Basic Science Investigations |
1 Department of Radiology, University of Washington, Seattle, Washington
2 Pacific Northwest National Laboratory, Richland, Washington
3 Department of Medicine, University of Washington, Seattle, Washington
| ABSTRACT |
|---|
|
|
|---|
-camera images and whole-body NaI probe counts were obtained to estimate 131I-antibody source-organ residence times as recommended by the MIRD Committee. The source-organ residence times for standard man or woman were adjusted by the ratio of the MIRD phantom organ mass to the CT-derived organ mass. Results: The mean radiation absorbed doses (in mGy/MBq) for our data using the MIRD model were lungs = 1.67; liver = 1.03; kidneys = 1.08; spleen = 2.67; and whole body = 0.3; and for CT volume-adjusted organ volumes (in mGy/MBq) were lungs = 1.30; liver = 0.92; kidneys = 0.76; spleen = 1.40; and whole body = 0.22. We determined the following correlation coefficients between the 2 methods for the various organs: lungs, 0.49 (P = 0.0001); liver, 0.64 (P = 0.004); kidneys, 0.45 (P = 0.0004); spleen, 0.22 (P = 0.0001); and whole body, 0.78 (P = 0.0001), for the residence times. For therapy, patients received mean 131I administered activities of 19.2 GBq (520 mCi) after adjustment for CT-derived organ mass compared with 16.0 GBq (433 mCi) that would otherwise have been given had therapy been based only using standard MIRD organ volumesa statistically significant difference (P = 0.0001). Conclusion: We observed large variations in organ masses among our patients. Our treatments were planned to deliver the maximally tolerated radiation dose to the dose-limiting normal organ. This work provides a simplified method for calculating patient-specific radiation doses by adjusting for the actual organ mass and shows the value of this approach in treatment planning for RIT.
Key Words: non-Hodgkins lymphoma radioimmunotherapy internal dosimetry MIRD models patient-specific dosimetry
| INTRODUCTION |
|---|
|
|
|---|
Biodistribution of 131I-tositumomab in patients shows prolonged blood clearance and related long residence in the blood pool and normal organs, including liver, spleen, lungs, and kidneys. The average whole-body clearance half-time is approximately 90 h. Although the aim of RIT is to irradiate tumor, the normal organs and tissues receive significant radiation dose from circulating radioactivity. In keeping with other treatment modalities such as chemotherapy and external-beam radiotherapy, it is important to achieve the best possible therapeutic ratio, delivering the highest possible tumor doses with the lowest possible normal tissue toxicity. The need to minimize normal organ effects was recognized early in the development of RIT, and treatment-planning strategies were introduced to achieve this goal (813). This goal is analogous to the inverse planning techniques used in intensity-modulated radiotherapy for external-beam radiotherapy.
Internal dosimetry methods for systemically administered, unsealed radiation sources continue to evolve. Treatment planning for systemic radionuclides requires assessments of radioactivity in the major organs over time to calculate the integral residence times that are needed to estimate radiation absorbed doses. Accurate measurement of the uptake and retention of the radiopharmaceuticals in organs and tissues is challenging, but a reasonably accurate assessment of radionuclide pharmacokinetics is feasible with current imaging techniques. The MIRD formalism, originally introduced as a method for estimating radiation absorbed doses from diagnostic radiopharmaceuticals, has now been widely applied in radionuclide therapy (14,15).
Recognizing the wide variation in patient sizes, the MIRD Committee developed a set of nonscalable anthropomorphic models (phantoms) for use within the schema, including a standard adult (74 kg) and a 15 y old (or representative female, 58 kg). Each model has fixed organ masses. Though the MIRD formalism provides the ability to estimate the radiation absorbed doses to patients, it does not take into account variations in whole-body and organ masses that are seen in individual patients. The use of simple geometric representations of human organs and tissues, as given in the MIRD schema, can indeed lead to either underestimation or overestimation of the true absorbed dose to organs of individual patients (16).
It is desirable to plan high-dose radionuclide therapy using patient-specific organ volumes. Since the radiation absorbed dose to an organ is inversely proportional to its mass, it is important to optimize the dose to the individual patients organ to minimize systematic errors. Several adjustments to correct for differences in body and organ sizes and radionuclide kinetics have been introduced by others (15,17). Patient-specific approaches that account for variations in organ masses should enable more accurate dose estimates and, consequently, better protection against overirradiation that could lead to serious normal tissue toxicities.
Prior clinical and radiobiologic dose-escalation studies have established reasonable estimates of the maximum tolerable radiation absorbed doses to the major organs and tissues (57,13). RIT delivers relatively low initial dose rates (0.10.2 Gy/h to the tumor) compared with other radiotherapy modalities (424 Gy/h) (18). The radiobiology of internal emitters is therefore somewhat different than that of external-beam radiotherapy or brachytherapy. The toxicities associated with high-dose RIT extend beyond bone marrow suppression and include cardiopulmonary effects (19), hepatic toxicities (20,21), nephrotoxicity, and gastrointestinal tract effects.
During the last decade, we have treated >200 NHL patients using high-dose 131I-labeled RIT. This report describes a simple method for adjusting radiation absorbed dose estimates using actual organ volumes. Our study involved a cohort of patients with relapsed or residual NHL who received novel high-dose myeloablative RIT with 131I-tositumomab. We examined the results of this study to evaluate the benefits of CT-derived organ volumetrics and adjustments to the dose estimates compared with unadjusted assessments and standard MIRD organ models.
| MATERIALS AND METHODS |
|---|
|
|
|---|
Patients
All patients entering these protocols had documented B-cell NHL in relapse after standard therapy or primary refractory disease, had evaluable disease, and were eligible to receive autologous stem-cell rescue. Only patients with tumors expressing CD20 antigen were eligible to enter the study. Entry criteria for the protocol required patients to have an Eastern Cooperative Oncology Group performance status score of >1, normal renal and liver function, and <25% cancer involvement of the bone marrow. Patients also underwent standard evaluations for stem-cell rescue eligibility, including collection of adequate number of stem cells. This trial was performed with the approval of the human subjects and radiation safety committees at the University of Washington and the Fred Hutchinson Cancer Research Center, Seattle. All patients signed informed consent and were enrolled into the study according to protocol requirements.
Radioimmunoconjugate
Radioiodination of murine monoclonal anti-CD20 antibody (tositumomab, anti-B1, immunoglobulin G2a [IgG2a]; Corixa Pharmaceuticals Inc.) was performed at the radiochemistry facility of the Division of Nuclear Medicine at the University of Washington using the chloramine-T labeling method (6,8,9).
Biodistribution Studies for Treatment Planning
A biodistribution study was conducted on each patient before treatment using 185370 MBq (510 mCi) 131I-tositumomab antibody (1.7 mg/kg) diluted to 25 mL with normal saline and infused intravenously. Anterior and posterior
-camera images of the chest, abdomen, and pelvis were obtained after trace dose administration and then again at 48, 96, and 120 h after infusion. Measurements of whole-body activity were also obtained at the same time points. Timeactivity curves for each organ were constructed from these data. Organ residence times derived from the integrated timeactivity curves were used to estimate the radiation absorbed dose to organs and tissues of the body. For patients <60 y of age, the maximum normal organ dose was set at 27 Gy (2,700 rad) and for patients >60 y, the maximum normal organ dose was set at 25 Gy (2,500 rad) (6).
Imaging Measurements
-Camera images of the lungs (using 99mTc-labeled macroaggregated albumin), liver, spleen (99mTc-sulfur colloid), and kidneys (99mTc-MAG3) were obtained before the 131I-trace-labeled antibody infusion to delineate these organs for quantifying the 131I uptake and retention over time. Transmission images of the chest and abdomen were obtained using a 74-MBq (2 mCi) 131I-filled flood source, from which we calculated the appropriate attenuation-correction factors. After infusion, serial planar (anterior and posterior)
-camera images of the chest, abdomen, and pelvis were obtained at approximately 0, 48, 96, and 120 h (8). A high-energy collimator was used with a 20% window centered on the 131I
-peak at 364 keV. A General Electric Maxxus dual-head
-camera with a dedicated Starcam computer was used for image acquisition. Serial background-subtracted whole-body counts were obtained at the same times as the
-camera images using a shielded detection probe with a 7.6-cm (3-in.) NaI scintillation crystal interfaced to a multichannel analyzer (model 261; Ludlum Corp.). Anterior and posterior whole-body counts were obtained with the detector directed toward the full height and width of the patient (standing as a point source, at 5-m distance). The geometric mean of the anterior and posterior whole-body counts was used to calculate whole-body retention. Whole-body counts were corrected for physical decay using an 131I standard to obtain the percentage injected activity remaining at each time point.
99mTc-Derived regions of interest (ROIs) were placed over acquired 131I-tositumomab images to determine the 131I counts in liver, lungs, spleen, and kidneys and to generate the fractional timeactivity curves for these organs. Timeactivity curves were constructed for each major source organ, were fit to first- or second-order exponential functions using commercially available mathematic software (CurveExpert 1.3, Daniel Hymas; Sigma-Plot v. 4.0, Jandel Scientific), and were integrated. Residence times for lungs, liver, spleen, kidneys, bone, and total body were calculated from the individual organ retention of radioactivity obtained from ROIs drawn for these organs. Counts were corrected for background radioactivity using a region drawn in the peripheral soft tissue and were decay-corrected using an 131I standard aliquot.
The amount of 131I activity needed for therapy was calculated to deliver a fixed radiation absorbed dose to the critical normal organ (lungs, liver, or kidneys) receiving the highest dose.
Radiation Absorbed Dose Estimates
Standard procedures described by the MIRD Committee of the Society of Nuclear Medicine were used to estimate radiation absorbed dose to normal organs (14). Results were expressed in units of radiation absorbed dose per unit administered 131I activity (mGy/MBq or cGy/mCi) using standard dose-assessment software (MIRDOSE2; Oak Ridge Associated Universities, Oak Ridge, TN).
Before dose calculations, we adjusted the organ residence times from the integral of the serial timeactivity curves by applying the ratio (unitless) of the standard organ volume to the CT-derived volume for each major organ. The resulting adjusted residence time was then used to calculate the radiation absorbed dose (mGy/MBq) using the MIRDOSE2 software for each patient treatment plan.
Adjusting the residence time is a simplified method to account for patient-specific organ mass in the dose calculation. Briefly, for individual organs and for the whole body, this correction may be made by multiplying the calculated source-organ residence time,
h, by the ratio of the defined reference man or reference woman organ mass (mMIRD) to the known organ mass (mactual):
![]() |
Statistical Analysis
The common approach to dose assessment is to ignore patient organ masses and use MIRD models. Differences between use of MIRD and patient-specific organ masses (with and without adjustment for the actual organ volumes) were analyzed using a paired t test. Values expressing the significance of the differences between the 2 datasets were analyzed.
| RESULTS |
|---|
|
|
|---|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Internal dose assessment using patient-specific factors appears to be a useful approach in RIT, just as computerized dose assessment has been the mainstay of external-beam treatment planning. Similar concepts of treatment planning have been used with other RIT regimes (10,24,25). In our trials, we developed patient-specific estimation of absorbed doses to normal organs to improve the overall accuracy of treatment planning. Our treatment protocols required treatment at targeted dose levels. Other investigators have also recognized the importance of patient-specific absorbed dose assessment before high-dose RIT (5,9,26).
MIRDOSE2 has been used by this institution since 1989 to maintain consistency in the way radiation doses are calculated for lymphoma and leukemia patients.
Other strategies have been used to determine treatment levels. Strategies based on body weight or surface area, as with administration of cytotoxic chemotherapy, can result in inadequate treatment of tumor or overtreatment of sensitive normal tissues when doses are escalated. Radioactivity and cytotoxic chemotherapy do not have similar behavior in the normal, nontarget organs and tissues. Though many chemotherapeutic drugs need to be activated in the body before they are cytotoxic (such as cyclophosphamide) and their toxicity is organ specific, the blood-pool radioactivity in an organ during RIT can result in delivery of a significant radiation absorbed dose.
Whole-body measurements of retained radiolabeled antibody are used by others to calculate treatment dose (11). The whole-body measurement approach assumes (a) uniform biodistribution of the radiotracer in all organs and tissues, (b) identical residence times for the whole body and critical organs, and (c) all tissues exhibit essentially equivalent radiosensitivity. The present study, together with earlier reports from our group and those from others, have shown wide variations in individual biodistributions of radiolabeled antibody among patients with NHL. Our data also show large variations in patient size and organ mass. We believe that these differences prove the need for patient-specific dosimetry and treatment planning, particularly when high-dose, curative therapy is administered (5,7,911,13,25).
The relative radiosensitivity indication of the ability of cells to repair sublethal damage of normal tissues also is highly variable. Normal organ radiosensitivity limits the absorbed dose that may be safely delivered during radiation therapy. As we are better able to estimate radiation doses to healthy internal organs and tissues and the tolerable limits on radiation toxicity, we will be better able to use this information in treatment planning. Experience treating multiple myeloma with 166Ho-1,4,7,10-tetraazacyclododecane-1,4,7,10-tetramethylene-phosphonic acid (a skeletal targeted radiotherapy) (27) and neuroendocrine cancer with 90Y-octreotide (a radiolabeled peptide analog of somatostatin) (28,29) illustrates the challenge of normal organ toxicity (which in both cases were the kidneys, remote from the intended target tissues) in high-dose radionuclide therapy.
Actual patient weights and organ sizes vary considerably from those volumes described in the standard dosimetry models. For example, spleen sizes in adult patients have varied from about 50 to >600 g, whereas the spleen size in the standard MIRD adult phantom is 183 g. Since radiation absorbed dose to an organ is inversely proportional to the mass of the organ for which dose is calculated, an adjustment to dose estimation should be made for organ mass when actual organ volumes are known from CT imaging. In this article, we describe a convenient method for correcting for known organ mass (15). This simplified approach corrects the ß-component (about 93% of the absorbed dose from 131I) but does not correct the
-component (<7% of the total). For most radionuclides, the ß self-irradiation dose in a source organ is the greater contributor to total organ dose (usually >90% of the total). This correction is accurate for pure ß-particleemitting radionuclides, such as 90Y and 32P. The simplified method does not correct for
-emitting radionuclides with very little ß-particle energy imparted to tissue but should be useful for most ß- and
-emitting radionuclides (such as 131I and 186Re). It is practical, yet accurate, compared with alternative methods for patient-specific dose assessment that require complete recalculation of S values for each source
target combination using patient-specific organ parameters. In practice, the use of CT images to display organ size and geometry combined with planar imaging or SPECT to display activity in the major organs can be used to calculate patient-specific S values for each source
target organ combination. However, recalculation of voxel S values can be a time-intensive and computer-intensive process that is not practical for most treatment-planning obligations.
For our study, we obtained organ volumes from CT scans of patients chest and abdomen to correct the source-organ residence times. Patient-specific residence times enable more accurate dose calculations for the major organs. The CT scan is a well-established method for estimating organ volumes (30,31).
| CONCLUSION |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
For correspondence or reprints contact: Joseph G. Rajendran, MD, Division of Nuclear Medicine, Box 356113, University of Washington, Seattle, WA 98195.
E-mail: rajan{at}u.washington.edu
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
J. N. Winter, D. J. Inwards, S. Spies, G. Wiseman, D. Patton, W. Erwin, A. W. Rademaker, B. B. Weitner, S. F. Williams, M. S. Tallman, et al. Yttrium-90 Ibritumomab Tiuxetan Doses Calculated to Deliver up to 15 Gy to Critical Organs May Be Safely Combined With High-Dose BEAM and Autologous Transplantation in Relapsed or Refractory B-Cell Non-Hodgkin's Lymphoma J. Clin. Oncol., April 1, 2009; 27(10): 1653 - 1659. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Buck, S. Nekolla, S. Ziegler, A. Beer, B. J. Krause, K. Herrmann, K. Scheidhauer, H.-J. Wester, E. J. Rummeny, M. Schwaiger, et al. SPECT/CT J. Nucl. Med., August 1, 2008; 49(8): 1305 - 1319. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Rajendran, A. K. Gopal, D. R. Fisher, L. D. Durack, T. A. Gooley, and O. W. Press Myeloablative 131I-Tositumomab Radioimmunotherapy in Treating Non-Hodgkin's Lymphoma: Comparison of Dosimetry Based on Whole-Body Retention and Dose to Critical Organ Receiving the Highest Dose J. Nucl. Med., May 1, 2008; 49(5): 837 - 844. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Cremonesi, M. Ferrari, C. M. Grana, A. Vanazzi, M. Stabin, M. Bartolomei, S. Papi, G. Prisco, G. Martinelli, G. Paganelli, et al. High-Dose Radioimmunotherapy with 90Y-Ibritumomab Tiuxetan: Comparative Dosimetric Study for Tailored Treatment J. Nucl. Med., November 1, 2007; 48(11): 1871 - 1879. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Mah and K. M. Spicer Comparison of Medium- and High-Energy Collimators for 131I-Tositumomab Dosimetry J. Nucl. Med. Technol., September 1, 2007; 35(3): 148 - 153. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. K. Gopal, J. G. Rajendran, T. A. Gooley, J. M. Pagel, D. R. Fisher, S. H. Petersdorf, D. G. Maloney, J. F. Eary, F. R. Appelbaum, and O. W. Press High-Dose [131I]Tositumomab (anti-CD20) Radioimmunotherapy and Autologous Hematopoietic Stem-Cell Transplantation for Adults >= 60 Years Old With Relapsed or Refractory B-Cell Lymphoma J. Clin. Oncol., April 10, 2007; 25(11): 1396 - 1402. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. K. Tan, R. W. Wassenaar, and W. Zeng Collimator Selection, Acquisition Speed, and Visual Assessment of 131I-Tositumomab Biodistribution in a Phantom Model J. Nucl. Med. Technol., December 1, 2006; 34(4): 224 - 227. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||