Motion correction of PET brain images through deconvolution: II. Practical implementation and algorithm optimization

Phys Med Biol. 2009 Feb 7;54(3):813-29. doi: 10.1088/0031-9155/54/3/022. Epub 2009 Jan 9.

Abstract

Image quality is significantly degraded even by small amounts of patient motion in very high-resolution PET scanners. When patient motion is known, deconvolution methods can be used to correct the reconstructed image and reduce motion blur. This paper describes the implementation and optimization of an iterative deconvolution method that uses an ordered subset approach to make it practical and clinically viable. We performed ten separate FDG PET scans using the Hoffman brain phantom and simultaneously measured its motion using the Polaris Vicra tracking system (Northern Digital Inc., Ontario, Canada). The feasibility and effectiveness of the technique was studied by performing scans with different motion and deconvolution parameters. Deconvolution resulted in visually better images and significant improvement as quantified by the Universal Quality Index (UQI) and contrast measures. Finally, the technique was applied to human studies to demonstrate marked improvement. Thus, the deconvolution technique presented here appears promising as a valid alternative to existing motion correction methods for PET. It has the potential for deblurring an image from any modality if the causative motion is known and its effect can be represented in a system matrix.

Publication types

  • Evaluation Study

MeSH terms

  • Algorithms*
  • Artifacts*
  • Computer Simulation
  • Humans
  • Image Enhancement / methods*
  • Image Interpretation, Computer-Assisted / methods*
  • Models, Neurological*
  • Movement*
  • Positron-Emission Tomography / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity