Skip to main content

Main menu

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI

User menu

  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart

Search

  • Advanced search
Journal of Nuclear Medicine
  • SNMMI
    • JNM
    • JNMT
    • SNMMI Journals
    • SNMMI
  • Subscribe
  • My alerts
  • Log in
  • Log out
  • My Cart
Journal of Nuclear Medicine

Advanced Search

  • Home
  • Content
    • Current
    • Ahead of print
    • Past Issues
    • JNM Supplement
    • SNMMI Annual Meeting Abstracts
    • Continuing Education
    • JNM Podcasts
  • Subscriptions
    • Subscribers
    • Institutional and Non-member
    • Rates
    • Journal Claims
    • Corporate & Special Sales
  • Authors
    • Submit to JNM
    • Information for Authors
    • Assignment of Copyright
    • AQARA requirements
  • Info
    • Reviewers
    • Permissions
    • Advertisers
  • About
    • About Us
    • Editorial Board
    • Contact Information
  • More
    • Alerts
    • Feedback
    • Help
    • SNMMI Journals
  • View or Listen to JNM Podcast
  • Visit JNM on Facebook
  • Join JNM on LinkedIn
  • Follow JNM on Twitter
  • Subscribe to our RSS feeds
Research ArticlePhysics and Instrumentation

A Multicenter Study on Observed Discrepancies Between Vendor-Stated and PET-Measured 90Y Activities for Both Glass and Resin Microsphere Devices

Silvano Gnesin, Justin K. Mikell, Maurizio Conti, John O. Prior, Thomas Carlier, Thiago V.M. Lima and Yuni K. Dewaraja
Journal of Nuclear Medicine May 2023, 64 (5) 825-828; DOI: https://doi.org/10.2967/jnumed.122.264458
Silvano Gnesin
1Institute of Radiation Physics, Lausanne University Hospital, University of Lausanne, Lausanne, Switzerland;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Justin K. Mikell
2Department of Radiation Oncology, University of Michigan, Ann Arbor, Michigan, and Department of Radiation Oncology, Washington University, St. Louis, Missouri;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Maurizio Conti
3Department of Molecular Imaging, Siemens Medical Solutions, Knoxville, Tennessee;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
John O. Prior
4Department of Nuclear Medicine and Molecular Imaging, Lausanne University Hospital, University of Lausanne, Lausanne, Switzerland;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Thomas Carlier
5Nuclear Medicine Department, University Hospital of Nantes, Nantes, France;
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Thiago V.M. Lima
6Department of Radiology and Nuclear Medicine, Luzerner Kantonsspital, Luzern, Switzerland; and
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site
Yuni K. Dewaraja
7Division of Nuclear Medicine, Department of Radiology, University of Michigan, Ann Arbor, Michigan
  • Find this author on Google Scholar
  • Find this author on PubMed
  • Search for this author on this site

Visual Abstract

Figure
  • Download figure
  • Open in new tab
  • Download powerpoint

Abstract

Dosimetry-guided treatment planning in selective internal radiation therapy relies on accurate and reproducible measurement of administered activity. This 4-center, 5-PET-device study compared the manufacturer-declared 90Y activity in vials with quantitative 90Y PET/CT assessment of the same vials. We compared 90Y PET-measured activity (APET) for 56 90Y-labeled glass and 18 90Y-labeled resin microsphere vials with the calibrated activity specified by the manufacturer (AM). Additionally, the same analysis was performed for 4 90Y-chloride vials. The mean APET/AM ratio was 0.79 ± 0.04 (range, 0.71–0.89) for glass microspheres and 1.15 ± 0.06 (range, 1.05–1.25) for resin microspheres. The mean APET/AM ratio for 90Y-chloride vials was 1.00 ± 0.04 (range, 0.96–1.06). Thus, we found an average difference of 46% between glass and resin microsphere activity calibrations, whereas close agreement was found for chloride solutions. We expect that the reported discrepancies will promote further investigations to establish reliable and accurate patient dosimetry and dose–effect assessments.

  • resin microspheres
  • glass microspheres
  • 90Y
  • PET/CT
  • activity

Selective internal radiation therapy (SIRT) with radioactive microspheres is an established liver-directed therapy for both primary liver cancer and liver metastases. Both 90Y glass and resin microspheres are used globally; they are Food and Drug Administration–approved in the United States, and they received the CE (Conformitè Europëenne) mark in the European Union.

Considerable evidence of dose–effect relationships for both tumor and nontumor liver have been demonstrated for 90Y SIRT (1). In particular, for glass microspheres, Garin et al. highlighted a dose–response relationship in a prospective randomized trial, which demonstrated that planned personalized dosimetry improves outcomes compared with standard single-compartment dosimetry for locally advanced hepatocellular carcinoma (2). Another area of investigation is focused on posttreatment SIRT dosimetry, which obviates most of the difficulties linked to the hypothesis that pretreatment imaging-based dosimetry is a robust surrogate of the actual delivered absorbed dose. In that respect, recent studies have suggested the benefit of 90Y PET–based dosimetry in hepatocellular carcinoma or cholangiocarcinoma (3,4).

For reliable dosimetry-guided treatment planning and dose–effect assessment from pretherapy imaging–based absorbed dose estimates, the net administered activity of 90Y microspheres should be accurately determined. Accurate assay of 90Y, an almost pure β-emitter, using activity meters is challenging compared with other radionuclides commonly used in nuclear medicine procedures (5). (In this work, we adopted the term activity meter for the reentrant well-type ionization chamber that is calibrated to convert a measured ionization current to an activity; this device is also colloquially referred to as a dose calibrator in North America. We reserved the use of dose for the absorbed dose in units of Gy.) The specific geometry and material composition of the source and its container affects the spectrum of Bremsstrahlung photons, hence affecting the activity meter measurement.

90Y PET imaging is also challenging because of the low true coincidence count rates associated with the low yield of positron emission (0.0032%). Despite this, there have been multiple reports demonstrating the quantification accuracy of 90Y PET in phantom studies when using state-of-the-art time-of-flight scanners (6–8). It should be noted that almost all phantom studies to date have used 90Y in the form of a chloride solution and not the microsphere devices themselves. Reasons may include the difficulty of suspending microspheres in a uniform distribution throughout a phantom compartment.

In this work, we used quantitative 90Y PET/CT imaging to measure the 90Y microsphere vial activity in air before SIRT with resin and glass microspheres at 4 institutions on 5 scanners: PET-1, a Biograph Vision 600 (Siemens Healthineers) at Lausanne University Hospital (Centre hospitalier universitaire vaudois [CHUV]); PET-2, a Discovery 690 (GE Healthcare) at CHUV; PET-3, a Biograph mCT 40 (Siemens Healthineers) at the University of Michigan; PET-4, a Biograph Vision 600 (Siemens Healthineers) at Luzerner Kantonsspital; and PET-5, a Biograph 40 mCT (Siemens Healthineers) at the University Hospital of Nantes. Additional data came from PET measurements of vials containing 90Y in chloride solution and as a liquefied resin. We compared the PET-measured activity with the activity on the calibration certificate supplied by the vendor for each vial with appropriate decay correction.

MATERIALS AND METHODS

We analyzed 90Y vials from 3 different manufacturers: 56 90Y-labeled glass microsphere vials (TheraSphere; Boston Scientific), 18 90Y-labeled resin microsphere vials (SIR-Spheres; Sirtex Medical), 4 vials containing 90Y-chloride solution (2 from Curium and 2 from Eckert and Ziegler), and 1 vial containing a solution of dissolved 90Y-labeled resin microspheres (liquefied resin). All vials were imaged in air at a single bed position centered on the 3 tomographic directions of the PET scanner to yield peak sensitivity.

Glass Microspheres

A first dataset of 43 90Y-labeled glass microsphere vials (0.7–6.3 GBq) was imaged on PET-1, with a subgroup of 8 of these vials (0.7–6.3 GBq) being additionally measured on PET-2. Another dataset of 13 glass microsphere vials (2.3–8.6 GBq) was imaged on PET-3.

Resin Microspheres

90Y-labeled resin microsphere vials (3.3–4.6 GBq) were imaged on PET-1 (n = 11), PET-2 (n = 1), or PET-4 (n = 6).

Chloride Solution

In addition to performing acquisitions on microsphere vials, we acquired PET/CT data for vials of liquid 90Y-chloride solution. Of these, 2 (0.4 and 2.5 GBq) were acquired on PET-1 and 2 (0.4 and 4.4 GBq) on PET-3.

Liquified Resin Microspheres

The vial with the solution of dissolved 90Y-labeled resin microspheres was from a prior study evaluating the reliability of measuring 90Y activity using PET performed at the University Hospital of Nantes in collaboration with the French National Standard Laboratory in Paris (LNHB [Laboratoire National Henri Becquerel]–CEA [Commissariat à l’énergie atomique et aux énergies alternatives]). For this purpose, a reference activity of resin microspheres (2.95 GBq) was first dissolved (9), measured using the triple- to double-coincidence ratio method with Cherenkov counting at LNHB-CEA, and then shipped to the University Hospital of Nantes for a PET/CT acquisition on PET-5.

PET/CT Reconstruction and Quantification

Supplemental Table 1 summarizes acquisition and reconstruction parameters for the different PET/CT devices (supplemental materials are available at http://jnm.snmjournals.org). Manufacturer-recommended reconstruction parameters were used for 90Y. Considering the measurements on PET-1 and PET-4 (the 2 Biograph Vision 600 devices), we tested both absolute and relative scatter corrections available with the manufacturer software. The relatively high noise associated with low-count 90Y PET can affect the scatter correction with relative scaling, and some bias can appear in the final quantification as reported previously (6). Since the preliminary quantitative assessment showed, as expected, no significant difference between the 2 scatter methods in the low-scatter setting (in air) of the current experiment (Supplemental Tables 2 and 3), we focus on results obtained with the absolute scatter correction.

In all PET scanners used in the current study, the software enables 90Y quantitation automatically from the local 18F system cross-calibration, accounting for the 90Y specific physical decay and positron branching ratio. The quantitative PET data were decay-corrected to the start of the PET acquisition. On the PET images, we defined cylindric (50 mm in diameter, 5 cm high) volumes of interest that encompassed the vials (diameters of 25, 35, and 25 mm for resin, glass, and 90Y-chloride, respectively), to minimize any signal loss due to partial-volume effects.

Manufacturer-Specified Activity Assessment

The manufacturer-specified calibrated activity was reported in a document shipped with the vial to the different hospitals. In this document, the manufacturer indicates the vial activity and the time of the calibration. After performing the manufacturer-specified procedure for establishing a local calibration factor, we routinely verified the activity by measuring the received vial in the local activity meter. Specifically, nominal manufacturer activity was used for resin microspheres, whereas the manufacturer-measured total activity (not nominal) was used for glass microspheres. The local versus certified manufacture activity was found to be within 5% at all centers.

Comparison of PET-Derived Activity with Manufacturer-Specified Value

For each measurement, the total PET activity measured in the vial volume of interest (APET) was compared with the vial activity reported in the manufacturer calibration sheet (AM) decay-corrected to the start of the PET acquisition, using the ratio APET/AM. We assessed for statistical differences in APET/AM ratios for the same microsphere type obtained in different PET scanners by applying ANOVA and multiple-comparison tests using the MATLAB statistical toolbox (version R2021a; MathWorks). A significant difference was considered present for P values of less than 0.05.

RESULTS

Figure 1 shows a dot-plot representation of the APET/AM distribution across all 4 vial types and all centers. Table 1 presents the summary statistics for all measurements (full data are available in Supplemental Presentation 1). The mean APET/AM ratio for 90Y glass spheres was 0.79 ± 0.04 (range, 0.71–0.89). No statistical differences in mean APET/AM for 90Y glass spheres were found between PET-1 and PET-3 (P = 0.43). Statistical differences were found between PET-1 and PET-2 (P = 0.009) and between PET-2 and PET-3 (P = 0.002). The mean APET/AM ratio for the resin spheres was 1.15 ± 0.06 (range, 1.05–1.25). In this case, no statistical difference was found between PET-1 and PET-4 (P = 0.072).

FIGURE 1.
  • Download figure
  • Open in new tab
  • Download powerpoint
FIGURE 1.

Distribution of APET/AM ratios for the 4 vial products tested in this study (i.e., 90Y-labeled glass microspheres, 90Y-chloride solution, 90Y-labeled resin microspheres, and 90Y-labeled liquefied resin). LUKS = Luzerner Kantonsspital, the 90Y-labeled liquefied resin (Nantes PET-5, purple dot) was associated to the resin’s category.

View this table:
  • View inline
  • View popup
TABLE 1.

Summary of APET/AM Results

The mean APET/AM measured in 90Y-chloride vials was 1.00 ± 0.04 (range, 0.96–1.06) (Table 1; Supplemental Table 4). The APET/AM measured for the liquefied resin spheres in PET-5 was 1.22 ± 0.12, whereas good agreement was found between the LNHB-CEA reference activity and the PET activity measurements, with a ratio of 1.01.

DISCUSSION

In this study, we used PET as an independent measure of activity for 90Y microsphere vials in air and compared this measure with the activity reported in the respective manufacturer’s calibration sheet for 90Y-labeled resin and glass microspheres and 90Y in chloride solution and liquefied resin. Although we report substantial discrepancies for resin and glass microspheres, close agreement is reported for the chloride solution. Furthermore, PET measurement of the liquefied resin activity is in excellent agreement with the national metrology laboratory reference measurement, suggesting an accurate PET quantification.

For the resin spheres, we reported an average APET/AM ratio of 1.15 ± 0.06 (i.e., AM underestimates APET by 13%); this value is compatible with the high-purity germanium National Institute of Standards and Technology–referred results of Graves et al., who recently reported a ratio of 1.233 ± 0.030 (10). In addition, we provided original data for the glass spheres showing a trend opposite that of resin, an APET/AM ratio of 0.79 ± 0.04 (i.e., AM systematically overestimating APET by 27%). Therefore, a relative difference of about 46% exists between the 2 manufacturers’ 90Y activity calibrations; that is, 1 Bq of 90Y measured in the activity reference frame of the glass microsphere manufacturer corresponds to 1.46 Bq in that of the resin microsphere manufacturer.

Quantitative PET imaging of 90Y is challenging, but it is enhanced by the state-of-the-art time-of-flight systems used in this study. However, an error in the PET-reconstructed activity may arise from a potentially inaccurate attenuation correction due to inadequate modeling of higher-density materials such as glass. The glass containers for the 90Y-chloride and resin microspheres have a minimal thickness. Such a thickness will have minor effects on PET-reconstructed activity. The following observations support this claim: first, our APET/AM for resin microspheres is consistent with prior studies (10,11) reporting that AM is underestimated using a measurement approach different from PET; second, our APET/AM is near unity for 90Y-chloride, for which activity meter measurements are well known with a traceable standard; and third, we obtained a near-unity value for the ratio of the LNHB-CEA reference activity to the PET activity, indicating the good agreement of the 2 methods (i.e., the coincidence Cherenkov counting and the PET) in estimating the vial activity. However, the combination of a thick glass V-Vial (Wheaton Industries, Inc.) bottom and glass microspheres settling at that bottom may lead to a combined glass thickness potentially great enough to introduce bias in the attenuation correction. To estimate the potential bias, we used cone-beam CT of a glass microsphere vial to create a high-resolution (0.1 mm) model of the geometry and material used in our study. Nominal linear attenuation coefficients were then assigned, and attenuation correction factors (ACFs) were calculated along a few lines of response. We compared ACFs from the high-resolution model with ACFs calculated from the CT-derived attenuation map used in the PET reconstruction. Assuming nominal values for diameter and total number of microspheres, with a packing ratio of 0.6, we estimated the potential ACF bias along evaluated lines of response to vary from −11% to +13%. The average ACF bias was no greater than 6%, which would move our results only slightly toward unity, leaving the APET/AM ratio at or below 0.85. The sensitivity of attenuation correction was also tested by increasing the CT numbers above 600 Hounsfield units by 20% and then performing PET reconstruction. The resulting activity concentration image had a maximum difference of 3.6%, demonstrating minimal sensitivity to changes in Hounsfield units.

Although primary measurements from national laboratories have been reported for both devices (12,13), any changes from the specific source and container tied to these measurements will impact the 90Y Bremsstrahlung energy spectrum and thereby the activity meter assay. One study reported a systematic bias of 4% due to likely changes in the acrylic shield used by glass microspheres (14). Monte Carlo simulations that model the composition and geometry of the vials, as well as the devices, might provide more insight on their impact on both the PET measurement and activity meter calibration but are beyond the scope of this work.

To the best of our knowledge, we believe this is the first report of such observed differences for 90Y glass microspheres between PET and vendor-stated activity. The purpose of this study is not to fully explain the discrepancies we have observed but to share our observations that suggest a significant bias when comparing PET quantification with vendor-stated activity for both glass and resin 90Y microsphere devices. Such differences would likely not affect clinical practice given the large number of patients safely and effectively treated to date with activities as stated by the vendors. However, it is important from a metrological standpoint to know the activities administered to patients; reporting true activities should enable more accurate radiobiologic modeling and dosimetry comparisons across devices and modalities. For example, our results should be considered within the context of studies reporting a lower biologic effect per Gray for glass versus resin microspheres when treating the same hepatic disease (15).

CONCLUSION

We have presented original data comparing quantitative PET and manufacturer-declared total activity in 90Y-labeled microspheres and 90Y-chloride vials. Manufacturer-declared vial activities were substantially different when measured by quantitative PET for glass (mean ratio, 0.79) and resin (mean ratio, 1.15), which showed opposite trends with a large relative difference of 46% between them. In 90Y-chloride vials, PET and manufacturer-declared activities agreed closely. We expect that the reported discrepancies will promote further investigations to establish reliable and accurate patient injected-activity measurement and thus consistent dosimetry and dose–effect relation assessments.

DISCLOSURE

Yuni Dewaraja acknowledges grant funding from NIH R01 EB022075. Maurizio Conti is full-time employee of Siemens Medical Solutions USA, Inc. No other potential conflict of interest relevant to this article was reported.

KEY POINTS

QUESTION: How accurate are vendor-specified calibrated activities used for therapy and absorbed dose assessment in 90Y SIRT?

PERTINENT FINDINGS: We compared quantitative 90Y PET measurements against vendor-specified calibrated activities in both glass and resin microsphere vials across multiple centers and devices. We found a large difference between PET measurements and reported vial activities (average, −21% for glass and +15% for resin).

IMPLICATIONS FOR PATIENT CARE: Accounting for the observed differences can lead to a shift of reported administered activity and absorbed dose thresholds in dose–effect studies.

ACKNOWLEDGMENTS

We thank James Hamill from Siemens Medical Solutions USA, Inc., for performing the attenuation correction sensitivity analysis, and we thank Dr. Christophe Bobin and his team (LNHB/CEA) for providing us with the dissolved resin source.

Footnotes

  • Published online Nov. 23, 2022.

  • © 2023 by the Society of Nuclear Medicine and Molecular Imaging.

REFERENCES

  1. 1.
    1. Weber M,
    2. Lam M,
    3. Chiesa C,
    4. et al
    . EANM procedure guideline for the treatment of liver cancer and liver metastases with intra-arterial radioactive compounds. Eur J Nucl Med Mol Imaging. 2022;49:1682–1699.
  2. 2.
    1. Garin E,
    2. Tselikas L,
    3. Guiu B,
    4. et al
    . Personalised versus standard dosimetry approach of selective internal radiation therapy in patients with locally advanced hepatocellular carcinoma (DOSISPHERE-01): a randomised, multicentre, open-label phase 2 trial. Lancet Gastroenterol Hepatol. 2021;6:17–29.
  3. 3.
    1. Veenstra EB,
    2. Ruiter SJS,
    3. de Haas RJ,
    4. Bokkers RPH,
    5. de Jong KP,
    6. Noordzij W
    . Post-treatment three-dimensional voxel-based dosimetry after yttrium-90 resin microsphere radioembolization in HCC. EJNMMI Res. 2022;12:9.
  4. 4.
    1. Willowson KP,
    2. Eslick EM,
    3. Bailey DL
    . Individualised dosimetry and safety of SIRT for intrahepatic cholangiocarcinoma. EJNMMI Phys. 2021;8:65.
  5. 5.
    1. Saldarriaga Vargas C,
    2. Bauwens M,
    3. Pooters INA,
    4. et al
    . An international multi-center investigation on the accuracy of radionuclide calibrators in nuclear medicine theragnostics. EJNMMI Phys. 2020;7:69.
  6. 6.
    1. Carlier T,
    2. Willowson KP,
    3. Fourkal E,
    4. Bailey DL,
    5. Doss M,
    6. Conti M
    . 90Y-PET imaging: exploring limitations and accuracy under conditions of low counts and high random fraction. Med Phys. 2015;42:4295–4309.
  7. 7.
    1. Willowson KP,
    2. Tapner M
    , QUEST Investigator Team, Bailey DL. A multicentre comparison of quantitative 90Y PET/CT for dosimetric purposes after radioembolization with resin microspheres: the QUEST Phantom Study. Eur J Nucl Med Mol Imaging. 2015;42:1202–1222.
  8. 8.
    1. Dewaraja YK,
    2. Devasia T,
    3. Kaza RK,
    4. et al
    . Prediction of tumor control in 90Y radioembolization by logit models with PET/CT-based dose metrics. J Nucl Med. 2020;61:104–111.
  9. 9.
    1. Lourenço V,
    2. Bobin C,
    3. Chisté V,
    4. et al
    . Primary standardization of SIR-Spheres based on the dissolution of the 90Y-labeled resin microspheres. Appl Radiat Isot. 2015;97:170–176.
  10. 10.
    1. Graves SA,
    2. Martin M,
    3. Tiwari A,
    4. Merrick M,
    5. Sunderland J
    . SIR-Spheres activity measurements reveal systematic miscalibration. J Nucl Med . 2022;63:1131–1135.
  11. 11.
    1. Selwyn R,
    2. Micka J,
    3. DeWerd L,
    4. Nickles R,
    5. Thomadsen B
    . Technical note: the calibration of 90Y-labeled SIR-Spheres using a nondestructive spectroscopic assay. Med Phys. 2008;35:1278–1279.
  12. 12.
    1. Mo L,
    2. Avci B,
    3. James D,
    4. et al
    . Development of activity standard for 90Y microspheres. Appl Radiat Isot. 2005;63:193–199.
  13. 13.
    1. Coursey BM,
    2. Calhoun JM,
    3. Cessna JT
    . Radioassays of yttrium-90 used in nuclear medicine. Nucl Med Biol. 1993;20:693–699.
  14. 14.
    1. Bergeon D,
    2. Cessna J,
    3. Zimmeran B
    . Challenges in migrating 90Y calibrations to new ionization chambers: geometry, height, and impurity effects [abstract]. J Nucl Med. 2018;59(suppl 1):428.
  15. 15.
    1. d’Abadie P,
    2. Hesse M,
    3. Jamar F,
    4. Lhommel R,
    5. Walrand S
    . 90Y TOF-PET based EUD reunifies patient survival prediction in resin and glass microspheres radioembolization of HCC tumours. Phys Med Biol. 2018;63:245010.
  • Received for publication June 2, 2022.
  • Revision received November 7, 2022.
SNMMI

© 2025 SNMMI

Powered by HighWire