Low-dose thallium-201 protocol with a cadmium-zinc-telluride cardiac camera

Nucl Med Commun. 2012 May;33(5):464-9. doi: 10.1097/MNM.0b013e3283504543.

Abstract

Objectives: Thallium-201 is efficient for myocardial perfusion imaging, but leads to relatively high radiation exposure in patients. The purpose of this study was to compare the efficiency of low-dose thallium-201 imaging with cadmium-zinc-telluride (CZT) cameras with regular-dose thallium-201 imaging with conventional cameras.

Methods: We prospectively studied 137 consecutive patients referred for stress myocardial perfusion imaging who had previously had a myocardial single-photon emission computed tomography with thallium-201. We injected at stress a low dose of thallium-201 (1.1 MBq/kg, 28 µCi/kg), performed a 5-7 min scan with a CZT camera (GE DNM 530c), and assessed redistribution imaging when the initial images were abnormal. We compared the CZT scan with the conventional dual-head tomographic camera scan taken previously with a regular dose of thallium-201.

Results: The average delay between both scans was 22 months. The stress dose was 88 ± 16 (2.38 ± 0.43 mCi) versus 125 ± 13 MBq (3.38 ± 0.34 mCi; a 30% reduction). The time for camera acquisition was 6 versus 13 min (a 54% reduction). The myocardial counts were increased two-fold with CZT (mean: 446 Kcounts). The quality of CZT images was better in 69% of the cases. There were 59 artifacts with conventional cameras and 29 artifacts with CZT (P<0.01). The diagnostic agreement was calculated in patients without clinical or angiographic changes between both scans (115 patients) and was high (97%). The effective dose at stress was less than 12 mSv. High myocardial counting allowed for further decrease in the injected activity, leading to an effective dose as low as 8 mSv.

Conclusion: With reduced activities of thallium-201 and low effective doses, the CZT camera provides reliable, high-quality imaging.

Publication types

  • Comparative Study

MeSH terms

  • Aged
  • Cadmium
  • Female
  • Humans
  • Male
  • Middle Aged
  • Myocardial Perfusion Imaging / instrumentation*
  • Prospective Studies
  • Radiation Dosage
  • Reproducibility of Results
  • Tellurium
  • Thallium Radioisotopes / administration & dosage*
  • Zinc

Substances

  • CdZnTe
  • Thallium Radioisotopes
  • Cadmium
  • Zinc
  • Tellurium