Detection of small pulmonary nodules in high-field MR at 3 T: evaluation of different pulse sequences using porcine lung explants

Eur Radiol. 2007 May;17(5):1341-51. doi: 10.1007/s00330-006-0422-z. Epub 2006 Sep 30.

Abstract

To evaluate two MR imaging sequences for the detection of artificial pulmonary nodules inside porcine lung explants. 67 agarose nodules ranging 3-20 mm were injected into ten porcine lungs within a dedicated chest phantom. The signal on T1-weighted images and radiopacity were adjusted by adding 0.125 mmol/l Gd-DTPA and 1.5 g/l of iodine. A T1-weighted three-dimensional gradient-echo (T1-3D-GRE; TR/TE:3.3/1.1 ms, slice:8 mm, flip-angle:10 degrees ) and a T2-weighted half-Fourier fast-spin echo sequence (T2-HF-FSE; TR/TE:2000/66 ms, slice:7 mm, flip-angle:90 degrees ) were applied in axial orientation using a 3-T system (Intera, Philips Medical Systems, Best, The Netherlands), followed by CT (16x0.5 mm) as reference. Nodule sizes and locations were assessed by three blinded observers. In nodules of >10 mm, sensitivity was 100% using 3D-GRE-MRI and 94% using the HF-FSE sequence. For nodules 6-10 mm, the sensitivity of MRI was lower than with CT (3D-GRE:92%; T2-HF-FSE:83%). In lesions smaller than 5 mm, the sensitivity declined to 80% (3D-GRE) and 53% (HF-FSE). Small lesion diameters were overestimated with both sequences, particularly with HF-FSE. This study confirms the feasibility of 3 T-MRI for lung nodule detection. In lesions greater than 5 mm, the sensitivity of the 3D-GRE sequence approximated CT (>90%), while sensitivity and PPV with the HF-FSE sequence were slightly inferior.

Publication types

  • Evaluation Study

MeSH terms

  • Animals
  • Confidence Intervals
  • Contrast Media
  • Gadolinium DTPA
  • Image Processing, Computer-Assisted
  • In Vitro Techniques
  • Lung Diseases / pathology*
  • Magnetic Resonance Imaging / methods*
  • Phantoms, Imaging
  • Sensitivity and Specificity
  • Swine
  • Tomography, X-Ray Computed

Substances

  • Contrast Media
  • Gadolinium DTPA