Skip to main content

Main menu

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI

User menu

  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart

Search

  • Advanced search
Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • View or Listen to JNM Podcast
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
Meeting ReportPhysics, Instrumentation & Data Sciences - Instrumentation

Physical Performance of a High-Resolution Total-Body PET Model with Sparse-Detector Rings

Sadek Nehmeh, Antonio Gonzalez Martinez, Alfonso Rios and Nicolas Karakatsanis
Journal of Nuclear Medicine June 2024, 65 (supplement 2) 241510;
Sadek Nehmeh
1Weill Cornell Medical College
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Antonio Gonzalez Martinez
2Institute for Instrumentation in Molecular Imaging, I3M-CSIC
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Alfonso Rios
3Ful-Body Imaging
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Nicolas Karakatsanis
4Weill Cornell Medical College, Cornell University
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Info & Metrics
Loading

Abstract

241510

Introduction: To evaluate the physical performance of a Monte Carlo (MC) model of a high-resolution clinical Total-Body PET model with sparse detector rings (TBSR-PET).

Methods: The TBSR-PET hosts 28 rings (64 cm in diameter), each with 80 detector blocks. Each detector block consists of an array of 8 (axial) x 1 (transaxial) semi-monolithic LYSO slabs of 3 (axial) x 24 (transaxial) x 20 (depth) mm3 each and an axial pitch of 3 mm, i.e. interleaved with axial gaps of width equal to the slabs' axial dimension. The scanner axial field-of-view (AFOV) is 142cm. MC simulation was carried out using the GATE package.

The physical performance (spatial resolution, sensitivity, and Image Quality) of the TBSR-PET model was assessed per NEMA NU 2-2012 standards. The system sensitivity was evaluated using a 70cm long line source (as per NEMA guidelines), as well as a 142cm line source that extended along the full scanner AFOV. The IEC phantom with 6 hot spheres (diameters: 10, 13, 17, 23, 28, and 37 mm) at a 4:1 target-to-background activity concentration ratio was simulated for 10 min centered at two AFOV positions: at the center and close to (12.6cm distance) one of the edges. The system matrix assumed an array of 24 (transaxial) x 20 (radial) virtual detector elements of 1 (transaxial) x 3 (axial) x 1 (depth) mm3 per slab. This was determined by an intrinsic FWHM resolution along the transaxial and depth dimensions of the slab of 1mm respectively, as estimated in a separate work using a Machine Learning approach. Image reconstruction was performed using CASToR (OSEM, 19 subsets, 2 iterations) with normalization, attenuation, randoms, scatter corrections, depth-of-interaction (DOI), inter-crystal scatter correction, and a time-of-flight resolution of 210 psec. The contrast-to-noise ratio (CNR) of the 10 mm diameter sphere was measured to optimize the iterations number. Resolution recovery was not included.

Results: TBSR-PET exhibited ~1mm FWHM spatial resolution in the radial, tangential, and axial directions at both the center and 1/4th the scanner AFOV positions. The scanner exhibited system sensitivities of 107.39 kcts/MBq and 75.82 kcts/MBq using the 70 cm and 142 cm long line sources respectively. All IEC phantom spheres in the reconstructed PET images were detectable. The IEC phantom images did not show any artifacts due to the sparse detector configuration. The percent contrast recovery (% CR) for the six IEC phantom spheres ranged between 67% and 91%, and between 67% and 92%, at the center and close to the edge of the AFOV respectively. Incorporation of inter-crystal scatter correction yielded an increase in the % CR of up to 18% and 30% for the 10mm diameter sphere at the center and close to the edge of the scanner AFOV, respectively, and that was inversely proportional to the sphere size. The background variability (% BV) for the six sphere sizes ranged between 7.07% - 5.60% and 7.92% - 4.38% at the center and close to the edge of the AFOV respectively. Inter-crystal scatter had no pronounced effect on the BV.

Conclusions: Sparse-detector rings configuration presents a potential cost-effective solution for Total-Body PET. Semi-monolithic scintillator allows achieving high intrinsic detector resolution, including DOI, and hence spatial resolution at no additional costs. However, reducing the image pixel size yields increased image noise, which was shown by the slight increase in % BV compared to that in commercial whole-body PET scanners with conventional pixel size (usually ~3mm2) at matched scan times. This can be compensated for by increasing the scan time.

Previous
Back to top

In this issue

Journal of Nuclear Medicine
Vol. 65, Issue supplement 2
June 1, 2024
  • Table of Contents
  • Index by author
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Physical Performance of a High-Resolution Total-Body PET Model with Sparse-Detector Rings
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
Physical Performance of a High-Resolution Total-Body PET Model with Sparse-Detector Rings
Sadek Nehmeh, Antonio Gonzalez Martinez, Alfonso Rios, Nicolas Karakatsanis
Journal of Nuclear Medicine Jun 2024, 65 (supplement 2) 241510;

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Physical Performance of a High-Resolution Total-Body PET Model with Sparse-Detector Rings
Sadek Nehmeh, Antonio Gonzalez Martinez, Alfonso Rios, Nicolas Karakatsanis
Journal of Nuclear Medicine Jun 2024, 65 (supplement 2) 241510;
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
  • Info & Metrics

Related Articles

  • No related articles found.
  • Google Scholar

Cited By...

  • No citing articles found.
  • Google Scholar

More in this TOC Section

  • Development of a portable and versatile PET System for Interactive Point-of-Care Imaging
  • Calibration Procedures for Multi-Center Dosimetry Studies with 177Lu Radiopharmaceuticals: Experience from the Canadian Cancer Trial Group PR.21 Trial (NCT 04663997)
  • Comparing PET and single photon planar imaging for F-18, Tc-99m, and At-211
Show more Physics, Instrumentation & Data Sciences - Instrumentation

Similar Articles

SNMMI

© 2025 SNMMI

Powered by HighWire