Split-dose technique for FDG PET/CT-guided percutaneous ablation: a method to facilitate lesion targeting and to provide immediate assessment of treatment effectiveness

Radiology. 2013 Jul;268(1):288-95. doi: 10.1148/radiol.13121462. Epub 2013 Apr 5.

Abstract

Purpose: To describe a split-dose technique for fluorine 18 fluorodeoxyglucose (FDG) positron emission tomography (PET)/computed tomography (CT)-guided ablation that permits both target localization and evaluation of treatment effectiveness.

Materials and methods: Institutional review board approved the study with a waiver of consent. From July to December 2011, 23 patients (13 women, 10 men; mean age, 59 years; range, 35-87 years) with 29 FDG-avid tumors (median size, 1.4 cm; range, 0.6-4.4 cm) were targeted for ablation. The location of the lesion was the liver (n = 23), lung (n = 4), adrenal gland (n = 1), and thigh (n = 1). Radiofrequency ablation was performed in 17 lesions; microwave ablation, in six; irreversible electroporation, in five; and cryoablation, in one. The pathologic condition of the tumor was metastatic colorectal adenocarcinoma in 18 lesions, primary hepatocellular carcinoma in one lesion, and a variety of metastatic tumors in the remaining 10 lesions. A total of 4 mCi (148 MBq) of FDG was administered before the procedure for localization and imaging guidance. At completion of the ablation, an additional 8 mCi (296 MBq) of FDG was administered to assess ablation adequacy. Results of subsequent imaging follow-up were used to determine if postablation imaging after the second dose of FDG reliably helped predict complete tumor ablation. Descriptive statistics were used to summarize the results.

Results: Twenty-eight of 29 (97%) ablated lesions showed no residual FDG activity after the second intraprocedural FDG dose. One patient with residual activity underwent immediate biopsy that revealed residual viable tumor and was immediately re-treated. Follow-up imaging at a median of 155 days (range, 92-257 days) after ablation showed local recurrences in two (7%) lesions that were originally negative at postablation PET.

Conclusion: Split-dose FDG PET/CT may be a useful tool to provide both guidance and endpoint evaluation, allowing an opportunity for repeat intervention if necessary. Further work is necessary to validate these concepts.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Catheter Ablation / methods*
  • Female
  • Fluorodeoxyglucose F18
  • Humans
  • Male
  • Middle Aged
  • Multimodal Imaging*
  • Neoplasms / diagnostic imaging*
  • Neoplasms / surgery*
  • Positron-Emission Tomography*
  • Radiopharmaceuticals
  • Retrospective Studies
  • Tomography, X-Ray Computed*

Substances

  • Radiopharmaceuticals
  • Fluorodeoxyglucose F18