A technique for using CT images in attenuation correction and quantification in SPECT

Nucl Med Commun. 1989 Feb;10(2):83-97. doi: 10.1097/00006231-198902000-00002.

Abstract

A technique is described for using computed tomography (CT) images for attenuation correction and quantification in SPECT. The CT images are aligned with the corresponding SPECT slices and the Hounsfield units are converted to linear attenuation coefficient values for the SPECT radionuclide. The attenuation coefficient map thus produced is used to provide the attenuation correction required in the SPECT reconstruction. The technique has been evaluated in both a non-anatomical and an anatomical phantom giving a mean accuracy in quantifying activity of various features in the phantoms of 2.6% (range 0.3%-4.0%). The value of performing scatter correction prior to attenuation correction in obtaining accurate quantification is demonstrated. The practicalities of applying the technique in patient studies are discussed.

MeSH terms

  • Abdomen / diagnostic imaging
  • Algorithms
  • Humans
  • Models, Structural
  • Scattering, Radiation
  • Thorax / diagnostic imaging
  • Tomography, Emission-Computed / methods*