoCo reconstructions was preferred over the non-MoCo reconstructions when motion was present and did not affect the scores when no motion was present, while the atlas-based analysis confirmed that motion correction does affect the quantitation of the reconstructions in the presence of motion.

Our study incorporated data from 3 scanners with very different geometries, with the axial field of view ranging from 157 to 250 mm, and all scanners benefitted from motion correction. The higher sensitivity and time-of-flight resolution of modern scanners allows for use of shorter frame durations for motion estimation and, hence, improved temporal sampling. To optimize temporal sampling, scanner-specific optimization may be necessary (17). Motion estimation and reconstruction were performed in a research setting and took approximately 2 h, which was about 30% longer than the non-MoCo reconstruction of the same dataset. Significantly faster reconstruction is expected with software optimization and dedicated hardware (e.g., graphics processing units) to ensure that the approach can be clinically feasible in future work.

This study had some limitations. We focused on 18F-FDG because it is the most common clinically used radiotracer. However, assuming that accurate motion estimates can be obtained with other radiotracers, we expect that motion correction would have a similar effect on reconstructions of such datasets. Accurate motion estimation has been demonstrated previously with 18F-florbetaben using this approach (17). Optical motion tracking was not available for comparison, as the data were processed retrospectively, and such a comparison was not the intention of this work. This work did not include examinations for which the activity distribution of the radiotracer inside the brain may change substantially during the scan, such as an 15O-H2O brain perfusion study, since the reference frame used for registration would not be representative of the entire dataset. Although an approach for motion estimation in such cases would be more challenging and was outside of the scope of the current work, we believe that a data-driven solution is possible; such a solution is the topic of ongoing research. Lastly, whereas image quality before and after motion correction was evaluated, the diagnostic implications of the MoCo images were not fully investigated. Considering the promising nature of our current results, we plan to further investigate the clinical impact of the application.

CONCLUSION

We have presented an evaluation of a data-driven technique for correction of head motion in brain PET imaging. We demonstrated that motion is prevalent among standard clinical datasets and that motion correction has a significant impact on reconstructions, both qualitatively and quantitatively. The application of motion correction was not detrimental to image quality or quantification when no motion was present. Because motion is a known confounder of clinical brain PET, using data-driven motion correction will likely have important implications for diagnostic and research studies in which motion may occur. Given that the proposed solution relies entirely on retrospective reconstruction, it can readily be adopted into routine PET imaging procedures.

DISCLOSURE

Matthew Spangler-Bickell and Timothy Deller are employees of GE Healthcare. Research reported in this publication was supported by the National Institute of Biomedical Imaging and Bioengineering of the National Institutes of Health under award R01EB026708. No other potential conflict of interest relevant to this article was reported.

KEY POINTS

QUESTION: How effective is the proposed data-driven head motion correction technique?

PERTINENT FINDINGS: Motion was observed in 70% of the study cohort, consisting of 50 consecutively acquired datasets. A reader study showed that all datasets that were deemed diagnostically unacceptable without motion correction (8%, 4/50) were then diagnostically acceptable with motion correction, with a significant improvement in cases of high motion. An 8-ROI atlas-based quantitative analysis concluded that motion correction had a significant impact on SUVmax (≤9%) in 65% of the ROIs.

IMPLICATIONS FOR PATIENT CARE: Patient motion would no longer be a concern for 18F-FDG PET brain imaging when using this technology; patients could be scanned regardless of motion risk, and any motion would be corrected.

Footnotes

  • Published online Jan. 27, 2022.

  • © 2022 by the Society of Nuclear Medicine and Molecular Imaging.

Immediate Open Access: Creative Commons Attribution 4.0 International License (CC BY) allows users to share and adapt with attribution, excluding materials credited to previous publications. License: https://creativecommons.org/licenses/by/4.0/. Details: https://jnm.snmjournals.org/page/permissions.

REFERENCES

  1. 1.↵
    1. Bloomfield PM,
    2. Spinks TJ,
    3. Reed J,
    4. et al
    . The design and implementation of a motion correction scheme for neurological PET. Phys Med Biol. 2003;48:959–978.
    OpenUrlCrossRefPubMed
  2. 2.↵
    1. Lopresti BJ,
    2. Russo A,
    3. Jones WF,
    4. et al
    . Implementation and performance of an optical motion tracking system for high resolution brain PET imaging. IEEE Trans Nucl Sci. 1999;46:2059–2067.
    OpenUrl
  3. 3.
    1. Daube-Witherspoon M,
    2. Yan Y,
    3. Green M,
    4. Carson R,
    5. Kempner K,
    6. Herscovitch P
    . Correction for motion distortion in PET by dynamic monitoring of patient position [abstract]. J Nucl Med. 1990;31(suppl):816.
    OpenUrl
  4. 4.
    1. Fulton RR,
    2. Meikle SR,
    3. Eberl S,
    4. Pfeiffer J,
    5. Constable CJ,
    6. Fulham MJ
    . Correction for head movements in positron emission tomography using an optical motion-tracking system. IEEE Trans Nucl Sci. 2002;49:116–123.
    OpenUrlCrossRef
  5. 5.↵
    1. Spangler-Bickell MG,
    2. Khalighi MM,
    3. Hoo C,
    4. et al
    . Rigid motion correction for brain PET/MR imaging using optical tracking. IEEE Trans Radiat Plasma Med Sci. 2019;3:498–503.
    OpenUrl
  6. 6.↵
    1. Kyme AZ,
    2. Aksoy M,
    3. Henry DL,
    4. Bammer R,
    5. Maclaren J
    . Marker-free optical stereo motion tracking for in-bore MRI and PET-MRI application. Med Phys. 2020;47:3321–3331.
    OpenUrl
  7. 7.↵
    1. Kyme AZ,
    2. Fulton RR
    . Motion estimation and correction in SPECT, PET and CT. Phys Med Biol. 2021;66(18).
  8. 8.↵
    1. Picard Y,
    2. Thompson CJ
    . Motion correction of PET images using multiple acquisition frames. IEEE Trans Med Imaging. 1997;16:137–144.
    OpenUrlCrossRefPubMed
  9. 9.↵
    1. Carson RE,
    2. Barker WC,
    3. Liow JS,
    4. Johnson CA
    . Design of a motion-compensation OSEM list-mode algorithm for resolution-recovery reconstruction for the HRRT. IEEE Nucl Sci Symp Conf Rec. 2003;5:3281–3285.
    OpenUrl
  10. 10.↵
    1. Rahmim A,
    2. Bloomfield P,
    3. Houle S,
    4. et al
    . Motion compensation in histogram-mode and list-mode EM reconstructions: beyond the event-driven approach. IEEE Trans Nucl Sci. 2004;51:2588–2596.
    OpenUrlCrossRef
  11. 11.↵
    1. Schleyer PJ,
    2. Dunn JT,
    3. Reeves S,
    4. Brownings S,
    5. Marsden PK,
    6. Thielemans K
    . Detecting and estimating head motion in brain PET acquisitions using raw time-of-flight PET data. Phys Med Biol. 2015;60:6441–6458.
    OpenUrl
  12. 12.↵
    1. Feng T,
    2. Yang D,
    3. Zhu W,
    4. Dong Y,
    5. Li H
    . Real-time data-driven rigid motion detection and correction for brain scan with listmode PET. In: 2016 IEEE Nuclear Science Symposium, Medical Imaging Conference and Room-Temperature Semiconductor Detector Workshop (NSS/MIC/RTSD) . IEEE;2016:1–4.
  13. 13.↵
    1. Rezaei A,
    2. Spangler-Bickell M,
    3. Schramm G,
    4. Van Laere K,
    5. Nuyts J,
    6. Defrise M
    . Rigid motion tracking using moments of inertia in TOF-PET brain studies. Phys Med Biol. 2021;66(18).
  14. 14.
    1. Jin X,
    2. Mulnix T,
    3. Gallezot JD,
    4. Carson RE
    . Evaluation of motion correction methods in human brain PET imaging: a simulation study based on human motion data. Med Phys. 2013;40:102503.
    OpenUrl
  15. 15.↵
    1. Mukherjee JM,
    2. Lindsay C,
    3. Mukherjee A,
    4. et al
    . Improved frame-based estimation of head motion in PET brain imaging. Med Phys. 2016;43:2443–2454.
    OpenUrlCrossRef
  16. 16.↵
    1. Spangler-Bickell MG,
    2. Deller TW,
    3. Bettinardi V,
    4. Jansen F
    . Ultra-fast list-mode reconstruction of short PET frames and example applications. J Nucl Med. 2021;62:287–292.
    OpenUrlAbstract/FREE Full Text
  17. 17.↵
    1. Spangler-Bickell MG,
    2. Hurley SA,
    3. Deller TW,
    4. et al
    . Optimizing the frame duration for data-driven rigid motion estimation in brain PET imaging. Med Phys. 2021;48:3031–3041.
    OpenUrl
  18. 18.↵
    1. Spangler-Bickell M,
    2. Hurley S,
    3. Pirasteh A,
    4. Perlman S,
    5. Deller T,
    6. McMillan A
    . Clinical evaluation of data-driven motion corrected reconstruction for PET brain imaging [abstract]. J Nucl Med. 2021;62(suppl 1):59.
    OpenUrl
  19. 19.↵
    1. Tisdall MD,
    2. Hess AT,
    3. Reuter M,
    4. Meintjes EM,
    5. Fischl B,
    6. Van Der Kouwe AJW
    . Volumetric navigators for prospective motion correction and selective reacquisition in neuroanatomical MRI. Magn Reson Med. 2012;68:389–399.
    OpenUrlCrossRefPubMed
  20. 20.↵
    Measuring transformation error by RMS deviation. Wellcome Centre Integrative Neuroimaging website. http://www.fmrib.ox.ac.uk/analysis/techrep/tr99mj1/tr99mj1/index.html. Published February 11, 2003. Accessed June 23, 2022.
  21. 21.↵
    1. Ahn S,
    2. Fessler JA
    . Globally convergent image reconstruction for emission tomography using relaxed ordered subsets algorithms. IEEE Trans Med Imaging. 2003;22:613–626.
    OpenUrlCrossRefPubMed
  22. 22.↵
    1. Spangler-Bickell MG,
    2. Deller T,
    3. Jansen F
    . Accelerated regularised list-mode PET reconstruction using subset relaxation. In: 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference (NSS/MIC). IEEE; 2019:1–3.
  23. 23.↵
    1. Deller TW,
    2. Ahn S,
    3. Jansen FP,
    4. et al
    .Implementation and image quality benefit of a hybrid-space PET point spread function. Presented at: IEEE Nuclear Science Symposium and Medical Imaging Conference; October 20, 2021; Yokohama, Japan [virtual].
  24. 24.↵
    1. Darkner S
    . FDG-PET template MNI152 1mm. UCPH Research website. https://research.ku.dk/search/result/?pure=en%2Fpublications%2Ffdgpet-template-mni152-1mm(5cf2edbb-e75b-4f90-ac4c-89e6ef1bc3f6).html. Published 2013. Accessed June 23, 2022.
  25. 25.↵
    Harvard-Oxford cortical and subcortical structural atlases. NeuroVault website. https://identifiers.org/neurovault.collection:262. Accessed June 23, 2022.
  26. 26.↵
    Advanced normalization tools. stnava website. http://stnava.github.io/ANTs/. Accessed June 23, 2022.
  27. 27.↵
    1. O’Neill TA
    . An overview of interrater agreement on Likert scales for researchers and practitioners. Front Psychol. 2017;8:777.
    OpenUrl
  28. 28.↵
    1. Cohen J
    . A coefficient of agreement for nominal scales. Educ Psychol Meas. 1960;20:37–46.
    OpenUrlCrossRef
  29. 29.↵
    1. Olkin I.
    1. Leven H.
    Robust tests for equality of variances. In: Olkin I., ed. Contributions to Probability and Statistics: Essays in Honor of Harold Hotelling. Stanford University Press; 1960:279–292.
  30. 30.↵
    Violin plots for Matlab. GitHub website. https://github.com/bastibe/Violinplot-Matlab. Accessed June 23, 2022.
  • Received for publication October 1, 2021.
  • Accepted for publication January 25, 2022.
PreviousNext
Back to top

In this issue

Journal of Nuclear Medicine: 63 (10)
Journal of Nuclear Medicine
Vol. 63, Issue 10
October 1, 2022
  • Table of Contents
  • Table of Contents (PDF)
  • About the Cover
  • Index by author
  • Complete Issue (PDF)
Print
Download PDF
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Evaluation of Data-Driven Rigid Motion Correction in Clinical Brain PET Imaging
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
Evaluation of Data-Driven Rigid Motion Correction in Clinical Brain PET Imaging
Matthew G. Spangler-Bickell, Samuel A. Hurley, Ali Pirasteh, Scott B. Perlman, Timothy Deller, Alan B. McMillan
Journal of Nuclear Medicine Oct 2022, 63 (10) 1604-1610; DOI: 10.2967/jnumed.121.263309

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Evaluation of Data-Driven Rigid Motion Correction in Clinical Brain PET Imaging
Matthew G. Spangler-Bickell, Samuel A. Hurley, Ali Pirasteh, Scott B. Perlman, Timothy Deller, Alan B. McMillan
Journal of Nuclear Medicine Oct 2022, 63 (10) 1604-1610; DOI: 10.2967/jnumed.121.263309
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
    • Visual Abstract
    • Abstract
    • MATERIALS AND METHODS
    • RESULTS
    • DISCUSSION
    • CONCLUSION
    • DISCLOSURE
    • Footnotes
    • REFERENCES
  • Figures & Data
  • Info & Metrics
  • PDF

Related Articles

  • No related articles found.
  • PubMed
  • Google Scholar

Cited By...

  • Enhancing the Diagnostic Accuracy of Amyloid PET: The Impact of MR-Guided PET Reconstruction
  • Google Scholar

More in this TOC Section

  • 212Bi-Macroaggregated Albumin Inhibited Mouse Melanoma Growth by Regulating Cell Cycle Checkpoint Markers Without Promoting Living Cell Repopulation
  • In Vivo Visualization and Quantification of Brain Heat Shock Protein 90 with [11C]HSP990 in Healthy Aging and Neurodegeneration
  • 161Tb Radioimmunotherapy as a Treatment for CD30-Positive Lymphomas
Show more Basic Science Investigation

Similar Articles

Keywords

  • PET
  • Image Reconstruction
  • data-driven motion correction
  • brain imaging
SNMMI

© 2025 SNMMI

Powered by HighWire