Is the image quality of I-124-PET impaired by an automatic correction of prompt gammas?

PLoS One. 2013 Aug 27;8(8):e71729. doi: 10.1371/journal.pone.0071729. eCollection 2013.

Abstract

Objectives: The aim of this study is to evaluate the quality of I-124 PET images with and without prompt gamma compensation (PGC) by comparing the recovery coefficients (RC), the signal to noise ratios (SNR) and the contrast to F-18 and Ga-68. Furthermore, the influence of the PGC on the quantification and image quality is evaluated.

Methods: For measuring the image quality the NEMA NU2-2001 PET/SPECT-Phantom was used containing 6 spheres with a diameter between 10 mm and 37 mm placed in water with different levels of background activity. Each sphere was filled with the same activity concentration measured by an independently cross-calibrated dose calibrator. The "hot" sources were acquired with a full 3D PET/CT (Biograph mCT®, Siemens Medical USA). Acquisition times were 2 min for F-18 and Ga-68, and 10 min for I-124. For reconstruction an OSEM algorithm was applied. For I-124 the images were reconstructed with and without PGC. For the calculation of the RCs the activity concentrations in each sphere were determined; in addition, the influence of the background correction was studied.

Results: The RCs of Ga-68 are the smallest (79%). I-124 reaches similar RCs (87% with PGC, 84% without PGC) as F-18 (84%). showing that the quantification of I-124 images is similar to F-18 and slightly better than Ga-68. With background activity the contrast of the I-124 PGC images is similar to Ga-68 and F-18 scans. There was lower background activity in the I-124 images without PGC, which probably originates from an overcorrection of the scatter contribution. Consequently, the contrast without PGC was much higher than with PGC. As a consequence PGC should be used for I-124.

Conclusions: For I-124 there is only a slight influence on the quantification depending on the use of the PGC. However, there are considerable differences with respect to I-124 image quality.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Fluorine Radioisotopes / chemistry
  • Gallium Radioisotopes / chemistry
  • Image Processing, Computer-Assisted / standards*
  • Iodine Radioisotopes / chemistry
  • Phantoms, Imaging
  • Positron-Emission Tomography / standards*
  • Radiopharmaceuticals / chemistry
  • Reference Standards
  • Signal-To-Noise Ratio
  • Tomography, Emission-Computed, Single-Photon / standards

Substances

  • Fluorine Radioisotopes
  • Gallium Radioisotopes
  • Iodine Radioisotopes
  • Radiopharmaceuticals