An accurate diagnosis of the cause of pulmonary hypertension determines the management and prognosis. Patients with chronic thromboembolic pulmonary hypertension (CTEPH) represent a distinct category of PH as they can potentially be offered curable surgery in the form of pulmonary endarterectomy. This operative procedure has gained wide acceptance and achieved good success rates, with a reported mortality rate of 4%8% (47).
Without treatment, the prognosis of CTEPH is very poor, with a quoted survival rate of 30% over a period of 5 y for patients with mean pulmonary artery pressure > 30 mm Hg (8,9).
Recent literature have shown that up to 3.8% of patients diagnosed with acute pulmonary embolism (PE) develop CTEPH at 2-y follow-up (10) and suggest that the incidence of CTEPH is much higher than previously thought (10,11). Furthermore, up to 63% of patients diagnosed with CTEPH may not have a previous history of documented PE (12).
Identifying CTEPH as the cause of PH is facilitated by several imaging techniques, including ventilationperfusion (V/Q) scintigraphy, multidetector CT pulmonary angiography (CTPA), high-resolution CT (HRCT), pulmonary digital subtraction angiography (DSA), and, more recently, magnetic resonance angiography (13,14).
The introduction of spiral CTPA has challenged the role of V/Q scanning in diagnosing acute PE (15). In the United Kingdom, it has become the first line of the investigation in patients with suspected acute PE (16).
Further advances in CT technology, with the introduction of multidetector CTPA and multiplanar reconstruction have further improved the sensitivity and specificity of this technique with small studies quoting values of 96% and 98%, respectively (17). However, a recent large multicenter study has shown these figures to be lower, with a sensitivity of 83% and a specificity of 96% for CTPA, with even a lower sensitivity for segmental (68%) and subsegmental (25%) branches (18).
This has been adopted by PIOPED (Prospective Investigation of Pulmonary Embolism Diagnosis) II investigators and is reflected in the new guidelines published in January 2007, which have approved multidetector CTPA as the first-line imaging procedure in most patients suspected of having PE (19).
However, the sensitivity of multidetector CTPA in detecting chronic thromboembolic pulmonary disease remains variable. There are few publications in this respect, including a study published by Pitton et al. showing multidetector CTPA to have a sensitivity of 70.4% for segmental and 63.6% for subsegmental branches when compared with pulmonary DSA (20).
On the other hand, several studies have shown that high-probability and intermediate-probability V/Q scintigraphy has a high sensitivity and specificity for distinguishing patients with chronic pulmonary thromboembolic disease from other causes of PH, to the extent that normal V/Q scintigraphy can comfortably rule out CTEPH (2127). To our knowledge, there is no published study that compares the value of V/Q scintigraphy with that of multidetector CTPA in the diagnosis of CTEPH. Our aim was to perform a head-to-head comparison of the 2 techniques.
| MATERIALS AND METHODS |
|---|
|
|
|---|
The identification of the PH clinical class (e.g., CTEPH vs. non-CTEPH etiology) was made in all patients in accordance with the guidelines of the European Society of Cardiology and the American College of Chest Physicians (13).
V/Q Scintigraphy
V/Q scintigraphy was performed using 99mTc-labeled macroaggregated albumin (99mTc-MAA) and 81mKr gas (81Kr). Images were acquired using an ECAM dual-head
-camera (Siemens Medical Solutions USA, Inc.). The perfusion images were acquired after intravenous administration of 100 MBq of 99mTc-MAA with the patient in the supine position. The images were acquired in 4 standard projections (anterior, posterior, right posterior oblique, and left posterior oblique) with a medium-energy, general-purpose collimator. An energy window of 20% centered on the energy peak of 99mTc (140 keV), and a matrix of 128 x 128 pixels was used and approximately 400,000 counts accumulated per view. Ventilation images using 81mKr were acquired immediately after each perfusion view. 81mKr was inhaled through a mouthpiece. The acquisition parameters were the same as those used for the perfusion study, but the 20% window was centered on the energy peak of 81mKr (190 keV). Images were reported from a Hermes workstation (Nuclear Diagnostics) by experienced nuclear medicine physicians. Images were interpreted according to the modified PIOPED criteria (28,29). A high-probability scan was suggestive of chronic thromboembolic pulmonary disease, whereas a low-probability scan was suggestive of nonchronic thromboembolic pulmonary disease etiology. For intermediate-probability scans, 2 separate datasets were produced and analyzed. In one dataset, intermediate-probability scans were considered as suggestive of chronic thromboembolic pulmonary disease and, in the other dataset, the intermediate-probability scans were considered as not suggestive of chronic thromboembolic pulmonary disease. This was in accordance with previous similar studies (22).
Multidetector CTPA
CTPA was acquired with multidetector row CT scanners. The first 24 studies were acquired with a Somaton Four Plus CT scanner (Siemens Medical Solutions) using 4 x 3 mm collimation and 2.5-mm reconstruction. From March 2002, the remaining 203 studies were acquired on a Light Speed CT scanner (GE Healthcare), using 8 x 1.25 mm collimation with 1.25-mm reconstruction. A bolus injection of 100 mL of nonionic iodinated contrast medium (300 mg/mL) was administered intravenously via a peripheral cannula at 4 mL/s via a power injector with a variable-start delay. This was determined using automatic bolus trigger software with a circular region of interest positioned at the level of main pulmonary trunk. The images were acquired in the craniocaudal direction with the z-axis coverage and the field of view chosen to include the entire thorax, from the apex to the base of the lungs.
All CTPA was reported by experienced radiologists using PACS workstations (GE Healthcare) as well as Siemens and General Electric workstations to assess multiplanar reconstructions. A report was considered as suggestive of chronic thromboembolic pulmonary disease if it stated visualization of the thrombus, calcified thrombus, recanalization, sudden change in vessel caliber, strictures, poststenotic dilatation, webs, or perfusion abnormality (3034). The presence of mosaic perfusion, defined as areas of increased and decreased attenuation and vascularity without evidence of destruction or displacement of pulmonary vessels, was also noted. However, its presence alone in the absence of other signs was not considered as diagnostic of chronic thromboembolic pulmonary disease (3537).
Pulmonary DSA
The indications for referring patients for pulmonary DSA included (a) clinical suspicion of CTEPH, (b) at least one positive imaging modality for chronic thromboembolic pulmonary disease, and (c) whether the patient was being assessed for pulmonary thrombendarterectomy. Pulmonary DSA was performed in 61 patients by experienced vascular radiologists via a femoral venous approach using a 7-French Grollman pigtail catheter (William Cook Europe) placed in the main pulmonary trunk. More selective studies of the intrapulmonary artery branches were performed in some patients with a 7-French headhunter catheter (Cordis Europa). Images were obtained with an HDI 5000 fluoroscopy unit (Philips) after administration of 20 mL of nonionic iodinated contrast medium (300 mg/mL), per DSA run, at 7 mL/s via a pump injector or by hand injection with a maximum volume limit of 300 mL. The minimum projections per hemithorax included frontal and posterior oblique projections, with extra views obtained as required.
| RESULTS |
|---|
|
|
|---|
In patients with CTEPH (group A, n = 78 patients), the final diagnosis was confirmed by pulmonary DSA in 61 patients. In the remaining 17 patients, there was strong clinical and imaging suspicion of chronic thromboembolic pulmonary disease. However, these patients did not consent to surgery or were considered unfit for surgery and, therefore, not referred for pulmonary DSA.
In patients with non-CTEPH (group B, n = 149 patients), the clinical and imaging data did not suggest thromboembolic etiology, so there was no clinical indication to refer them for pulmonary DSA. The final diagnosis of PH of nonthromboembolic etiology was achieved by a consensus based on clinical and imaging data discussed at a multidisciplinary meeting.
Patients with CTEPH: Group A
CTPA was reported as showing features suggestive of chronic thromboembolic pulmonary disease in 40 of 78 patients but was falsely negative in 38 of 78 cases, giving a sensitivity of 51%. The V/Q scintigraphy showed high probability in 75 patients, intermediate probability in 1 patient, and low probability in 2 patients. When the intermediate-probability scans were considered suggestive of chronic thromboembolic pulmonary disease (grouped with the high probability), the V/Q sensitivity was 97.4%, and when intermediate probability was considered as negative for chronic thromboembolic pulmonary disease (grouped with low probability), the sensitivity was 96% (Tables 1 and 2).
|
|
The specificity of CTPA was 99%. The V/Q specificity was 90% when the intermediate-probability scans were analyzed as suggestive of chronic thromboembolic pulmonary disease and 95% when the intermediate scans were considered negative for chronic thromboembolic pulmonary disease (Tables 1 and 2).
Mosaic Perfusion
Mosaic perfusion was reported in 53 of the total 227 CTPA scans (23.34%). Among those, 43 (18.94%) were in group A and 10 (4.41%) were in group B.
Thus, the CTPA showed evidence of mosaic perfusion in 55% (43/78) of patients with CTEPH (group A).
| DISCUSSION |
|---|
|
|
|---|
In our study, we approach the V/Q scintigraphy analysis from 2 perspectives. First (V/Q (1)), we considered both the high- and intermediate-scans as suggestive of chronic pulmonary thromboembolic disease, which resulted in a sensitivity of 97.4%, specificity of 90%, accuracy of 92.5%, negative predictive value (NPV) of 98.5%, and positive predictive value (PPV) 83.5%, respectively (Table 2).
When we analyzed the data considering just the high-probability scans as suggestive of chronic pulmonary thromboembolic disease (V/Q (2)), the sensitivity was 96.2%, whereas the specificity increased to 94.6%. The overall accuracy was 95.2% with a NPV of 97.9% and a PPV of 90.3% (Table 2).
The sensitivity of CTPA in our study was significantly lower compared with V/Q scintigraphy in detecting chronic thromboembolic pulmonary disease changes. However, it was similar with that reported by Pitton et al., who had analyzed 994 vessel segments in 14 patients with CTEPH and had both CTPA and selective pulmonary DSA (20). Their analysis showed CTPA to be significantly inferior to selective DSA, with a 67% overall concordance in diagnosing CTEPH. Furthermore, the nonocclusive changes of the vessels were significantly underdiagnosed by CT, with a concordance between CT and DSA of 23.1% (20).
False-Negative V/Q Scintigraphy
The 2 patients from group A who had their V/Q scintigraphy interpreted as low probability did not have the diagnosis of CTEPH confirmed on pulmonary DSA. The final diagnosis was made by CTPA, which showed calcified thrombus in one patient and narrowed but patent pulmonary arterial branches in the other patient. By corroborating the patient clinical history and the CT findings, it was concluded that these patients most likely had previous PE and the pulmonary vessels had recanalized.
V/Q scintigraphy that was interpreted as intermediate probability was confirmed on pulmonary DSA to be CTEPH. The CT was reported as showing severe emphysematous changes but failed to show signs of CTEPH. It was the V/Q result together with the clinical history that prompted the clinicians to request pulmonary DSA.
False-Positive V/Q Scintigraphy
Eight V/Q scans were reported as high probability and 7 were reported as intermediate probability among the non-CTEPH group. None of these patients had pulmonary DSA. The clinical diagnoses were idiopathic pulmonary arterial hypertension, pulmonary arterial hypertension associated with atrial septal defect, pulmonary venoocclusive disease, and PH related to parenchymal lung disease, such as emphysema or pulmonary fibrosis (Table 3).
|
| CONCLUSION |
|---|
|
|
|---|
| FOOTNOTES |
|---|
| References |
|---|
|
|
|---|
Related articles in JNM:
This article has been cited by other articles:
![]() |
K. M. Chin and L. J. Rubin Pulmonary arterial hypertension. J. Am. Coll. Cardiol., April 22, 2008; 51(16): 1527 - 1538. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. J. van Beers, B. L. F. van Eck-Smit, M. R. Mac Gillavry, C. F. J. van Tuijn, J. W. J. van Esser, D. P. M. Brandjes, M. C. Kappers-Klunne, A. J. Duits, B. J. Biemond, J.-J. B. Schnog, et al. Large and Medium-Sized Pulmonary Artery Obstruction Does Not Play a Role of Primary Importance in the Etiology of Sickle-Cell Disease-Associated Pulmonary Hypertension Chest, March 1, 2008; 133(3): 646 - 652. [Abstract] [Full Text] [PDF] |
||||
![]() |
National Pulmonary Hypertension Centres of the UK Consensus statement on the management of pulmonary hypertension in clinical practice in the UK and Ireland Heart, March 1, 2008; 94(Suppl_1): i1 - i41. [Full Text] [PDF] |
||||
![]() |
National Pulmonary Hypertension Centres of the UK Consensus statement on the management of pulmonary hypertension in clinical practice in the UK and Ireland Thorax, March 1, 2008; 63(Suppl_2): ii1 - ii41. [Full Text] [PDF] |
||||
![]() |
L. M. Freeman Don't Bury the V/Q Scan: It's as Good as Multidetector CT Angiograms with a Lot Less Radiation Exposure J. Nucl. Med., January 1, 2008; 49(1): 5 - 8. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |
| JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY | THE JOURNAL OF NUCLEAR MEDICINE |