[Radiation exposure to patients caused by single-photon transmission measurement in PET]

Nuklearmedizin. 2000 Nov;39(7):204-8.
[Article in German]

Abstract

Aim: The aim of the study was the determination of the radiation exposure to the patient caused by single-photon transmission measurement for 3D whole-body PET.

Material and method: Single-photon-transmission measurement is performed using two Cs-137 point-sources (E gamma = 662 keV, A = 2*614 MBq) on a 3D PET scanner (ECAT ART). During a simulation of a whole body transmission scan (axial length: 75 cm, 6 contigous bed positions) dose measurements with thermoluminescent dosimeters were carried out using a thorax and an abdomen phantom. Following the guidelines of the ICRU report No. 60 an estimation of the effective dose caused by a single-photon transmission measurement was calculated.

Results: For a total acquisition time of 360 min (6 beds with an acquisition time of 60 min per bed) the absorbed doses amounted to: surface (xyphoid) 189 microGy, heart 196 microGy, lungs 234 microGy, vertebra 240 microGy, liver 204 microGy, gonads 205 microGy, thyroid 249 microGy and bladder 185 microGy resulting in a conversion factor of 1.7*10(-4) mSv/(h*MBq). The estimation of the effective dose for a patient's transmission (acquisition time of 3.2 min per bed) yields a value of 11 microSv. An estimation of the ratio of the conversion factors for transmission measurements in single-photon- and in coincidence mode (two Ge-68/Ga-68 rod sources of 40 MBq each), respectively, resulted in a value of 0.18. The comparison of the effective doses caused by single-photon transmission and by emission measurement (injection of 250 MBq of FDG) yields a ratio of 2.3*10(-3).

Conclusion: The radiation exposure of the patient caused by the transmission measurement for 3D whole-body-PET can be neglected. In comparison with the coincidence-transmission using uncollimated line sources of low activity the radiation exposure is still reduced using single photon trans-mission with collimated point sources of high activity.

Publication types

  • English Abstract

MeSH terms

  • Computer Simulation
  • Humans
  • Phantoms, Imaging*
  • Radiation Dosage*
  • Tomography, Emission-Computed
  • Tomography, Emission-Computed, Single-Photon* / instrumentation
  • Tomography, Emission-Computed, Single-Photon* / methods
  • Whole-Body Irradiation