Background: We compared the reproducibility of thallium 201 and technetium 99m sestamibi (MIBI) gated single photon emission computed tomography (SPECT) measurement of myocardial function using the Germano algorithm (J Nucl Med 1995;36:2138-47).
Methods and results: Gated SPECT acquisition was repeated in the same position in 30 patients who received Tl-201 and in 26 who received Tc-99m-MIBI. The quantification of end-diastolic volume (EDV), end-systolic volume (ESV), and ejection fraction (EF) on Tl-201 and Tc-99m-MIBI gated SPECT was processed independently with Cedars-Sinai QGS (Quantitative Gated SPECT) software. The reproducibility of the measurement of ventricular function on Tl-201 gated SPECT was compared with that of Tc-99m-MIBI gated SPECT. Correlation between the 2 measurements for volumes and EF was excellent for the repeated gated SPECT studies of Tl-201 (r = 0.928 to 0.986, P <.05) and Tc-99m-MIBI (r = 0.979 to 0.997, P <.05). However, Bland-Altman analysis revealed the 95% limits of agreement (2 SDs) for volumes and EF were narrower by repeated Tc-99m-MIBI gated SPECT (EDV 14.1 mL, ESV 9.4 mL, EF 5.5%) than by repeated Tl-201 gated SPECT (EDV 24.1 mL, ESV 18.6 mL, EF 10.3%). The root-mean-square values of the coefficient of variation for volumes and EF were smaller by repeated Tc-99m-MIBI gated SPECT (EDV 2.1 mL, ESV 2.7 mL, EF 2.3%) than by repeated Tl-201 gated SPECT (EDV 3.2 mL, ESV 3.5 mL, EF 5.2%).
Conclusions: QGS provides an excellent correlation between repeated gated SPECT with Tl-201 and Tc-99m-MIBI. However, Tc-99m-MIBI provides more reproducible volumes and EF than Tl-201. Tc-99m-MIBI gated SPECT is the preferable method for the clinical monitoring of ventricular function.