Abstract
We previously reported the basic performance of a prototype small cadmium telluride (CdTe) γ-camera (SSGC) intended for use in radioguided surgeries. In this study, we sought to confirm the favorable previous results and to extend the preliminary findings to examine the efficacy of the SSGC in an animal study and a clinical setting for sentinel lymph node biopsy. Methods: The prototype SSGC (1,024 pixels; field of view, 44.8 × 44.8 mm), equipped with a parallel-hole collimator, was used in both animal and clinical studies. 99mTc-phytate (18.5 MBq) was injected into the tongues and legs of 6 rabbits. In the clinical study, 74 MBq of 99mTc-phytate was injected into peritumoral regions in 8 patients with oral cancer. The detection of hot nodes by the SSGC was compared with that by a conventional scintillation γ-camera (CGC). Results: The SSGC detected 29 hot nodes in images of 6 rabbits after injection. The number of hot nodes was the same as the number seen in CGC studies, but the CGC required a longer acquisition time to produce comparable images. There were no differences between the SSGC and the CGC in terms of activity ratios and hot node-to-background ratios. The biodistribution of 99mTc-phytate in removed tissues was evaluated by contact radiography, and radioactivity was assayed with a γ-well counter. The mean ± SD radioactivity in specimens was 0.15% ± 0.15%, with a range of 0.01%–0.62%. In the clinical study, the SSGC detected 30 hot nodes with a 5- to 60-s acquisition time at 4 h after injection. The SSGC documented all hot nodes depicted by the CGC at 4 h after injection. Conclusion: The SSGC showed significant potential for the detection of sentinel lymph nodes in lymphoscintigraphy. The results of the studies suggested that the SSGC facilitates the exploration of hot nodes in sentinel lymph node biopsy.
- lymphoscintigraphy
- oncology
- semiconductor γ-camera
- CdTe
- handheld γ-camera
- sentinel lymph node biopsy
- radioguided surgery
- oral cancer
Radioguided surgery has been accomplished with the use of γ-probes, primarily for sentinel lymph node biopsy. However, γ-probes can provide only radioactivity counts and variable-pitch audio output based on the intensity of the detected radioactivity. Allowing the surgeon to see the spatial distribution of the increased radioactivity could result in more accurate delineation of the surgical field, thus facilitating the surgery. Therefore, we developed a prototype small semiconductor γ-camera (SSGC) that provides images of radioactivities for surgical use (1).
For radioguided surgery, the γ-camera should be miniaturized or handheld to facilitate its application in the restricted space of a surgical field. A semiconductor detector containing cadmium telluride (CdTe) or cadmium zinc telluride (CdZnTe) has shown excellent performance and quality for imaging (2–7). The semiconductor detector is considered to be the most suitable for miniaturization because it does not require the relatively massive photomultiplier tube that characterizes a traditional scintillation/photomultiplier tube detector. The solid-state technology allows for the fabrication of a camera head that is smaller, lighter, and more portable. CdTe or CdZnTe appears to be the technology of choice for developing an intraoperative imaging system.
The CdTe detector that we chose for the production of our SSGC resulted in better basic performance, including energy resolution and spatial resolution, than that achieved with a conventional scintillation γ-camera (CGC). The preliminary results of a simulation study for sentinel lymph node biopsy with the SSGC suggested that this camera would identify sentinel node radioactivity during the procedure (1).
Several reports of handheld small γ-cameras have been published (1,8–15), but studies to date have not verified the effectiveness of small γ-cameras in depicting nodes with increased radioactivities in the clinical setting of sentinel lymph node biopsy. In this study, we aimed to confirm our earlier results regarding the basic performance of the SSGC and to extend the preliminary findings by using the camera in an animal study and a clinical setting.
MATERIALS AND METHODS
SSGC
The camera head consists of a “pixelized” CdTe module (32 × 32 individual elements, total of 1,024 pixels) attached to a collimator (Acrorad Co. Ltd.) (1). The field of view is 44.8 × 44.8 mm. The camera head measures 152 × 166 × 65 mm. The collimator consists of a stack of 100 pieces of 0.1-mm-thick tungsten mesh sheets, each having square openings measuring 1.2 × 1.2 mm and a space of 0.2 mm to fit the geometry of CdTe pixels. The camera head weight is 2.7 kg, and the head is attached to the holding arm of the rack (Fig. 1) (1).
SSGC. (Left) SSGC head. (Right) Entire SSGC system.
The energy resolution (full width at half maximum) for a single element at the center of the field of view is 4.2% at 141 keV (99mTc), and the mean energy resolution of the CdTe detector arrays is approximately 7.8%. The spatial resolution, represented by the full width at half maximum, is 1.56 ± 0.05 (mean ± SD) mm. The detection efficiency is 0.30 count per second per kilobecquerel (99mTc) (1).
CGC
A CGC (SNC-5100R; matrix size, 512 × 512 pixels; field of view, 51 × 38 cm; Shimadzu) equipped with an NaI(Tl) crystal detector was used for comparison of CGC- and SSGC-generated images. The system resolution at the 99mTc energy (141 keV) was 10.4%, and the spatial resolution for 99mTc was 3.5 mm. A parallel-hole low-energy (high-resolution) collimator was used for imaging.
Animal Study Design
Six healthy male Japanese white rabbits (JW AIZ) (mean weight, 3.23 kg; range, 3.06–3.55 kg) were anesthetized with an injection of sodium pentobarbital (30 mg/kg intravenously) through the ear vein and placed in a prone position on a fixation bed. Stannous phytate kits were obtained from Daiichi Radioisotope Laboratories. A 99mTc-phytate solution was prepared by mixing the content of the product and 99mTc-pertechnetate. With a 27-gauge needle attached to a 1.0-mL syringe, a dose of 18.5 MBq of 99mTc-phytate in 0.2 mL was injected into 5 sites: the right side of the tongue (submucosally), forelegs (right side subdermally, left side intramuscularly), and posterior legs (right side subdermally, left side intramuscularly). 99mTc-phytate lymphoscintigraphy of the whole body was performed for all 6 rabbits.
Rabbits were imaged with the CGC at 1, 2, 4, and 6 h after injection with 10 min of exposure for each whole static image (512 × 512 pixels). Immediately after each CGC imaging study, the detector of the SSGC was placed on the body surface for each of the specified regions, and images were obtained with 5-, 10-, and 15-s acquisition times. Acquisition times for the SSGC were based on previous results (1). The routine acquisition time for the CGC was applied to produce comparable images.
Regions of interest (ROIs) were placed over the hot node and a surrounding background region on CGC-generated static images by use of square templates (15.2 × 15.2 mm) to obtain uptake ratios. The radioactivity in each ROI was integrated for 10 min. The isotope uptake was expressed as a ratio of the hot-node activity to the background activity (HN/BG) to compare potential changes in the regional 99mTc-phytate distribution over time.
HN/BG ratios also were determined in SSGC studies with square ROIs of 11 × 11 pixels (15.4 × 15.4 mm; approximately identical in size to the ROIs in the CGC studies), as in the CGC studies but with a different acquisition time (15 s).
After completion of the last imaging study, rabbits were euthanized by injection of pentobarbital at 100 mg/kg of body weight or more via the ear vein and dissected to remove tissues containing increased radioactivity in the neck and foreleg regions but not the posterior leg regions. Both SSGC and CGC were used to search for increased radioactivity. The biodistribution of radioactivity in excised specimens was confirmed by autoradiography. The radioactivity in excised tissues was measured with a γ-well counter. Tissue specimens were fixed in formalin, embedded in paraffin, and stained with hematoxylin–eosin. The study was approved by the Ethics Committee for Animal Research, The Nippon Dental University School of Life Dentistry at Niigata.
Autoradiography was performed by use of a phosphorimager (FLA2000; Fuji Film) with an imaging plate (IP) (BAS-SR2040; Fuji Film). The excised radioactive specimens were weighed, measured in length, and placed on the IP covered with a sheet of film of polyvinylidene. The radioactivity was exposed on the IP for 5 and 10 min to produce autoradiographic images in a darkroom, and the IP was processed to image the radioactivity by use of the FLA2000 device.
The radioactivities in the specimens and a set of reference 99mTc radioactivities were assayed with an Autowell γ-system (ARC-370 M; Aroca). The radioactivities were corrected for decay time, calculated, and expressed as percentages (activity in the specimen relative to the initial injected radioactivity).
Data on percentage variability were expressed as mean ± SD. A 2-tailed paired Student t test was used for 2-group comparisons for SSGC versus CGC. Probability values of less than 0.05 were considered statistically significant.
Clinical Study Design
Eight patients (4 men and 4 women; age [mean ± SD], 53.3 ± 21.0 y) with oral cancer were enrolled in the clinical study at Niigata Hospital (The Nippon Dental University School of Life Dentistry at Niigata). Primary lesions were in the tongue (n = 7) and lower gingiva (n = 1), and the stages were N0 (n = 3), N1 (n = 2), and N2c (n = 3). Sentinel node lymphoscintigraphy was performed after the injection of 99mTc-phytate (Daiichi Radioisotope Laboratories). The radiopharmaceutical was injected into 4 healthy sites surrounding the tumor and approximately 5 mm from the edge of the tumor (18.5 MBq/0.25 mL × 4) with a 27-gauge needle after the application of topical anesthetic. The injection involved creating a vesicle to attain administration to a submucosal depth. Although an N2c neck lesion was not an indication for sentinel lymph node biopsy, the 3 patients who had such lesions were included in the study simply for lymphoscintigraphy.
Sentinel lymph node biopsy had not been introduced to the hospital at the time of this clinical study. Informed consent was obtained from the patients after explanation of the nature and possible risks of their voluntary participation in the clinical study. This research was approved by The Ethics Committee, The Nippon Dental University School of Life Dentistry at Niigata.
The same SSGC as that used in the animal study was applied in the clinical study. First, comparative CGC images were obtained at 1, 2, 4, 6, and 24 h after injection. SSGC studies followed CGC studies at each time setting, and images were obtained with 5-, 10-, and 15-s acquisition times. Occasionally, additional 30- or 60-s images were obtained. Comparable CGC images were obtained with routine 10-min acquisition times. Head and neck imaging with CGC was performed with patients in the prone position. Posterior-to-anterior and lateral projection images were completed with 10-min acquisition times. SPECT also was performed at 4 h after injection. ROIs were placed over the hot node and the surrounding background area on both SSGC and CGC static images, as in the animal study, to obtain HN/BG ratios. The lymphoscintigrams obtained with both the SSGC and the CGC were interpreted by 3 independent radiologists who were experts in the field of body imaging. The uptake of 99mTc-phytate was evaluated subjectively by consensus of the radiologists.
The HN/BG ratios were expressed as mean ± SD and compared. An ANOVA was used to compare the ratios for the SSGC and the CGC. Probability values of less than 0.05 were considered statistically significant.
RESULTS
Animal Study
In the animal study, the SSGC detected a total of 29 hot nodes in images of 6 rabbits, each at 2, 4, and 6 h after injection (15-s acquisition time). The number of hot nodes detected was the same as the number detected by the CGC: 11 in the neck region, 6 in the left anterior leg region, 7 in the right anterior leg region, 2 in the left posterior leg region, and 3 in the right posterior leg region. The SSGC clearly depicted the increased radioactivity at shorter acquisition times (5, 10, and 15 s) than the CGC (standard 600-s acquisition time) (Fig. 2).
Comparison of lymphoscintigrams of rabbits. (A) (Left) CGC image shows hot nodes at 4 h after injection with acquisition time of 600 s. Hot node in left foreleg region is circumscribed by rectangular frame. (Right) SSGC image at 4 h after injection with 5-s acquisition time shows same hot node. (B) (Left) Two sentinel nodes in neck region are shown in rectangular frame on CGC image after 600-s acquisition time. (Right) These 2 sentinel nodes are shown on SSGC image with 15-s acquisition time (4 h after injection). Fl = foreleg; inj = injection site; lt = left; rt = right.
The HN/BG ratios of 2-h images (9 nodes) and 4-h images (11 nodes) of 6 rabbits were calculated in both γ-camera studies. The HN/BG ratios for the SSGC were 38.1 ± 32.9 (2 h) and 39.3 ± 30.5 (4 h). The HN/BG ratios for the CGC were 31.6 ± 35.4 (2 h) and 29.4 ± 26.6 (4 h). There were no significant differences in the HN/BG ratios between the γ-cameras (P = 0.864) (Fig. 3).
Comparison of HN/BG ratios between SSGC and CGC in animal study (9 hot nodes at 2 h; 11 hot nodes at 4 h), demonstrating no significant differences (P = 0.864).
All 22 tissues in the neck and anterior leg regions with increased radioactivities were excised and confirmed by microscopic examination to be lymph nodes. An autoradiographic study showed that all 22 lymph nodes were radioactive. The weights (mean ± SD) of the lymph nodes were 0.14 ± 0.09 g in the neck, 0.05 ± 0.07 g in the anterior leg regions, and 0.09 ± 0.08 g for all 22 lymph nodes. The mean maximum lengths of the lymph nodes were 12.9 ± 6.4 mm in the neck, 5.5 ± 3.8 mm in the anterior leg regions, and 8.9 ± 6.0 mm for all 22 lymph nodes.
Radioactivity in the excised specimens was assayed with a γ-well counter. The mean percentages of radioactivity in the excised lymph nodes relative to the initial injected radioactivity were 0.21% ± 0.18% for lymph nodes in the neck, 0.12% ± 0.11% for lymph nodes in the anterior leg regions, and 0.15% ± 0.15% for all 22 lymph nodes. The percentages ranged from 0.01% to 0.62% (Fig. 4 and Table 1). The 3 lymph nodes containing 0.01% of the initial injection volume (18.5 MBq of 99mTc) were detected by both γ-cameras at 6 h after injection.
Confirmation of radioactivity in excised specimens by autoradiography. (A) (Left) CGC image at 4 h after injection with 600-s acquisition time. (Right) SSGC image at 4 h after injection with 15-s acquisition time. (B) (Left) Three specimens excised from right foreleg region contained high levels of radioactivity (percentage of initial injection of 99mTc-phytate). (Right) Autoradiographic image of specimens.
Lymph Node Weight and Percentage of Dose (99mTc) in Nodes
Clinical Study
The SSGC detected all hot nodes that were detected by the CGC in 8 patients with oral cancer (10 neck sites). The numbers of hot nodes in the neck were 31, 29, 30, 27, and 16 at 1, 2, 4, 6, and 24 h after injection, respectively (Table 2). The hot nodes were revealed in the following regions at 4 h after injection: 7 hot nodes at level I, 18 hot nodes at level II, 3 hot nodes at level III, and 2 hot nodes at level IV. The node-level classification is from the American Joint Committee on Cancer (2002). Eighteen of the 30 hot nodes (60%) were localized at level II. The SSGC revealed the hot nodes with short acquisition times (Fig. 5).
(A) T2N1M0 (pN0) squamous cell carcinoma of tongue in 58-y-old man. (Left) CGC image shows neck hot nodes (2 h after injection, 600-s acquisition time). (Middle) SSGC image clearly shows 4 hot nodes with short (15-s) acquisition time. (Right) Sentinel nodes are shown with acquisition time of 60 s at 24 h after injection. (B) T1N1M0 (pN0) squamous cell carcinoma of tongue in 34-y-old man. (Left) CGC image shows neck hot nodes (24 h after injection, 600-s acquisition time). (Middle) SSGC image clearly shows lymphatic vessel at 6 h after injection with acquisition time of 15 s. (Right) Three hot nodes are shown at 24 h after injection (acquisition time, 30 s).
Number of Nodes and HN/BG Ratios for Nodes with Increased Radioactivity Identified by SSGC and CGC at 4 Hours After Injection
Eighteen hot nodes from 8 patients were chosen for determining the HN/BG ratios for the SSGC and the CGC to compare image quality. The mean HN/BG ratios of the 18 hot nodes for the SSGC were 10.9 ± 7.1, 9.6 ± 6.3, and 4.3 ± 1.7 at 2, 4, and 24 h after injection, respectively. The mean HN/BG ratios of the hot nodes for the CGC were 8.8 ± 5.6, 7.1 ± 4.4, and 3.1 ± 1.7 at 2, 4, and 24 h after injection, respectively. There were no significant differences in the HN/BG ratios between the γ-cameras (P = 0.969) (Fig. 6).
Comparison of HN/BG ratios between SSGC and CGC in clinical study (n = 18), demonstrating no significant differences (P = 0.969).
DISCUSSION
Sentinel lymph node biopsy is an established and widely available procedure for breast cancer and melanoma (16–25). It is also used for the management of head and neck cancer and other malignancies. Various 99mTc radiocolloids are used for this technique. In breast cancer, the doses of injected radiopharmaceuticals have been reported to be 5–10 MBq (18), 10–20 MBq (19), and 3.7–111 MBq (20). Similar doses (15–111 MBq) have been used for the procedure in patients who have oral cancer (21–25).
When a sentinel lymph node biopsy is performed, only a very small amount of radiopharmaceutical accumulates in the sentinel node or nodes. Although the injection methods and 99mTc-labeled particles differed, the average accumulated radioactivities relative to the injected doses per lymph node in breast cancer studies were reported to be 0.9% and 0.60% (26,27), and the range was 0.001%–2.5% (27,28). Some surgeons have estimated the percentage of the injected dose per lymph node to be about 0.15%.
It is a challenge to visualize the small amount of radioactivity in the target tissue by using a small γ-camera during the procedure. In our previous simulation study, the SSGC reliably detected small amounts of radioactivity (150 kBq; 0.15% of an estimated injected dose of 100 MBq of 99mTc) in simulated nodes in a water bath at up to a 30-mm underwater depth after 18 h. Even after 24 h, when the radioactivity had decreased to 1/16th of its initial value (9.4 kBq), the simulated sentinel nodes still could be recognized with short acquisition times (1).
In the animal study, the SSGC clearly depicted sentinel lymph nodes after 99mTc-phytate injection. The percentage of injected radioactivity (18.5 MBq) in hot nodes ranged from 0.01% to 0.62%, with an average of 0.15% at 6 h after the injection. At the time of the last imaging study, the initial radioactivity had decreased by almost 50% with the physical decay of 99mTc, indicating that the SSGC could image the radioactivity of at least 0.00093 MBq in a hot node. The SSGC revealed all hot nodes that were revealed by the CGC, and the acquisition times of the SSGC were very short (5, 10, or 15 s) compared with the usual 5 or 10 min required for the CGC. The absence of radioactivity in the lymphatic basin was confirmed easily by the SSGC after the excision of a hot node or nodes, especially close to the injection site. Furthermore, the SSGC provided simple verification of radioactivity in specimens after surgery.
The clinical study also showed the favorable performance of the SSGC in detecting hot nodes. The SSGC detected all hot nodes detected by the CGC, but with short acquisition times (5, 10, 15, 30, and 60 s). In both animal and clinical studies, sentinel lymph nodes, even lymphatic channels in some images, were clearly shown by the SSGC. The SSGC could reveal more details than the CGC at shorter acquisition times. In the previous simulation study, we tested the SSGC with the “shine through” phenomenon, in which too much radioactivity from the injection site interferes with the detection of a small amount of nodal radioactivity situated close to the injection site during the exploration of sentinel lymph nodes with a γ-probe. The SSGC could resolve a simulated sentinel lymph node situated close to the injection site. In both of the present studies, the SSGC demonstrated lymph nodes neighboring the injection site.
Real-time imaging also was attained in the clinical experiment; such imaging may provide surgeons with increased maneuverability for surgical procedures.
The favorable results obtained in our animal and clinical studies were primarily attributable to the excellent basic performance of the SSGC, described earlier in this article. The basic performance seems to be notable, even among small γ-cameras with scintillation-based technology (8–12) or semiconductor technology (13,14).
The camera head for handheld small γ-cameras weighs as little as 820 g–2.7 kg (1,10,11,13–15). The field of view of such cameras varies: 20 × 20 mm (9), 24 mm in diameter (12), 32 × 32 mm (13), 40 mm in diameter (10), 40 × 40 mm (14), 44 × 44 mm (15), 44.8 × 44.8 mm (1,15), and 95 × 95 mm (15). However, these handheld small γ-cameras may not replace γ-probes completely in radioguided surgery because their detector heads still obscure the surgical field. The SSGC does offer a tremendous advantage with visualization of the target uptake during surgery. Furthermore, surgeons can use the SSGC to confirm the location of radioactivity in excised specimens during surgery. Recently, the usefulness of a handheld miniature γ-camera for parathyroidectomy was reported (11). In a commentary about the article, Rubello and Mariani speculated that the combined use of a γ-probe and the handheld miniature γ-camera would be practical for surgery, especially when the target tissues are located ectopically or deeply (29). We agree with that approach for sentinel lymph node biopsy; the simultaneous use of the devices better facilitated the exploration of hot nodes than did the independent use of each device in the animal study.
We have reported the performance of an experimental type of clinical SSGC that has been modified for better maneuverability during surgery. The size of the detector head of this experimental SSGC was greatly decreased by the architectural design of the configuration of the detector. The weight of the device was minimized to 1.4 kg (Fig. 7) (30,31).
SSGC designed for clinical use.
CONCLUSION
A prototype SSGC for surgical use was able to resolve the small amount of radioactivity in sentinel nodes in an animal study of sentinel lymph node biopsy. The mean radioactivity in imaged nodes was 0.15% of the initial injected volume of 99mTc-phytate (18.5 MBq). Furthermore, the SSGC clearly depicted increased uptake in the necks of patients with oral cancer at shorter acquisition times than did the CGC. These depictions were confirmed by a semiquantitative method (determination of HN/BG ratios) in both animal and clinical studies. We verified the efficacy of the SSGC for detecting a small amount of radioactivity in hot nodes during the procedure. The SSGC can offer clear visualization of sentinel nodes during surgery.