Assessment of myocardial adrenergic innervation with a solid-state dedicated cardiac cadmium-zinc-telluride camera: first clinical experience

Eur Heart J Cardiovasc Imaging. 2014 May;15(5):575-85. doi: 10.1093/ehjci/jet258. Epub 2013 Dec 17.

Abstract

Aims: To investigate the relationships between regional adrenergic innervation heterogeneity, myocardial perfusion, and contractile function obtained by means of a low-dose imaging protocol with a cadmium-zinc-telluride (CZT) dedicated camera.

Methods: Twenty-eight patients with or without ischaemic heart disease underwent (123)I-metaiodobenzylguanidine (MIBG) planar scintigraphic and CZT early and delayed evaluations followed by (99m)Tc-tetrofosmin rest gated CZT with a single-day protocol. The heart-to-mediastinum ratio and the washout rate were computed from planar (123)I-MIBG images. The summed (123)I-MIBG defect scores (SS-MIBG) were semi-quantitatively assessed from CZT images. The summed rest score (SRS), summed motion score (SMS), and summed thickening score (STS) were quantitated from (99m)Tc-tetrofosmin images.

Results: Sixteen patients showed a depressed left ventricular systolic function [ejection fraction (EF)<50%]. They presented higher SRS (P = 0.007), SMS (P < 0.001), STS (P < 0.001), and early SS-MIBG (P = 0.007) values than those with normal contractile function. Interestingly, higher early SS-MIBG values, index of regional sympathetic innervation heterogeneity, clustered with more elevated SRS (P = 0.023), and more impaired measures of regional and global left ventricle systolic function, i.e. SMS (P = 0.046), STS (P = 0.014), and EF (P = 0.027). At multivariate analysis, a higher SRS (P = 0.039) remained the only independent predictor of more elevated early SS-MIBG values. In the 20 of 28 ischaemic patients, the correlations between early SS-MIBG and SMS (P = 0.017) and also STS (P = 0.036) were further confirmed. The effective dose of the investigation was 4.2 ± 0.72 mSv.

Conclusions: An altered early SS-MIBG, assessed with a low-dose imaging protocol and a CZT cardiac camera, identifies patients with more impaired myocardial perfusion and contractile function.

Keywords: 123I-MIBG; CZT camera; Cardiac innervation; Heart-to-mediastinum ratio.

MeSH terms

  • 3-Iodobenzylguanidine
  • Adrenergic Neurons*
  • Aged
  • Cadmium
  • Female
  • Heart / diagnostic imaging
  • Heart / innervation*
  • Humans
  • Male
  • Middle Aged
  • Myocardial Ischemia / diagnostic imaging*
  • Organophosphorus Compounds
  • Organotechnetium Compounds
  • Radiographic Image Interpretation, Computer-Assisted
  • Radiopharmaceuticals
  • Tellurium
  • Tomography, Emission-Computed, Single-Photon / methods*
  • Zinc

Substances

  • CdZnTe
  • Organophosphorus Compounds
  • Organotechnetium Compounds
  • Radiopharmaceuticals
  • Cadmium
  • 3-Iodobenzylguanidine
  • technetium tc-99m tetrofosmin
  • Zinc
  • Tellurium