Skip to main content

Main menu

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI

User menu

  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart

Search

  • Advanced search
Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • View or Listen to JNM Podcast
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
Meeting ReportPoster - Technologist

Optimization of Injected Dose for Myocardial Flow Quantification in 13N ammonia PET with Time of Flight Scanner. Noise Equivalent Count Rate analysis

Yoko Kaimoto
Journal of Nuclear Medicine May 2021, 62 (supplement 1) 3033;
Yoko Kaimoto
1Tokyo Women's Medical University Tokyo Japan
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
  • Article
  • Figures & Data
  • Info & Metrics
Loading

Abstract

3033

Introduction: Myocardial blood flow quantification (MBF) is one of the distinctive features for cardiac PET. It is commonly known that the MBF is not influenced by injected dose because the input function (IF) is obtained from time activity curve (TAC). However, there is a substantial risk of count-loss when obtaining TAC because high radioactivity of perfusion tracer goes through left-ventricular cavity sized 10-40mL within a few minutes. We aimed to determine the optimal dose for MBF using noise equivalent count rate analysis (NECR).

Methods: We used a high resolution, time of flight scanner with LYSO crystal (Biograph mCT, Siemens) for phantom study. 10mL of syringe was placed in thyroid uptake neck phantom to mimic end-systolic left ventricular cavity. 1500MBq of 13N was filled in the syringe, and a set of single session with 3D list-mode scan for 120sec (60sec acquisition followed by 60sec stand-by) was repeated 31 times to simulate the acquisition with injected dose of 1500 down to 20MBq. This session was repeated for 6 days (6 sessions with 186 scans in total) to obtain mean value. Radioactivity (MBq/mL) was measured by drawing volume of interest on the 3D-OSEM reconstruct images. Random and net true count rate were calculated and compared to the estimated injected doses. NECR curve analysis was used to determine the optimal injected dose.

Results: The attenuation curves showed best correlation at dose of 185, 370 and 740 MBq (r = 1.0, 1.0, and 1.0; p <0.001, <0.001, and <0.001, respectively). NECR analysis revealed that the random counts were significantly suppressed for the injected dose with 22 to 40 MBq/mL, while the count rate started to decrease over 40MBq/mL (figure). The optimized injected dose was determined as 222 to 555MBq.

Conclusions: Simulative analysis for 13N derived MBF using cylindrical phantom mimicking end-systolic left ventricular cavity confirmed that the risk of count-loss was minimized at the recommended dose in guideline. This result confirmed the feasibility of myocardial flow quantification and can be supportive data in clinical routine.

Figure
  • Download figure
  • Open in new tab
  • Download powerpoint
Previous
Back to top

In this issue

Journal of Nuclear Medicine
Vol. 62, Issue supplement 1
May 1, 2021
  • Table of Contents
  • Index by author
Article Alerts
Sign In to Email Alerts with your Email Address
Email Article

Thank you for your interest in spreading the word on Journal of Nuclear Medicine.

NOTE: We only request your email address so that the person you are recommending the page to knows that you wanted them to see it, and that it is not junk mail. We do not capture any email address.

Enter multiple addresses on separate lines or separate them with commas.
Optimization of Injected Dose for Myocardial Flow Quantification in 13N ammonia PET with Time of Flight Scanner. Noise Equivalent Count Rate analysis
(Your Name) has sent you a message from Journal of Nuclear Medicine
(Your Name) thought you would like to see the Journal of Nuclear Medicine web site.
Citation Tools
Optimization of Injected Dose for Myocardial Flow Quantification in 13N ammonia PET with Time of Flight Scanner. Noise Equivalent Count Rate analysis
Yoko Kaimoto
Journal of Nuclear Medicine May 2021, 62 (supplement 1) 3033;

Citation Manager Formats

  • BibTeX
  • Bookends
  • EasyBib
  • EndNote (tagged)
  • EndNote 8 (xml)
  • Medlars
  • Mendeley
  • Papers
  • RefWorks Tagged
  • Ref Manager
  • RIS
  • Zotero
Share
Optimization of Injected Dose for Myocardial Flow Quantification in 13N ammonia PET with Time of Flight Scanner. Noise Equivalent Count Rate analysis
Yoko Kaimoto
Journal of Nuclear Medicine May 2021, 62 (supplement 1) 3033;
Twitter logo Facebook logo LinkedIn logo Mendeley logo
  • Tweet Widget
  • Facebook Like
  • Google Plus One
Bookmark this article

Jump to section

  • Article
  • Figures & Data
  • Info & Metrics

Related Articles

  • No related articles found.
  • Google Scholar

Cited By...

  • No citing articles found.
  • Google Scholar

More in this TOC Section

Poster - Technologist

  • Development of a novel phantom to evaluate the characteristics of dual-isotope simultaneous acquisition in a small-animal SPECT-PET/CT system
  • Seeing the big picture: The importance of reviewing the entire field of view in Myocardial Perfusion Imaging and the role of the Nuclear Medicine Technologist
  • Impacts of improved TOF timing resolutions on cold contrast of PET images
Show more Poster - Technologist

Technologist Poster Session

  • Development of a novel phantom to evaluate the characteristics of dual-isotope simultaneous acquisition in a small-animal SPECT-PET/CT system
  • Seeing the big picture: The importance of reviewing the entire field of view in Myocardial Perfusion Imaging and the role of the Nuclear Medicine Technologist
  • Impacts of improved TOF timing resolutions on cold contrast of PET images
Show more Technologist Poster Session

Similar Articles

SNMMI

© 2025 SNMMI

Powered by HighWire