Abstract
Objective: Head motion occuring during brain PET studies leads to image blurring and to bias in measured local quantities. Our first objective was to implement an accurate list-mode-based rigid motion correction method for PET data acquired with the mMR synchronous Positron Emission Tomography/Magnetic Resonance (PET/MR) scanner. Our second objective was to optimize the correction for [11C]-PIB scans using simulated and actual data with well-controlled motions. Results: An efficient list-mode based motion correction approach has been implemented, fully optimized and validated using simulated as well as actual PET data. The average spatial resolution loss induced by inaccuracies in motion parameter estimates as well as by the rebinning process was estimated to correspond to a 1 mm increase in Full Width Half Maximum (FWHM) with motion parameters estimated directly from the PET data with a temporal frequency of 20 secs. The results show that it can be safely applied to the [11C]-PIB scans, allowing almost complete removal of motion induced artifacts.The application of the correction method on a large cohort of 11C-PIB scans led to the following observations: i) more than 21% of the scans were affected by a motion greater than 10 mm (39% for subjects with Mini-Mental State Examination -MMSE scores below 20) and ii), the correction led to quantitative changes in Alzheimer-specific cortical regions of up to 30%. Conclusion: The rebinner allows an accurate motion correction at a cost of minimal resolution reduction. The application of the correction to a large cohort of [11C]-PIB scans confirmed the necessity to systematically correct for motion for quantitative results.