Accuracy and reliability of quantitative measurements in coronary arteries by multi-slice computed tomography: experimental and initial clinical results

Clin Radiol. 2001 Jun;56(6):466-74. doi: 10.1053/crad.2001.0687.

Abstract

Aim: To evaluate the accuracy of non-invasive measurements within coronary arteries by multi-slice computed tomography (MSCT). We present experimental as well as clinical data.

Materials and methods: Silicon tubes simulating coronary arteries (outer diameter 6 mm, lumen diameter within stenotic area 2 mm) were used for experimental studies. Clinical data were derived from 15 patients in whom vessel diameters were assessed by MSCT, intracoronary ultrasound (ICUS) and quantitative coronary angiography (QCA). MSCT were performed in a Somatom Volume Zoom(trade mark)CT system (Siemens, Forchheim, Germany) at 2 collimated slice widths (2.5 mm, 1.0 mm).

Results: Outer silicon tube diameters were overestimated by MSCT (6.56 mm +/- 0.32 mm). All measurements revealed significantly better results on 1.0 collimation compared to 2.5 mm collimation (outer diameter: 6.36 mm +/- 0.22 mm vs 6.76 mm +/- 0.27 mm, P < 0.0001; lumen diameters: 1.83 mm +/- 0.14 mm vs 1.51 mm +/- 0.19 mm, P < 0.0001). The comparison of vessel diameters within human coronary arteries revealed comparable results between ICUS and MSCT (4.89 mm +/- 0.67 mm vs 4.91 mm +/- 0.71 mm, P = 0.79, r = 0.79, P < 0.0001). QCA-measurements showed significantly lower results (3.67 +/- 0.71, P < 0.0001, r = 0.62, P < 0.001).

Conclusions: Experimental as well as initial clinical results indicate acceptable reliability and accuracy of quantitative measurements by MSCT, when using thin collimated slice widths. Partial volume effects lead to a systematic overestimation of vessel size. MSCT has the potential to become an important non-invasive diagnostic tool in patients with coronary artery disease.

MeSH terms

  • Coronary Angiography / instrumentation
  • Coronary Angiography / methods
  • Coronary Angiography / standards*
  • Humans
  • Phantoms, Imaging
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / instrumentation
  • Tomography, X-Ray Computed / methods
  • Tomography, X-Ray Computed / standards*