Validation of global and segmental left ventricular contractile function using gated planar technetium-99m sestamibi myocardial perfusion imaging

J Am Coll Cardiol. 1994 Jan;23(1):141-5. doi: 10.1016/0735-1097(94)90512-6.

Abstract

Objectives: The purpose of this study was to test the hypothesis that segmental wall motion analysis determined from gated planar technetium-99m sestamibi myocardial imaging is reproducible and agrees well with echocardiographic data.

Background: Technetium-99m sestamibi is a new radiopharmaceutical recently approved for myocardial perfusion imaging. Its advantages include a dosimetry that allows use of a dose 10 to 15 times higher than that of thallium-201. As a result, myocardial counts are markedly improved and images can be collected in a gated mode to potentially allow assessment of global and segmental ventricular function. However, the reproducibility and accuracy of technetium-99m sestamibi imaging for measurement of global and segmental left ventricular function have not been evaluated or compared with those of a standard ventricular function technique, such as echocardiography.

Methods: We studied 136 patients referred for clinical technetium-99m sestamibi imaging. One-day rest-stress planar technetium-99m sestamibi protocols were used, gating the stress images. After technetium-99m sestamibi imaging, all patients had standard rest two-dimensional echocardiography. Global and segmental technetium-99m sestamibi and echocardiographic left ventricular contraction was graded qualitatively as normal or abnormal using a four-point grading system.

Results: Interobserver and intraobserver agreement was extremely high for global and segmental technetium-99m sestamibi wall motion analysis, with absolute agreements ranging from 0.92 to 1.00 and corresponding kappa values of 0.74 to 1.00 (p < 0.00001). Agreement with global and segmental echocardiographic wall motion was similarly very high, with absolute agreements ranging from 0.93 to 1.00 and corresponding kappa values of 0.75 to 1.00 (p < 0.00001).

Conclusions: Gated technetium-99m sestamibi cardiac imaging provides information with regard to rest global and segmental left ventricular systolic function that is highly reproducible and agrees very well with results of two-dimensional echocardiography.

MeSH terms

  • Adult
  • Echocardiography
  • Female
  • Gated Blood-Pool Imaging*
  • Heart / diagnostic imaging*
  • Humans
  • Male
  • Middle Aged
  • Myocardial Contraction*
  • Observer Variation
  • Reproducibility of Results
  • Technetium Tc 99m Sestamibi*
  • Ventricular Function, Left*

Substances

  • Technetium Tc 99m Sestamibi