Multiphase contrast-enhanced CT with highly concentrated contrast agent can be used for PET attenuation correction in integrated PET/CT imaging

Eur J Nucl Med Mol Imaging. 2012 Feb;39(2):316-25. doi: 10.1007/s00259-011-1919-5. Epub 2011 Nov 29.

Abstract

Purpose: State-of-the-art positron emission tomography/computed tomography (PET/CT) systems incorporate multislice CT technology, thus facilitating the acquisition of multiphase, contrast-enhanced CT data as part of integrated PET/CT imaging protocols. We assess the influence of a highly concentrated iodinated contrast medium (CM) on quantification and image quality following CT-based attenuation correction (CT-AC) in PET/CT.

Methods: Twenty-eight patients with suspected malignant liver lesions were enrolled prospectively. PET/CT was performed 60 min after injection of 400 MBq of (18)F-fluorodeoxyglucose (FDG) and following the biphasic administration of an intravenous CM (400 mg iodine/ml, Iomeron 400). PET images were reconstructed with CT-AC using any of four acquired CT image sets: non-enhanced, pre-contrast (n-PET), arterial phase (art-PET), portal venous phase (pv-PET) and late phase (late-PET). Normal tissue activity and liver lesions were assessed visually and quantitatively on each PET/CT image set.

Results: Visual assessment of PET following CT-AC revealed no noticeable difference in image appearance or quality when using any of the four CT data sets for CT-AC. A total of 44 PET-positive liver lesions was identified in 21 of 28 patients. There were no false-negative or false-positive lesions on PET. Mean standardized uptake values (SUV) in 36 evaluable lesions were: 5.5 (n-PET), 5.8 (art-PET), 5.8 (pv-PET) and 5.8 (late-PET), with the highest mean increase in mean SUV of 6%. Mean SUV changes in liver background increased by up to 10% from n-PET to pv-PET.

Conclusion: Multiphase CT data acquired with the use of highly concentrated CM can be used for qualitative assessment of liver lesions in torso FDG PET/CT. The influence on quantification of FDG uptake is small and negligible for most clinical applications.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Contrast Media / pharmacology*
  • Diagnostic Imaging / methods
  • False Positive Reactions
  • Female
  • Fluorodeoxyglucose F18 / pharmacology*
  • Humans
  • Image Processing, Computer-Assisted
  • Male
  • Middle Aged
  • Multimodal Imaging / methods*
  • Positron-Emission Tomography / methods*
  • Tomography, X-Ray Computed / methods*

Substances

  • Contrast Media
  • Fluorodeoxyglucose F18