|
|
|||||||||
Clinical Investigations |
1 Department of Nuclear Medicine, University Hospital Essen, Essen, Germany
2 Institute of Medicine, Research Centre Jülich, Jülich, Germany
3 Department of Nuclear Medicine, Friedrich-Alexander-University Erlangen, Erlangen, Germany
4 Department of Physics, University of Wuppertal, Wuppertal, Germany
| ABSTRACT |
|---|
|
|
|---|
Key Words: PET/CT motion artifacts positioning aids
| INTRODUCTION |
|---|
|
|
|---|
Nevertheless, spatial coregistration throughout the extended coaxial imaging range is sometimes compromised by involuntary patient motion during the combined examination. Respiratory motion, for example, has been shown to potentially yield CT and PET data that are misregistered in anatomic regions of high mobility, such as the lower thorax and the diaphragm (79). If significant, this misregistration can propagate through CT-based attenuation correction into a biased tracer distribution (7,9,10).
Involuntary patient motion may also be due to relaxation of muscles during the combined examination, which takes about 2035 min for a whole-body (WB) acquisition (11,12). The likelihood of muscle relaxation increases with scanning time; therefore, body regions such as the head and neck, with the largest time differences between the CT and the PET portions, are prone to misregistration (Fig. 1).
|
| MATERIALS AND METHODS |
|---|
|
|
|---|
We assume a standardized WB PET/CT acquisition protocol (12), with patients being positioned headfirst supine in the scanner, with the neck centered inside the gantry (Fig. 2A). First, a topogram was acquired, followed by a spiral CT scan in the craniocaudal direction (0
t
tCT, where t = scan time and tCT = time at the end of the CT scan). The patient was then moved toward the rear of the gantry, where a multibed PET emission scan commenced with retracing of the imaging range in the craniocaudal direction. We assumed a coaxial imaging range of 8 bed positions with an emission scan time of 3.5 min per bed position. Therefore, the neck of the subject was back in the original position inside the gantry after about 30 min. Figure 2B illustrates the position of the subjects with respect to the coordinate system of the motion-tracking device.
|
|
For setup C (12 subjects), subjects were placed on a foam bean bag (MED-TEC). This type of support is also referred to as vacuum-lock bag since an attached pump is used to extract the air from the bag, thus providing an adaptable and reusable patient support for single-time or repeated scanning in a fixed position. In our study, the vacuum-lock bag was butterfly-shaped, 45 x 60 x 2 cm, and filled with 2-mm diameter foam pellets of a low density. First, the bag was slightly filled with air and fitted to the subjects neck. The technologist then extracted the air from the bag via the attached electric pump, while the cushion was supported further and fitted more tightly to the neck and lower head (Fig. 3C). Because this vacuum-lock bag did not fit into the curved head holder used for setup B, it rested on the pallet without further restraints.
Setup D (12 subjects) combined setups B and C. A reusable small (35 x 60 x 1 cm) vacuum-lock bag (Fig. 3D) was custom-made (Additec GmbH) for the Institute of Medicine in Jülich to replace the disposable construction-foam supports (Fig. 3B). The small vacuum-lock bag was also placed inside the curved head holder (B).
None of the volunteers was claustrophobic or had undergone a WB PET or PET/CT study before. At a mean age of 27 y, volunteers for this simulation study were younger than the general oncology patients referred to the participating PET/CT site (University Hospital Essen). Similar to patients undergoing standard clinical PET/CT, the subjects in this study were asked to remain still for the entire simulated scan procedure.
Infrared Camera to Track Patient Motion
An optical tracking system (Polaris; Northern Digital) was used to monitor the position of the head and neck during the simulated scan. The tracking device used infrared light to monitor the location of several passive, reflective markers. The device consisted of a camera (emitter and receiver) that was mounted behind the PET tomograph facing the rear gantry (Fig. 2B), a set of 3 target markers, and another set of reference markers. The reference markers were mounted on the rear of the gantry and remained stationary inside the active area of the camera. The 3 nonstationary, reflecting target markers were mounted on a plastic frame that attached to a pair of plastic goggles. All subjects were asked to wear those goggles fitted tightly against their head and supported by their nose and ears, with the marker points held in place just above the top of their head, while they faced the infrared camera (15). The active field of measurement of the tracking device was 100 cm, starting 85 cm from the infrared camera. The manufacturer quotes a tracking accuracy of 0.35 mm root mean square within this range.
Estimating Motion of Head and Neck in WB PET/CT Studies
All volunteers were positioned head-first, supine, with their arms down, with their legs resting on a low-density support, and with their head and neck centered inside the field of view of the tomograph. The bed position was fixed during the first 2.5 min to simulate the topogram and spiral CT scan (Fig. 2A), which was assumed to correspond to the time needed to acquire a topogram and a spiral CT scan as part of the clinical PET/CT examination (tCT). With the subject lying still, the bed was then moved out of the tomograph by 21 cm to simulate the first bed position (lower pelvis). The bed was moved back in by 3 cm after every 3.5 min, to simulate the duration of an individual bed position, until, at 27.5 min (tPET_neck), the subjects head was back in the original position. The distance of the simulated bed steps was chosen such that the target points were always in the active field of view of the infrared tracking system.
Motion of the head and neck was estimated for each subject from the difference in the position of the 3 target points at the time of the last bed position (neck), at tPET_neck, with respect to the initial position at tCT. The absolute difference, that is, the PET/CT misregistration, was reported in 3 rotational and 3 translational components. The 3 rotational parameters are the pitch, yaw, and roll, which describe the rotation around the y-, x-, and z-axes, respectively (Fig. 2B). The 3 translational displacements,
x,
y, and
z, were measured in the x, y, and z directions, respectively.
Motion of the marker points was tracked for all subjects in the 4 different positioning scenarios (Fig. 3AFig. 3D). For each positioning scenario, we report the mean of the absolute rotational (pitch, yaw, roll) and translational displacements (
x,
y,
z). For each subject in setups AD, we calculated the absolute displacement,
r, as the square root of the sum of the squared translational displacements
x,
y, and
z and report the mean of the individual
r values, <
r>, for setups AD. Similarly, we report the mean of the average rotational displacements, <rot>, for setups AD.
| RESULTS |
|---|
|
|
|---|
|
|
Overall, the mean rotational displacement was reduced to 0.4° (P < 0.05), 0.7° (P = 0.1), and 0.3° (P < 0.05) for setups B, C, and D, respectively, compared with 1° for setup A. The respective average translational displacements, <
r>, were 2 mm (P < 0.05), 4 mm (P = 0.06), and 1.4 mm (P < 0.05), compared with 7 mm for setup A.
| DISCUSSION |
|---|
|
|
|---|
Within this limitation, we were able to show that misregistration of the head and neck was largest when patients were positioned without restraint of the area of interest (setup A). The measured average translational and rotational misalignment was 7 mm and 1°, respectively, and the maximum was 12 mm and 2°, respectively, with a large spread of head-and-neck motion in the vertical direction (
y) (Table 1 and Fig. 4A). Motion of the head and neck was reduced significantly when the curved head holder was combined with the custom-fitted foam insert (setup B). The resulting average translational and rotational misalignment was only 2 mm and 0.4°, respectively, with a maximum misalignment of 8 mm or 1° in different subjects. However, this setup was ineffective in reducing the translational misregistration along the main scanner axis, as indicated by the relatively large spread in
z (Fig. 4B). It appears likely that the neck was restrained insufficiently as a result of a nonevenly distributed foam support, with more volume being pushed into the head holder than below the neck in front of it, thus indicating the importance of careful patient positioning even with additional positioning aids at hand.
By comparison, the butterfly-shaped vacuum-lock bag of setup C reduced, in particular, the roll rotation around the main scanner axis z and the translation along the anteroposterior direction (
y). The efficacy of setup C can be explained by the support volume that extends even beyond the bilateral boundaries of the curved head holder in setup B and thus reduces both the misregistration and the intersubject variability of the misregistration around the main scanner axis z (Fig. 4C). Nevertheless, combination of the curved head holder and the vacuum-lock bag (setup D) resulted in the most significant and least variable reduction in translational and rotational misalignment (Table 1), because of the particularly rigid encasement of the head and neck at either side of the lower head and along the entire axial range of the head and neck (Fig. 3).
In practice, however, similarly curved head holders are not available commercially. To evaluate the efficacy of a versatile patient-positioning scheme in clinical PET/CT practice, we propose combining a generic arm support (Fig. 5A) made of semirigid foam (MED-TEC) and the vacuum-lock bag from setup C. In combination with the custom-fitted vacuum-lock bag, which is placed between the neck of the patient and the foam-based armrest (Fig. 5B), a potentially effective support of the head and neck for lengthy WB PET/CT studies might be provided.
|
Because motion of the head and neck could not be monitored in real time, we evaluated the misalignment of the head and neck on the reconstructed WB images. We extracted sets of contiguous axial CT and PET images that covered the range from the tip of the cranium to the lower neck from the WB PET/CT image volumes. The resulting 20 subvolumes were transferred to a commercially available image registration platform (Fusion7D; Mirada Solutions). For each patient, automatic linear registration of the CT and PET volumes was performed, and the mean translational and rotational displacement of the PET images with respect to the CT images was calculated with reference to the center of the image volume. The difference in motion of the head and neck, that is, the displacement of the head and neck between the CT and PET portions of the combined examination, was accepted as a measure of the efficacy of the particular patient positioning.
We found that the average rotational and translational displacement in group 1with the armrest onlywas 1.2° ± 0.6° and 6 ± 5 mm, respectively (Fig. 6A). This displacement was reduced in group 2with the additional vacuum-lock supportto 0.9° ± 0.3° and 2 ± 1 mm, respectively. The reduction in translation was statistically significant (Student t test, P < 0.05). The maximum displacement was reduced from 2.5° and 15 mm in group 1 to 1.4° and 3.7 mm in group 2 (different subjects). Figures 6B and 6C show examples of fused PET/CT images of patients from group 1 and 2, respectively, before and after linear image registration.
|
Adequate patient positioning in clinical routine requires support structures of low density, particularly when used in combination with CT equipment. All positioning aids in our simulation and clinical studies were of low CT density (900 HU to 1,000 HU) and did not cause image artifacts. Short setup times are essential for routine use of positioning aids, to limit exposure of the staff to radiation from the injected patient. In our study, a single technologist could complete the patient setup in 4 min (setup B) or less, with an average setup time of 2 min.
Routinely applied patient supports should also be reusable and, if possible, adaptable to the patient. Although alignment was better with the construction foam (setup B) than with the vacuum-lock bag alone (setup C), the accumulated costs for using disposable foam-based positioning aids in a high-throughput scenario can be quite high. In contrast, vacuum-lock bags are reusable and can be tailored to specific needs, such as positioning the entire trunk or the neck only, by varying the size, packing fraction, and coating of the bags. Furthermore, vacuum bags do not cause discomfort to the patient. Although the cost of these bags may be offsetting, this cost may soon be justified in reasonable-throughput scenarios when compared with the cost of disposable foam inserts and the volume of waste they create.
Finally, our simulation and patient studies indicated that the accuracy of coregistration in the head and neck in WB PET/CT studies was more reliable when additional positioning aids were used. Any residual misregistration between CT and PET (as shown for the head and neck in this study) may partially invalidate PET/CT image fusion and propagate into a biased tracer distribution after CT-based attenuation correction. This motion-induced bias can be estimated from the effect of head motion on PET of the brain as shown by an independent study (15) that indicated visual and quantitative distortion of the recovered PET tracer distribution when the motion of the head exceeded one half the full width at half maximum of the spatial resolution of the PET scanner (16). Our clinical study showed that the mean misregistration in the head and neck could be reduced to 2 mm, which is less than the full width at half maximum of current WB PET tomographs. Nevertheless, when imaging of the head and neck is the primary objective, separate acquisition protocols for the neck and torso should be considered to help increase patient comfort and eliminate systematic artifacts, such as those caused by the position of the arms with respect to the CT field of view (12,17).
| CONCLUSION |
|---|
|
|
|---|
| ACKNOWLEDGMENTS |
|---|
| FOOTNOTES |
|---|
For correspondence or reprints contact: Thomas Beyer, PhD, Department of Nuclear Medicine, University Hospital Essen, Hufelandstrasse 55, D 45122 Essen, Germany.
E-mail: thomas.beyer{at}uni-essen.de
| REFERENCES |
|---|
|
|
|---|
Related articles in JNM:
This article has been cited by other articles:
![]() |
C. Nahmias, C. Lemmens, D. Faul, E. Carlson, M. Long, T. Blodgett, J. Nuyts, and D. Townsend Does Reducing CT Artifacts from Dental Implants Influence the PET Interpretation in PET/CT Studies of Oral Cancer and Head and Neck Cancer? J. Nucl. Med., July 1, 2008; 49(7): 1047 - 1052. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Kuehl, P. Veit, S. J. Rosenbaum, A. Bockisch, and G. Antoch Can PET/CT Replace Separate Diagnostic CT for Cancer Imaging? Optimizing CT Protocols for Imaging Cancers of the Chest and Abdomen J. Nucl. Med., January 1, 2007; 48(1_suppl): 45S - 57S. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. J. Montgomery, K. Thielemans, M. A. Mehta, F. Turkheimer, S. Mustafovic, and P. M. Grasby Correction of Head Movement on PET Studies: Comparison of Methods J. Nucl. Med., December 1, 2006; 47(12): 1936 - 1944. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Brechtel and T. Beyer Reply: Adequate Evaluation of Image Registration in Hybrid PET/CT. J. Nucl. Med., September 1, 2006; 47(9): 1557 - 1558. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | RSS | TABLE OF CONTENTS |
| JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY | THE JOURNAL OF NUCLEAR MEDICINE |