Repeatability of metabolically active tumor volume measurements with FDG PET/CT in advanced gastrointestinal malignancies: a multicenter study

Radiology. 2014 Nov;273(2):539-48. doi: 10.1148/radiol.14132807. Epub 2014 May 26.

Abstract

Purpose: To evaluate the feasibility and repeatability of various metabolically active tumor volume ( MATV metabolically active tumor volume ) quantification methods in fluorine 18 fluorodeoxyglucose ( FDG fluorine 18 fluorodeoxyglucose ) positron emission tomography (PET)/computed tomography (CT) in a multicenter setting and propose the optimal MATV metabolically active tumor volume method together with the minimal threshold for future response evaluation studies.

Materials and methods: The study was approved by the institutional review board of all four participating centers, and patients provided written informed consent. Thirty-four patients with advanced gastrointestinal malignancies underwent two FDG fluorine 18 fluorodeoxyglucose PET/CT examinations within 1 week. MATV metabolically active tumor volume s were defined semiautomatically with 27 variations of tumor delineation methods with different reference values. Feasibility was determined as the percentage of successful tumor segmentations per MATV metabolically active tumor volume method. Repeatability was determined with intraclass correlation coefficients, Bland-Altman plots, and limits of agreement ( LOA limit of agreement s) of the percentage difference between the test and repeat test measurements. In addition, LOA limit of agreement variability per center was investigated.

Results: In total, 136 lesions were identified. Feasibility of tumor segmentation ranged from 54% to 100% (74-136 of 136 lesions); repeatability was evaluated for 19 MATV metabolically active tumor volume methods with feasibility of greater than 95%. The median MATV metabolically active tumor volume derived with 50% threshold of mean standardized uptake value ( SUV standardized uptake value ) of a sphere of 12-mm diameter with highest local intensity ( SUVhp mean SUV of a sphere of 12-mm diameter with highest local intensity ), which may not include the voxel with highest SUV standardized uptake value corrected for local background, was 5.7 and 6.1 mL for test and retest scans, respectively, with a relative LOA limit of agreement of 36.1%. Comparable repeatability was found between centers. A difference in uptake time between scan 1 and 2 of 15 minutes or longer had a minor negative influence on repeatability.

Conclusion: MATV metabolically active tumor volume measured with 50% of SUVhp mean SUV of a sphere of 12-mm diameter with highest local intensity corrected for local background is recommended in multicenter FDG fluorine 18 fluorodeoxyglucose PET/CT studies on the basis of a high feasibility (96%) and repeatability ( LOA limit of agreement of 36.1%).

Publication types

  • Clinical Trial, Phase I
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Canada
  • Female
  • Fluorodeoxyglucose F18
  • Gastrointestinal Neoplasms / diagnostic imaging
  • Gastrointestinal Neoplasms / pathology*
  • Gastrointestinal Neoplasms / therapy
  • Humans
  • Male
  • Middle Aged
  • Multimodal Imaging*
  • Positron-Emission Tomography
  • Radiopharmaceuticals
  • Reproducibility of Results
  • Tomography, X-Ray Computed
  • United States

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18