|
|
||||||||
Basic Science Investigations |
1 Service Hospitalier Frédéric Joliot, Commissariat à lEnergie Atomique, Orsay, France
2 Department of Radiology, University of Pittsburgh, Pittsburgh, Pennsylvania
3 Imaging Research Laboratory, Department of Radiology, University of Washington, Seattle, Washington
4 CTI, Knoxville, Tennessee
| ABSTRACT |
|---|
|
|
|---|
Key Words: PET whole-body acquisitions noise equivalent count rate
| INTRODUCTION |
|---|
|
|
|---|
The aim of this study was to derive guidelines on the activity level of the radiotracer (dose) that should be injected into the patient to optimize the statistical quality of the data for 2D and 3D whole-body PET acquisitions. For that purpose, we used the noise equivalent count (NEC) rate (5) as a figure of merit. This issue was not addressed on replicate patient datasets because of ethical considerations. Instead, an anthropomorphic phantom was used to mimic realistic whole-body 18F-FDG PET imaging conditions for the torso. We performed 2D and 3D whole-body acquisitions of the phantom on an ECAT EXACT HR+ scanner (CTI, Knoxville, TN/Siemens Medical Systems, Inc., Hoffman Estates, IL) (6,7) and measured NEC rates for different activity levels for several organs of interest. Then, we compared count rates measured from the anthropomorphic phantom and from a series of clinical patient scans to extrapolate results from this phantom experiment to clinical practice. Good correlations between the phantom and the patient scans were obtained for prompt and delayed coincidence count rates measured over the FOV as a function of the single-photon rate. We then demonstrated that the NEC index can be approximated as a function of the total prompt and delayed count rates over the FOV on the basis of the assumptions explained below. Using the measurements obtained from the whole-body phantom over a large range of activity, we fitted polynomial models to the prompt and delayed coincidence rates as a function of single-photon rate, allowing us to derive an expression of the patient NEC rate as a function of the single-photon rate. Finally, we demonstrated a good correlation between the patient single-photon rate and the mean dose per weight at acquisition start normalized by the body mass index (BMI). This allowed estimation of the patient NEC rate as a function of the injected dose and BMI. Conversely, it also allowed estimation of the injected dose that maximized the NEC rate as a function of patient height and weight.
The NEC index has been widely used to compare scanner performance for brain imaging (810). More recently, the NEC performance of different scanners has been investigated for whole-body imaging based on phantom or patient data (1113). The originality of our work, compared with previous studies, is the use of a realistic model of 18F-FDG distribution in the human torso, in particular accounting for accurate activity distribution outside the FOV, which can significantly affect the NEC rate. Moreover, by investigating correlations between phantom and patient studies, we were able to demonstrate the consistency of our model and to extrapolate results to clinical practice.
| MATERIALS AND METHODS |
|---|
|
|
|---|
|
|
|
The rate of random coincidences was measured using a delayed coincidence window. The transmission scan was used to determine the fraction of the FOV f subtended by the phantom to determine the rate of random coincidences R within the object boundaries.
To account for the effect of activity sources outside the FOV, we estimated the scatter distribution in each acquisition mode using the method of Cherry et al. (18) based on the comparison of the 2D and 3D sinograms of the direct planes. The sinograms acquired in 2D mode were corrected for detector efficiency variations and then corrected for scatter using the 1-dimensional deconvolution method of Bergström et al. (19). The resulting sinograms were considered as the reference for the distribution of true coincidences. The 2D and 3D scatter sinograms were calculated by subtracting this reference from the 2D and 3D emission sinograms after efficiency normalization and correction for isotope decay for the 3D acquisitions. The transmission scan was used to define an object mask that was applied to the sinograms to determine the rate of scattered coincidences S within the phantom boundaries.
Finally, the true event rate T was calculated from the prompt coincidence rate (sum of the true, scattered, and random coincidences) by subtracting the scatter sinograms derived previously and the delayed coincidences.
We used a linear function to fit the variations of the single-photon rate SP as a function of the activity concentration A for data points at which the dead time was negligible and for acquisitions centered on the thorax and abdomen. This determined the parameters
2D and
3D in:
![]() | (Eq. 1) |
Deviations from the linearity of Equation 1 will occur because of scanner dead-time effects. The dead-time effect on the single-photon rate was measured by comparing the fitted and measured single-photon rates at higher activity levels. The same methodology was used to estimate the dead time for the true coincidence rates.
NEC Calculation
The NEC rate formula is defined (Eq. 7 of Strother et al. (5)), for the case of noiseless scatter correction, as:
![]() | (Eq. 2) |
Using Equation 2 and the phantom count rates measured as described in the previous section, we calculated the NEC rates in 2D and 3D modes for acquisitions centered on the brain, thorax, and abdomen of the whole-body phantom as a function of the activity A.
Estimation of Patient NEC Rates as a Function of Single-Photon Rates
For the patient studies, we express the NEC rate in terms of the directly measured prompt coincidence rate P = TFOV + SFOV + RFOV (where TFOV, SFOV, and RFOV are, respectively, the true, scattered, and random coincidences measured over the entire FOV), the delayed coincidence rate D = RFOV, and the scatter fraction
s = S/(S + T). Because T = TFOV and R = fRFOV, and if we approximate S
SFOV as is typically done by assuming a forward peaked scatter distribution (5,8,20), we can write:
![]() | (Eq. 3) |
sFOV is the scatter fraction using the entire scanner FOV, defined as
sFOV = SFOV/(SFOV + T). A result similar to Equation 3 was derived by Barnes et al. (20) for brain imaging applications. Because S < SFOV, the estimated NEC rates are reduced compared with those that would be measured following the NU 2-2000 standards of the National Electrical Manufacturers Association ([NEMA] Rosslyn, VA) (21). This does not, however, significantly affect the location of the peak NEC rate with respect to activity in the phantom or patient. This expression of the NEC rate depends on, first, the activity distribution that determines the count rates P and D and, second, the terms f and
sFOV that depend on patient morphology. To estimate NEC rates for patient studies, we examined the correlation of the prompt and delayed coincidence rates and single-photon rates between the phantom and patient datasets following the method of Smith et al. (13). Two bed positions centered on the thorax and abdomen were selected for comparison. Twenty-one 3D whole-body scans (14 male, 7 female) and ten 2D whole-body scans (3 male, 7 female) of clinical oncology patients were selected on the criterion that they matched the positions of the anthropomorphic phantom organs inside and outside the FOV. PET was performed as part of research protocols approved by local ethical review committees. Patient body weights averaged 67 ± 13 kg and ranged from 45 to 90 kg. Patient heights averaged 170 ± 9 cm and ranged from 150 to 182 cm. For both patient height and patient weight, median values were not significantly different from average values. Patients received a dose corresponding to approximately 3.7 MBq/kg (100 µCi/kg) and were scanned 1 h after injection. The mean activity concentration in the patients at the start of scanning was 2.5 ± 0.5 MBq/kg (67 ± 14 µCi/kg). To estimate the fraction f of the FOV subtended by the patient cross-sectional shape, the semiaxis values of each patients torso were measured on transaxial slices of the emission scan of the thorax, assuming an elliptic shape. The mean values were 110 ± 11 mm for the semiminor axis and 148 ± 11 mm for the semimajor axis of the ellipse. The fraction f was then numerically computed for each patient considering only the LORs perpendicular to the scanner axis (no oblique LORs). This gave a mean fraction f of 0.47 ± 0.04.
On the basis of the method of Smith et al. (13), the variations of the prompt and random coincidence rate over the FOV as a function of the single-photon rate were plotted for the phantom and patient datasets. As shown in the Results, we found that curves for the phantom and patients were superimposed. This allowed us to use the phantom count rates measured over a large range of activity, unlike patient data, to adjust a polynomial model to the variations of the prompt coincidence rate P as a function of the single-photon rate SP and derive the function h in:
![]() | (Eq. 4) |
To determine patient NEC rate, we also confirmed that the delayed coincidence rate D is fully determined by the single-photon rate as:
![]() | (Eq. 5) |
s is a proportionality constant accounting for the scanner sensitivity (s) and the coincidence time window (2
) (22).
Substituting P and D in Equation 3 by their expression as a function of the single-photon rate in Equations 4 and 5 gives:
![]() | (Eq. 6) |
Equation 6 allows predicting the patient NEC rate as a function of single-photon rate by determining the function h in Equation 4 and the proportionality constant 2
s in Equation 5 from the measured phantom data. This, however, requires fixing the parameters f and
sFOV. Based on our series of patients, we found that f ranged from 0.43 to 0.51. To estimate the influence of this factor on the NEC rate, the NEC curves were plotted with f set to these 2 extreme values. Results indicated that f had no significant effect on the location of the peak NEC rate, which is the main point of interest in this study. Thus, f was considered as a constant set to the estimated average value of 0.47. Similarly, Equation 6 indicates that uncertainties in the global scale factor (1 -
sFOV)2 may affect the magnitude of the NEC rate but not the location of the peak value. Based on these assumptions, the relative patient NEC rate may be expressed as:
![]() | (Eq. 7) |
Then, if we establish a relationship between SP and some index related to the dose injected into the patient, we can derive the variation of the NEC as a function of patient dose.
Single-Photon and NEC Rates as a Function of Patient Dose
We investigated the patient single-photon rates as a function of the dose per weight d/w, where w is the patient weight in kg and d is the injected dose in megabecquerels corrected for the delay between the injection and the acquisition. A linear regression was applied to the variation of the 3D single-photon rate SP3D as a function of the mean activity per weight similar to Equation 1:
![]() | (Eq. 8) |
To improve the correlation between SP3D and c by accounting for differences in habitus, which influences attenuation, we also normalized c by the relative BMI (23,24):
![]() | (Eq. 9) |
= c · (BMI/
BMI
) and
BMI
is the average BMI corresponding to a 70-kg and 170-cm-tall patient (
BMI
= 24.2 kg/m2). The BMI index is used as a general measure of obesity and was calculated as BMI = w/h2, where h is the patient height in meters.
The correlations between the 2D single-photon rate SP2D and the activity concentrations c and
were not derived from the set of 2D clinical data because of the small sample size (10 patients). Instead, these correlations were derived from those measured on the 3D datasets, assuming that the geometric similarity between the phantom and patient studies with respect to single-photon rate allows us to write
and, thus,
![]() | (Eq. 10) |
The combination of Equations 7 and 9 gives the predicted NEC rate as a function of
:
![]() | (Eq. 11) |
This allows estimating the injected dose that maximizes NEC rate for both 2D and 3D acquisitions as a function of patient height and weight.
| RESULTS |
|---|
|
|
|---|
|
|
|
2D and
3D in Equation 1. These parameters (in units of Mcps/kBq/mL) for the thorax and abdomen were, respectively, 0.32 and 0.47 in 2D mode and 1.16 and 1.54 in 3D mode. By comparing the fitted and the measured single-photon rates, we found that the dead time for 2D mode single-photon rates was negligible over the range of activity concentration considered in this study (maximum activity concentration A of 25 kBq/mL). The corresponding dead-time fractions for the 3D single-photon rates are reported in Table 4.
|
|
|
For the torso, the NEC rates measured in 3D mode were higher than the 2D values up to an activity concentration A of 10 kBq/mL for the thorax and 11.5 kBq/ml for the abdomen (Figs. 4B and 4C). The maximum 3D NEC rates were obtained for an activity concentration of 9 kBq/mL for the thorax and 7 kBq/mL for the abdomen. The true coincidence dead times at the peak 3D NEC rate were 10% for the thorax and 13% for the abdomen. The corresponding single-photon dead times were 3% for the thorax and 4% for the abdomen.
Estimation of Patient NEC Rates as a Function of Single-Photon Rates
Figure 5 compares the variations of the prompt and random coincidence rates over the FOV as a function of the single-photon rates measured for the anthropomorphic phantom data and the clinical datasets in 2D and 3D modes for bed positions centered on the thorax and abdomen. Patient and phantom data in 3D mode were superimposed for single-photon rates between 5 and 9 Mcps, corresponding to an activity concentration A ranging from 2.5 to 9.5 kBq/mL. In 2D mode, phantom and patient data were superimposed for single-photon rates between 2 and 4 Mcps. Variations of the prompt coincidence rate were well described by a second-order polynomial function. Coefficients of these polynomials are reported in Figures 5A and 5B for 2D acquisitions centered on the thorax and the abdomen, respectively, and in Figures 5C and 5D for 3D acquisitions. This determines the function h in Equation 4. We also verified the validity of Equation 5 for the random coincidence rate. The constant 2
s was, respectively, 1,220 ns in 2D mode and 2,500 ns in 3D mode.
|
= c · (BMI/
BMI
).
|
, with r2 = 0.66 for the thorax and r2 = 0.74 for the abdomen (Figs. 6B and 6D).
The linear relationship between SP3D and c or
in Equations 8 and 9 is valid as long as the dead-time effect on the rate of registered single photons is kept negligible. We found that the single-photon dead time in 3D exceeded 10% for single-photon rates in the thorax and abdomen higher than 17 Mcps (Table 4). This value of 10% single-photon dead time was considered as the upper limit of validity of the linear relationship between SP3D and c or
.
The proportionality constants
2D for the thorax and the abdomen were derived from the 3D values using Equation 12.
2D equaled 0.85 for the thorax and the abdomen.
Estimating Dose Required to Achieve Peak 2D and 3D NEC Rates
Combining Equations 7 and 9, we were then able to predict the 2D and 3D relative NEC variations as a function of
on the basis of Equation 11. Figure 7 shows the predicted 2D and 3D NEC rates for the thorax and the abdomen as a function of the dose injected to a standard 70-kg and 170-cm-tall patient (BMI = 24.2 kg/m2) and assuming a 90-min uptake period. The estimated patient peak NEC rates for the thorax and the abdomen were never reached in 2D mode within the range of activity considered. For 3D acquisitions, the peak NEC rates were achieved for a mean injected dose of 450 MBq (12.2 mCi) for the thorax and of 500 MBq (13.5 mCi) for the abdomen.
|
BMI
= 24.2 kg/m2) would be achieved for a mean starting-time activity of 4.1 MBq/kg. This activity concentration, as previously derived from Figure 7B, corresponds to an injected dose of 500 MBq (13.5 mCi), assuming a 90-min uptake period, or 413 MBq (11.2 mCi) for a 60-min uptake.
|
| DISCUSSION |
|---|
|
|
|---|
= c · (BMI/
BMI
), significantly improved the correlation (Figs. 6B and 6D). There was, however, remaining dispersion in the patient single-photon rates expressed as a linear function of
for the 21 patients studied. This was likely due to intersubject physiologic and biodistribution variations and urinary excretion. These correlations allowed estimating the injected dose that optimizes the NEC rate for 2D and 3D acquisitions. Results showed that neither of the 2D or 3D acquisition modes performs best overall according to NEC for torso imaging, but relative performance depends on allowable dose and the local imaging environment, which is affected by factors such as the activity outside the axial FOV. The peak 2D NEC rates for the thorax and the abdomen were never reached within an acceptable range of activity, suggesting that the optimal injected dose in 2D mode according to NEC should be driven by ethical considerations based on the maximal allowable dose delivered to the patient. For 3D acquisitions, the injected dose that optimizes the NEC rate as a function of weight and height is given in Figure 8.
The main limitation to the predicted NEC rate for patients in Figure 7 is that these curves are based on the assumption of the linear relationships between SP and A in Equation 1 and between SP and
in Equation 9. These linear relationships become invalid at high activity because of dead-time effects. We chose a 10% single-photon dead time as the upper limit of the linear range. The maximum single-photon dead time in 2D mode was lower than 5% within the range of activity considered in this study, so that we can be confident about the predicted NEC rates for 2D acquisitions (Fig. 7). For 3D acquisitions, the shape of the NEC curves for injected-dose values higher than 710 MBq (19 mCi) should be interpreted with caution, as they correspond to single-photon dead times higher than 10%. It is likely that fitting a nonlinear relationship between SP and A and between SP and
would improve the accuracy of the predicted NEC rate for this range of elevated dose by accounting for dead-time effects. The peak 3D NEC rate, however, is in the linear range of Equations 1 and 9. As a consequence, the curves of the predicted injected dose per weight as a function of BMI presented in Figure 8 are not affected by dead time.
Another limitation of the predicted NEC rate for patients is the fixed value of the fraction f of the FOV subtended by the object. As mentioned in the methods, we estimated the error on the location of the 3D peak NEC rate by plotting the NEC curves corresponding to the 2 extreme values of f = 0.43 and f = 0.51. The maximum difference in the activity concentration corresponding to the peak NEC rate was less than 5%, thus suggesting that our approximation in fixing the parameter f in Equation 6 does not significantly affect the results of this study. The scatter fractions used to estimate the NEC curves in Figure 7 were set to the values measured for the whole-body phantom. Thus, these curves do not provide accurate information on the absolute NEC rates for patients, since this value will depend on the scatter fractions. This scale change, however, will not affect the activity at which the peak NEC rate is reached.
The complex shape of the anthropomorphic phantom did not allow measurement of the scatter fraction with an inserted line source, as recommended by the NEMA procedure (25). As a consequence, the scatter fractions for the 20-cm-diameter cylinder in Table 3 differ slightly from the NEMA protocol values (6).
The 3D NEC rates for the brain in Figure 4A correlate well with the 3D NEC rate measured for a 10-cm-diameter (1,217 mL) cylinder on an HR+ scanner (26), although differences in phantom geometry may explain why the NEC curve for the brain phantom tends to reach the peak NEC rate at a lower activity. Our measurements from the brain phantom also correlate well with those reported by Wienhard et al. (27), comparing the ratio of the 3D NEC rates to the 2D NEC rates for cerebral studies on patients, and also with the work of Bailey (28) suggesting that 3D acquisitions detect fewer random events than 2D acquisitions for the same rate of true events for brain-type geometries (Figs. 3A and 3B).
The ratios of 3D NEC to 2D NEC for torso imaging were consistent with the NEC ratios of a patient whole-body scan reported by Townsend et al. (26) for the HR+ scanner. They also correlated well with the phantom measurements of Laforest et al. (12) on the HR+ scanner, suggesting that the peak 3D NEC is achieved for an activity concentration lower than 10 kBq/mL. However, the ratios of 3D to 2D NEC rates were lower in our experiment, and this finding may be explained by differences in the estimated scatter fractions. Discrepancies were observed when our results were compared with those of Badawi et al. (11), who reported NEC rates for the ECAT 951 (CTI/Siemens) scanner using the elliptic chest phantom of the European Economic Community, and with the NEC performances reported by Smith et al. (13) for the C-PET (ADAC Laboratories/Philips Medical Systems, Inc., Milpitas, CA) scanner based on a 20-cm-diameter by 70-cm-long cylinder. These discrepancies originate from differences in scanner characteristics and phantom shapes. In addition, the measurements by Badawi et al. did not include the effect of activity outside the axial FOV (11).
Finally, as we have already noted, the NEC figure of merit allows estimating the statistical quality of the data in the acquisition with regard to the injected dose to the patient. The NEC rate does not account either for inaccurate correction methods of the scattered or random coincidences or for resolution and noise correlation changes introduced by processing and reconstruction of the data. Further studies are required to compare the image quality obtained from 2D and 3D acquisitions, such as human or numeric observer performance studies (29,30). This is the purpose of our ongoing work based on the methodology developed for use with volumetric image display (31,32).
| CONCLUSION |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
For correspondence or reprints contact: Carole Lartizien, PhD, CERMEP, Hôpital Neuro-Cardiologique, 59, Boulevard Pinel, 69003 Lyon, France.
E-mail: clartizien{at}netcourrier.com
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
B. J. Pichler, H. F. Wehrl, and M. S. Judenhofer Latest Advances in Molecular Imaging Instrumentation J. Nucl. Med., June 1, 2008; 49(Suppl_2): 5S - 23S. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. Bonutti, E. Cattaruzzi, E. Cragnolini, M. Floreani, C. Foti, M. R. Malisan, E. Moretti, O. Geatti, and R. Padovani COUNT-RATE ANALYSIS FROM CLINICAL SCANS IN PET WITH LSO DETECTORS Radiat Prot Dosimetry, April 24, 2008; (2008) ncn153v1. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. El Fakhri, P. A. Santos, R. D. Badawi, C. H. Holdsworth, A. D. Van Den Abbeele, and M. F. Kijewski Impact of Acquisition Geometry, Image Processing, and Patient Size on Lesion Detection in Whole-Body 18F-FDG PET J. Nucl. Med., December 1, 2007; 48(12): 1951 - 1960. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. S. Yap, J. Czernin, M. C. Fishbein, R. B. Cameron, C. Schiepers, M. E. Phelps, and W. A. Weber Evaluation of Thoracic Tumors With 18F-Fluorothymidine and 18F- Fluorodeoxyglucose-Positron Emission Tomography. Chest, February 1, 2006; 129(2): 393 - 401. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. D. Badawi and M. Dahlbom NEC: Some Coincidences Are More Equivalent Than Others J. Nucl. Med., November 1, 2005; 46(11): 1767 - 1768. [Full Text] [PDF] |
||||
![]() |
C. C. Watson, M. E. Casey, B. Bendriem, J. P. Carney, D. W. Townsend, S. Eberl, S. Meikle, and F. P. DiFilippo Optimizing Injected Dose in Clinical PET by Accurately Modeling the Counting-Rate Response Functions Specific to Individual Patient Scans J. Nucl. Med., November 1, 2005; 46(11): 1825 - 1834. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. Brasse, P. E. Kinahan, C. Lartizien, C. Comtat, M. Casey, and C. Michel Correction Methods for Random Coincidences in Fully 3D Whole-Body PET: Impact on Data and Image Quality J. Nucl. Med., May 1, 2005; 46(5): 859 - 867. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Brogsitter, T. Gruning, R. Weise, P. Wielepp, O. Lindner, R. Korfer, and W. Burchert 18F-FDG PET for Detecting Myocardial Viability: Validation of 3D Data Acquisition J. Nucl. Med., January 1, 2005; 46(1): 19 - 24. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lubberink, R. Boellaard, A. P. van der Weerdt, F. C. Visser, and A. A. Lammertsma Quantitative Comparison of Analytic and Iterative Reconstruction Methods in 2- and 3-Dimensional Dynamic Cardiac 18F-FDG PET J. Nucl. Med., December 1, 2004; 45(12): 2008 - 2015. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Lartizien, P. E. Kinahan, and C. Comtat A Lesion Detection Observer Study Comparing 2-Dimensional Versus Fully 3-Dimensional Whole-Body PET Imaging Protocols J. Nucl. Med., April 1, 2004; 45(4): 714 - 723. [Abstract] [Full Text] [PDF] |
||||
![]() |
T.-C. Yen, K.-K. Ng, S.-Y. Ma, H.-H. Chou, C.-S. Tsai, S. Hsueh, T.-C. Chang, J.-H. Hong, L.-C. See, W.-J. Lin, et al. Value of Dual-Phase 2-Fluoro-2-Deoxy-D-Glucose Positron Emission Tomography in Cervical Cancer J. Clin. Oncol., October 1, 2003; 21(19): 3651 - 3658. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Knesaurek, J. Machac, B. R. Krynyckyi, and O. D. Almeida Comparison of 2-Dimensional and 3-Dimensional 82Rb Myocardial Perfusion PET Imaging J. Nucl. Med., August 1, 2003; 44(8): 1350 - 1356. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||