Left ventricular systolic/diastolic function evaluated by quantitative ECG-gated SPECT: comparison with echocardiography and plasma BNP analysis

Ann Nucl Med. 2005 Sep;19(6):447-54. doi: 10.1007/BF02985571.

Abstract

Objective: The aim of this study was to evaluate the left ventricular (LV) functional parameters calculated using quantitative electrocardiography (ECG)-gated myocardial perfusion single photon emission computed tomography (QGS). In addition to LV systolic parameters, diastolic parameters were compared with those by ultrasound echocardiography (UCG) and also with plasma B-type natriuretic peptide (BNP) concentrations.

Methods: We examined 46 patients with various forms of heart disease. By the QGS data with 16 framing data acquisition using technetium (Tc)-99m methoxyisobutylisonitrile (MIBI) perfusion, we calculated the following parameters: LV end-diastolic volume (EDV), end-systolic volume (ESV), ejection fraction (EF), peak filling rate (PFR), filling rate during the first third of the filling time (1/3FR) and first third filling fraction (1/3FF). By UCG, we measured mitral early to atrial (E/A) wave velocity ratio and pulmonary venous inflow systolic/diastolic (S/D) ratio as diastolic functional parameters. Plasma BNP concentrations were also measured.

Results: There was a significant correlation between LVEDV, ESV and EF measured by QGS and UCG (EDV, r = 0.71, p < 0.001; ESV, r = 0.82, p < 0.001; EF, r = 0.75, p < 0.001). The PFR, 1/3FR and 1/3FF obtained by QGS correlated positively with E/A ratio (PFR, r = 0.54, p < 0.001; 1/3FR, r = 0.61, p < 0.001; 1/3FF, r = 0.42, p < 0.01) and negatively with S/D ratio (PFR, r = -0.40, p < 0.01; 1/3FR, r = -0.38, p < 0.05; 1/3FF, r = -0.39, p < 0.01) obtained by UCG. Plasma BNP concentrations in EF < 50% patients were greater than those in EF > or = 50% patients (335.2 +/- 60.2 vs. 101.2 +/- 41.3 pg/ml, p < 0.01, both n = 17). Plasma BNP levels were also compared between higher and lower 1/3FF patients matched for LVEF. Plasma BNP concentrations in 1/3FF < 35% patients were significantly greater than those in 1/3FF > or = 35% patients (312.9 +/- 62.5 vs. 120.5 +/- 32.8 pg/ml, p < 0.05, both n = 14).

Conclusions: The degree of LV systolic and diastolic dysfunctions evaluated by QGS correlated with that by UCG or BNP. The QGS functional parameters offer useful information regarding cardiac failure.

Publication types

  • Clinical Trial
  • Controlled Clinical Trial
  • Validation Study

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Aged, 80 and over
  • Echocardiography / methods*
  • Female
  • Gated Blood-Pool Imaging / methods*
  • Humans
  • Male
  • Middle Aged
  • Natriuretic Peptide, Brain / blood*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Statistics as Topic
  • Tomography, Emission-Computed, Single-Photon / methods*
  • Ventricular Dysfunction, Left / blood*
  • Ventricular Dysfunction, Left / diagnosis*

Substances

  • Natriuretic Peptide, Brain