Fully three-dimensional reconstruction for a PET camera with retractable septa

IEEE Trans Med Imaging. 1991;10(4):505-12. doi: 10.1109/42.108584.

Abstract

A fully 3-D reconstruction algorithm has been developed to reconstruct data from a 16 ring PET camera (a Siemens/CTI 953B) with automatically retractable septa. The tomograph is able to acquire coincidences between any pair of detector rings and septa retraction increases the total system count rate by a factor of 7.8 (including scatter) and 4.7 (scatter subtracted) for a uniform, 20 cm diameter cylinder. The reconstruction algorithm is based on 3-D filtered backprojection, expressed in a form suitable for the multi-angle sinogram data. Sinograms which are not measured due to the truncated cylindrical geometry of the tomograph, but which are required for a spatially invariant response function, are obtained by forward projection. After filtering, the complete set of sinograms is backprojected into a 3-D volume of 128x128x31 voxels using a voxel-driven procedure. The algorithm has been validated with simulation, and tested with both phantom and clinical data from the 953B.