The effects of incomplete breath-holding on 3D MR image quality

J Magn Reson Imaging. 1997 Nov-Dec;7(6):1132-9. doi: 10.1002/jmri.1880070628.

Abstract

The purpose of this study was to investigate how fast three-dimensional (3D) MR image quality is affected by breath-holding and to develop an optimal breath-holding strategy that minimizes artifact in the event of an incomplete breath-hold. A computer model was developed to study variable-duration breath-holds during fast 3D imaging. Modeling was validated by 3D gradient-echo imaging performed on 10 volunteers. Signal-to-noise ratio (SNR) and image blur were measured for both simulated and clinical images. Insights gained were applied to clinical 3D gadolinium-enhanced MR angiography. Breath-holding significantly improved abdominal 3D MR image quality. Most of this benefit could be achieved with a breath-hold fraction of 50% if it occurred during acquisition of central k space. Breath-holding during peripheral k-space acquisition, however, had no significant benefit. Respiratory motion artifact on fast 3D MRI occurring when a patient fails to suspend respiration for the entire scan duration can be minimized by collecting central k space first (centric acquisition) so that premature breathing affects only the acquisition of peripheral k space.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Artifacts
  • Computer Simulation
  • Contrast Media
  • Female
  • Gadolinium
  • Humans
  • Magnetic Resonance Angiography
  • Magnetic Resonance Imaging*
  • Male
  • Respiration / physiology*