|
|
||||||||
CLINICAL INVESTIGATIONS |
Department of Radiology, Emory University School of Medicine, Atlanta, Georgia
| ABSTRACT |
|---|
|
|
|---|
Key Words: 2D PET 3D PET 82Rb heart
| INTRODUCTION |
|---|
|
|
|---|
An advantage of 3D mode is that collecting many more lines of response permits injecting far less radioactivity, which has the dual benefit of reducing the radiation exposure to the patient and reducing the amount of 82Rb needed. Because the counting rate in 3D mode is several times greater than that in 2D mode for equal radioactivity in the field of view, potentially less than half as much 82Rb could be used. The cost of a 82Rb generator is determined primarily by the amount of 82Sr (parent isotope of 82Rb) that the generator contains, so the cost to obtain a 3D 82Rb scan could be substantially less than that of a 2D scan. This may lead to cheaper and more widespread use of myocardial imaging with 82Rb.
Disadvantages of 3D scanning are the counting rate limit of the scanner, increased random events, and increased scatter. The counting rate in 2D mode after injecting 1480 MBq 82Rb approaches the limit of the ECAT EXACT 921 scanner (CTI/Siemens, Knoxville, TN). In 3D mode the radioactivity must be scaled back so as not to saturate the camera. When the radioactivity is scaled back so that the scanner is counting at near its maximal rate in 3D mode, the number of detected true events may not be much more than that of the higher dose 2D study. Therefore, the potential benefit of collecting more data in a 3D study may be small.
One limitation of scanning in a high counting rate environment is that the random counting rate increases much more rapidly than does the true counting rate as a function of radioactivity in and near the field of view. The random counting rate is the product of the counting rates in each of the 2 responding detectors and the coincidence time window (5). Hence, whereas the true counting rate scales linearly with radioactivity in the field of view, the random rate scales as the square of radioactivity in and near the field of view. The 3D case is worse because removal of the septa allows the detectors to view radioactivity further outside the field of view so more radioactivity contributes to the background rate. Because of this, the counting rate for which the true and random counting rates are equal is generally less in 3D scanning than in 2D scanning. Above this rate, the detected events are dominated by random events.
When the septa are removed, each individual detector is sensitive to radiations from a much larger area. For a given total counting rate, the fraction of random events recorded will be greater when scanning in 3D mode (6). The number of random events can be estimated and subtracted from the dataset. Comparing 3D with 2D scanning at the scanners maximal counting rate, the number of detected true events in 3D scanning will be less because the rate of detecting prompt events (true events + random events + scatter) is limited by the maximal counting rate of the scanner, and uncertainty in the estimate of random events will have a greater effect on the noise in the final 3D images. These effects can be quantified by calculating noise equivalent counting rates (NECs) (7). DeGrado et al. (8) have shown that the peak NEC is greater in 2D than in 3D scanning for a uniform phantom scanned with an Advance scanner (a Bi4Ge3O12 [BGO]-based scanner similar to the one we used) (General Electric Medical Systems, Milwaukee, WI). Badawi et al. (9) showed that the peak NEC and relative performance between 2D and 3D scanning on a 951R scanner (CTI/Siemens) are dependent on the geometry of the object being scanned and the amount of radioactivity in the field of view. These works suggest that 2D scanning will be superior in high counting rate scanning, but it is necessary to perform a direct comparison between 2D and 3D scanning for the current scanner and cardiac application with 82Rb.
In 3D mode, the number of scattered events approaches half of all recorded events (6). Because of this, for proper quantitation, a scatter correction must be applied to the 3D data. A large literature relating to 3D scatter correction methods exists (1012). Some of these methods, particularly the dual-energy window technique, should be applicable to cardiac imaging but none has been specifically tested in the very high counting rate situation of 82Rb cardiac imaging.
The authors of 2 reports have concluded that, under the proper conditions, 3D scanning is advantageous for high counting rate H215O neuroactivation studies (13,14). However, large differences between the size of the object in the field of view and the distribution of background radioactivity make inferring 3D performance for the current cardiac application difficult. This study was initiated to determine if the potential benefit of increasing the number of true counts in 3D mode would offset the detriment of increased random and scattered events. Of primary interest is to determine whether the same clinical information can be obtained from a 3D protocol as from the standard 2D protocol.
| MATERIALS AND METHODS |
|---|
|
|
|---|
All images were reconstructed with segmented attenuation correction derived from a 7-min measured transmission scan (15). Images were reconstructed with filtered backprojection using a Hahn smoothing filter cutoff at 1 cycle per centimeter. The recommended scatter correction supplied by the scanner manufacturer (CTI/Siemens software, version 7.1b) was used. The final reconstructed resolution was measured by scanning a 82Rb line source (5 cm off center) in a water-scattering medium and found to be 12.3 mm tangentially and 15.3 mm radially.
Thirty-three consecutive patients who were scheduled for resting 82Rb scans at the Nuclear Medicine Department were studied. Our standard 82Rb protocol calls for a preview scan for positioning the patient in the scanner, followed by the resting 82Rb scan, and then a transmission scan. Because the preview scan can be done equally well in either 2D or 3D mode, the preview scan was collected in 3D mode so that it could be compared with the resting scan collected in 2D mode. Between 1300 and 1850 MBq were injected for the resting scan (depending on the age of the Rb generator). Nineteen patients were injected with 740 MBq and 14 patients were injected with 370 MBq for the 3D preview scan. The duration of each scan on each patient was 7 min.
After reconstruction, all images were oriented into short-axis views and then loaded into the Emory Cardiac Toolbox (Emory University, Atlanta, GA). The toolbox was used to create polar maps of the 2D and 3D scans, which were displayed side by side for comparison. The 2D scan was taken as the gold standard. Four physicians who were experienced in reading 82Rb studies reviewed the polar maps. The display included the raw polar map data and a blackout image in which all pixels >2.5 SDs from the mean were blacked out. The blackout was relative to the normal 82Rb database included in the Emory Cardiac Toolbox. The physicians were asked 3 questions: Is there a defect in either or both of the polar maps? Would the findings in your dictation be different between the 2 scans? Would your impression and recommendations for future patient care be different? All physicians viewed the images independently. No attempts were made to reconcile differing opinions.
Phantom studies were performed to simulate the human 82Rb studies. The thoracic phantom (Data Spectrum Corp., Hillsborough, NC) with the cardiac insert and lungs filled with Styrofoam beads (Dow Chemical Co., Midland, MI) was used. Regions of interest were drawn on the 2D images from a healthy subject who was administered 1480 MBq 82Rb to determine typical radioactivity concentrations for filling the phantom. The total system counting rates for true and random events were also noted from the human study. The quantity of radioactivity added to the background was estimated to be that necessary to make the true counting rates match in the human and the phantom experiments. In addition, to account for sources out of the field of view, two 3-L cylinders were filled with radioactive solution and placed on the scanner bed in roughly the location of the head and the bladder. The amount of radioactivity in these containers was intended to produce the same random counting rate in the phantom studies as in the human study.
The beginning of the first experiment was delayed, which caused the measured region-of-interest values to undershoot the desired radioactivity levels as well as the number of recorded true and random events. The experiment was also suboptimal in 2 other areas: The measured radioactivity in the cardiac wall was less than the true radioactivity because of the partial-volume effect (16,17), and the radioactivity concentration was reduced more than expected because of the Styrofoam beads in the lungs. Because of these limitations, a second experiment was performed with the radioactivity in the phantom chambers adjusted accordingly. The number of detected true events was adjusted by adding additional radioactivity to the background of the phantom. Moving the head and body cylinders closer to the phantom increased the number of detected random events. The radioactivity concentrations and other parameters for the 2 experiments are detailed in Table 1.
|
After reconstruction, the images were analyzed quantitatively by evenly spacing 100 points over the defect-free portion of the myocardial wall. Each point was approximately 1 cm from its nearest neighbor. Because the true radioactivity concentration at these points was equal, the mean divided by the SD of the measured values is the signal-to-noise ratio (SNR) in the images. This measure takes into account all sources of noise.
| RESULTS |
|---|
|
|
|---|
The physicians evaluations are summarized in Table 2. On average, their impressions differed between the 2D and 3D studies in 53% of the cases. The most noticeable difference between the 3D and 2D studies was decreased uptake in the apex area on the 3D scans. Defects generally appeared much larger on the 3D scans. No significant difference in physicians attitudes was found between injecting 370 or 740 MBq when scanning in the 3D mode. The physicians recommended, on average, different follow-up care for 10 of the 33 patients. For 8 patients, a consensus of the physicians (3 or 4 in agreement) believed that subsequent treatment would be different. Three of the 4 physicians did not believe that the 3D studies could be used as a replacement for the 2D studies without further investigations. Transaxial images of representative cases are shown in Figure 1, and the corresponding polar maps are shown in Figure 2.
|
|
|
|
|
|
|
| DISCUSSION |
|---|
|
|
|---|
Because the scanner is near the saturation limit, injecting a greater amount of 82Rb would be detrimental to image quality, and the potential advantage of switching to 3D scanning to collect additional counts is not supported. The situation would be improved if the random counting rate were decreased. The coincidence window is 12 nS for the scanner used in this study. If this window could be reduced to 6 nS, the number of recorded random events would be cut in half and the number of prompt events would be reduced as well. This would improve images from both scanning modes, but the improvement would be greater for the 3D acquisitions because cutting the random events in half would reduce the 3D total counting rate by 40% and the 2D rate by 28%. It is not clear which acquisition method would be preferable in this case.
In addition to greater noise in the 3D studies, half of the images were read as being significantly different. The primary difference was a new defect, or a larger defect, at the apex on the 3D scans. The image quality at the edge of the thorax is particularly bad in the 3D studies, as is seen in Figure 1. The reason for this is not known but a possibility is the scatter correction. The scatter correction algorithm relies on the estimated emission and attenuation images. If these images have artifacts, then the scatter correction may be erroneous and lead to strong lateral streak artifacts (18). The situation may be improved by iterating the reconstruction but this was not tried. In any case, the unacceptable noise level on the 3D images renders investigation of 3D scatter correction moot for this project.
| CONCLUSION |
|---|
|
|
|---|
This study supports a general rule of thumb for PET scanning. If there are sufficient counts to perform a study in 2D mode, then that study should be performed in 2D mode. Three-dimensional scanning is advantageous when the counting rate is much lower and the benefit from additional events far outweighs any loss attributed to increasing the fraction of random and scatter events. This may be (and often is) the case when the amount of injected radioactivity is limited for nonscanner-related reasons (e.g., radiation safety concerns). However, this is not the case with 82Rb and the ECAT EXACT scanner. We recommend that 82Rb scanning continue to be performed in 2D mode until improvements in the scanner allow greater counting rate or a method is developed to shield radioactivity outside the field of view (or both).
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
For correspondence or reprints contact: John R. Votaw, PhD, Department of Radiology, Emory University School of Medicine, 1364 Clifton Rd., N.E., Atlanta, GA 30322.
| REFERENCES |
|---|
|
|
|---|
This article has been cited by other articles:
![]() |
M. Ibaraki, S. Miura, E. Shimosegawa, S. Sugawara, T. Mizuta, A. Ishikawa, and M. Amano Quantification of Cerebral Blood Flow and Oxygen Metabolism with 3-Dimensional PET and 15O: Validation by Comparison with 2-Dimensional PET J. Nucl. Med., January 1, 2008; 49(1): 50 - 59. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. Schepis, O. Gaemperli, V. Treyer, I. Valenta, C. Burger, P. Koepfli, M. Namdar, I. Adachi, H. Alkadhi, and P. A. Kaufmann Absolute Quantification of Myocardial Blood Flow with 13N-Ammonia and 3-Dimensional PET J. Nucl. Med., November 1, 2007; 48(11): 1783 - 1789. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. A. Lodge, R. D. Badawi, R. Gilbert, P. E. Dibos, and B. R. Line Comparison of 2-Dimensional and 3-Dimensional Acquisition for 18F-FDG PET Oncology Studies Performed on an LSO-Based Scanner J. Nucl. Med., January 1, 2006; 47(1): 23 - 31. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Schwaiger, S. Ziegler, and S. G. Nekolla PET/CT: Challenge for Nuclear Cardiology J. Nucl. Med., October 1, 2005; 46(10): 1664 - 1678. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Brogsitter, T. Gruning, R. Weise, P. Wielepp, O. Lindner, R. Korfer, and W. Burchert 18F-FDG PET for Detecting Myocardial Viability: Validation of 3D Data Acquisition J. Nucl. Med., January 1, 2005; 46(1): 19 - 24. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Lubberink, R. Boellaard, A. P. van der Weerdt, F. C. Visser, and A. A. Lammertsma Quantitative Comparison of Analytic and Iterative Reconstruction Methods in 2- and 3-Dimensional Dynamic Cardiac 18F-FDG PET J. Nucl. Med., December 1, 2004; 45(12): 2008 - 2015. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Knesaurek, J. Machac, B. R. Krynyckyi, and O. D. Almeida Comparison of 2-Dimensional and 3-Dimensional 82Rb Myocardial Perfusion PET Imaging J. Nucl. Med., August 1, 2003; 44(8): 1350 - 1356. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Knesaurek New Developments in PET Instrumentation: Quo Vadis PET? J. Nucl. Med., December 1, 2001; 42(12): 1831 - 1832. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY | THE JOURNAL OF NUCLEAR MEDICINE |