Variable-density spiral-in/out functional magnetic resonance imaging

Magn Reson Med. 2011 May;65(5):1287-96. doi: 10.1002/mrm.22722. Epub 2010 Nov 30.

Abstract

A variable-density spiral k-space trajectory is introduced for brain functional magnetic resonance imaging. The proposed spiral trajectory consists of an Archimedean spiral from the k-space origin to an arbitrary fraction r of the maximum k-space radius, extending beyond this point with a variable-density spiral in which the sampling density decreases as the k-space radius increases. It, therefore, permits a reduction in readout time at the expense of undersampling only the high spatial frequencies, in which the energy in T(2)*-weighted brain images is low. The trajectory was implemented in a two-dimensional spiral-in/out sequence, and single-shot high-resolution (1.72 × 1.72 mm(2) in-plane) functional magnetic resonance imaging data were acquired from human volunteers. Compared with a two-shot fully Archimedean spiral sequence with the same spatial coverage and total scan time, the variable-density sequence yielded greater activation magnitudes with improved temporal efficiency and minor artifacts.