Gated myocardial perfusion SPECT underestimates left ventricular volumes and shows high variability compared to cardiac magnetic resonance imaging -- a comparison of four different commercial automated software packages

BMC Med Imaging. 2010 May 25:10:10. doi: 10.1186/1471-2342-10-10.

Abstract

Background: We sought to compare quantification of left ventricular volumes and ejection fraction by different gated myocardial perfusion SPECT (MPS) programs with each other and to magnetic resonance (MR) imaging.

Methods: N = 100 patients with known or suspected coronary artery disease were examined at rest with 99 mTc-tetrofosmin gated MPS and cardiac MR imaging. Left ventricular end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV) and ejection fraction (EF) were obtained by analysing gated MPS data with four different programs: Quantitative Gated SPECT (QGS), GE MyoMetrix, Emory Cardiac Toolbox (ECTb) and Exini heart.

Results: All programs showed a mean bias compared to MR imaging of approximately -30% for EDV (-22 to -34%, p < 0.001 for all), ESV (-12 to -37%, p < 0.001 for ECTb, p < 0.05 for Exini, p = ns for QGS and MyoMetrix) and SV (-21 to -41%, p < 0.001 for all). Mean bias +/- 2 SD for EF (% of EF) was -9 +/- 27% (p < 0.01), 6 +/- 29% (p = ns), 15 +/- 27% (p < 0.001) and 0 +/- 28% (p = ns) for QGS, ECTb, MyoMetrix, and Exini, respectively.

Conclusions: Gated MPS, systematically underestimates left ventricular volumes by approximately 30% and shows a high variability, especially for ESV. For EF, accuracy was better, with a mean bias between -15 and 6% of EF. It may be of value to take this into consideration when determining absolute values of LV volumes and EF in a clinical setting.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Aged
  • Algorithms
  • False Negative Reactions
  • Female
  • Gated Blood-Pool Imaging / methods*
  • Humans
  • Image Enhancement / methods
  • Image Interpretation, Computer-Assisted / methods*
  • Imaging, Three-Dimensional / methods
  • Magnetic Resonance Imaging, Cine / methods*
  • Male
  • Middle Aged
  • Pattern Recognition, Automated / methods
  • Perfusion Imaging / methods
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Software Validation
  • Software*
  • Stroke Volume*
  • Tomography, Emission-Computed, Single-Photon / methods*
  • Ventricular Dysfunction, Left / diagnosis*