Semi-automated algorithm for calculating heart-to-mediastinum ratio in cardiac Iodine-123 MIBG imaging

J Nucl Cardiol. 2011 Feb;18(1):82-9. doi: 10.1007/s12350-010-9313-4. Epub 2010 Nov 23.

Abstract

Background: We have developed novel software for semi-automatically measuring heart-to-mediastinum (H/M) ratio in cardiac Iodine-123 (¹²³I)-labeled meta-iodobenzylguanidine (MIBG) imaging. Our aim is to improve the reproducibility of the H/M ratio using the semi-automated method as opposed to the manual method.

Methods and results: The software algorithm automatically determined the mediastinal region of interest (ROI) using information from ¹²³I-MIBG uptake of the heart, lung, liver, and thyroid after a cardiac circular ROI was manually set. A total of 37 patients who underwent both early and delayed ¹²³I-MIBG imaging was retrospectively selected. The heart-to-mediastinum (H/M) ratios were calculated by both semi-automated and manual methods and assessed for the intra- and inter-observer variability. All H/M ratios were classified into three groups: normal, slight abnormality, and severe abnormality. The H/M ratios with the new method were higher than those obtained manually (P < .001). In the test-retest reliability, the intra-class correlation coefficient from the semi-automated method showed excellent reproducibility for early (0.99) and delayed (0.99) imaging. The Bland-Altman plots demonstrated better agreement using the semi-automated method (a range of 95% limits -0.11 to 0.10) than that in the manual method (-0.34 to 0.27). The inter-observer agreement was also good using the semi-automated method (κ = 0.866).

Conclusions: The H/M ratio using the semi-automated method showed high reproducibility in both early and delayed imaging.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • 3-Iodobenzylguanidine* / pharmacokinetics
  • Algorithms*
  • Female
  • Heart Diseases / diagnostic imaging*
  • Heart Diseases / metabolism
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Male
  • Mediastinum / diagnostic imaging*
  • Mediastinum / physiopathology
  • Middle Aged
  • Myocardium / metabolism
  • Observer Variation
  • Pattern Recognition, Automated / methods*
  • Positron-Emission Tomography / methods
  • Radiopharmaceuticals / pharmacokinetics
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tissue Distribution

Substances

  • Radiopharmaceuticals
  • 3-Iodobenzylguanidine