7 T body MRI: B1 shimming with simultaneous SAR reduction

Phys Med Biol. 2007 Sep 7;52(17):5429-41. doi: 10.1088/0031-9155/52/17/022. Epub 2007 Aug 21.

Abstract

The high frequency of the radiofrequency (RF) fields used in high field magnetic resonance imaging (MRI) results in electromagnetic field variations that can cause local regions to have a large specific absorption rate (SAR) and/or a low excitation. In this study, we evaluated the use of a B1 shimming technique which can simultaneously improve the B+1 homogeneity and reduce the SAR for whole body imaging at 7 T. Optimizations for four individual anatomies showed a reduction up to 74% of the peak SAR values with respect to a quadrature excitation and a simultaneous improvement of the B+1 homogeneity varying between 39 and 75% for different optimization parameters. The average SAR was reduced with approximately 50% for all optimizations. The optimized phase and amplitude settings from an elliptical phantom model were applied to four realistic human anatomy models to evaluate whether a generic application without prior knowledge of the detailed human anatomy is possible. This resulted in an average improvement of the B+1 homogeneity of 37% and an average reduction of the maximum and average SAR of 50 and 55%, respectively. It can be concluded that this generic method can be used as a simple method to improve the prospects of 7 T body imaging.