CT findings after radiofrequency ablation in rabbit livers: comparison of internally cooled electrodes, perfusion electrodes, and internally cooled perfusion electrodes

J Vasc Interv Radiol. 2007 Nov;18(11):1417-27. doi: 10.1016/j.jvir.2007.08.010.

Abstract

Purpose: To compare the computed tomography (CT) findings of radiofrequency (RF) ablation of rabbit livers with the use of internally cooled electrodes, perfusion electrodes, and internally cooled perfusion (ICP) electrodes.

Materials and methods: RF ablation zones were created in 24 rabbit livers in vivo in three groups, each treated at 30 W for 30 minutes with an electrode with a 1-cm active portion: one group was treated with an internally cooled electrode (group A), one with a perfusion electrode (group B), and one with an ICP electrode (group C). In the latter two groups, 0.9% saline solution mixed with contrast medium (9:1 ratio) was infused at a rate of 1 mL/min. Immediate CT images were ascertained in all groups and follow-up CT was performed in the latter two groups. CT findings of the RF ablation zone and surrounding liver parenchyma were evaluated and compared among groups.

Results: On immediate unenhanced CT, infused contrast medium was noted in the center of the ablation zone without extravasation. Marginal linear or tubular low densities of the RF ablation zone were noted only in groups B (n=4) and C (n=3). Liver infarction was more frequently noted in groups B (n=5) and C (n=7) than in group A (n=3). Periportal tracking was noted more frequently in groups B (n=7) and C (n=8) than in group A (n=1). Pericaval or perihepatic tracking was noted only in groups B (n=3) and C (n=3). Extensive portal venous gas was noted in one animal in group C.

Conclusions: On CT, RF ablation with the perfusion or ICP electrode seemed to show more severe surrounding parenchymal changes than RF ablation with an internally cooled electrode.

Publication types

  • Comparative Study
  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Catheter Ablation / instrumentation*
  • Catheter Ablation / methods
  • Cold Temperature
  • Electrodes*
  • Equipment Design
  • Equipment Failure Analysis
  • Liver / diagnostic imaging*
  • Male
  • Perfusion / instrumentation*
  • Perfusion / methods
  • Rabbits
  • Tomography, X-Ray Computed / methods*