1 March 2008 Noninvasive Raman tomographic imaging of canine bone tissue
Matthew V. Schulmerich, Jacqueline H. Cole, Kathryn A. Dooley, Michael D. Morris, Jaclynn Kreider, Steven A. Goldstein, Subhadra Srinivasan, Brian W. Pogue
Author Affiliations +
Abstract
Raman spectroscopic diffuse tomographic imaging has been demonstrated for the first time. It provides a noninvasive, label-free modality to image the chemical composition of human and animal tissue and other turbid media. This technique has been applied to image the composition of bone tissue within an intact section of a canine limb. Spatially distributed 785-nm laser excitation was employed to prevent thermal damage to the tissue. Diffuse emission tomography reconstruction was used, and the location that was recovered has been confirmed by micro-computed tomography (micro-CT) images.

With recent advances, diffuse tomography shows promise for in vivo clinical imaging.1, 2 In principle, algorithms developed for fluorescence imaging in tissue can be applied to Raman signals. Although the Raman effect is weaker than fluorescence, the scattered signal is detectable, and thus tomography is achievable. Here we demonstrate the first diffuse tomography reconstructions based on Raman scatter.

Raman mapping and imaging are well-established techniques for examining material surfaces.3 Subsurface mapping of simple planar objects was reported recently4, 5 using fiber optic probes with spatially separated injection and collection fibers.6 Noninvasive measurements of bone Raman spectra were demonstrated at depths of 5mm below the skin.5

Bone is promising for Raman tomography because the spectra are rich in compositional information,7 which reflects bone maturity and health. Spectroscopically measured bone composition changes have been correlated with aging8 and susceptibility to osteoporotic fracture.9 The Raman spectrum of bone mineral is easily distinguished from the spectra of proteins and other organic tissue constituents, facilitating recovery of even weak signals by multivariate techniques.

Assessments of bone quantity and quality are essential to detect and monitor fracture risk and fracture healing with disease or injury. Common sites for fracture with osteoporosis are the spine, proximal femur, and distal radius. Stress fractures are most frequently seen in the weight-bearing sites of the tibia and metatarsals. Fracture risk depends on bone geometry, architecture, and material properties, as well as the nature of applied load (magnitude, rate, and direction). As a result, noninvasive imaging and nondestructive analysis methods have been developed to assess many of these bone attributes that are increasingly important to clinical practice and basic research in orthopedics.10 Current clinical in vivo methods include dual-energy x-ray absorptiometry (DXA), quantitative computed tomography (QCT), magnetic resonance imaging (MRI), ultrasound, and most recently, high-resolution peripheral QCT. Ex vivo analyses of bone specimens from patients or animals have also utilized these and other techniques.

In this study, we couple micro-computed tomography (micro-CT) and diffuse optical tomography with Raman spectroscopy to recover spatial and composition information from bone tissue ex vivo. We demonstrate the first reconstruction-based recovery of Raman signals through thick tissues to yield molecular information about subsurface bone tissue. Reconstructions from transcutaneous Raman measurements are challenging, because layers of skin, muscle, fat, and connective tissue lie over the bone sites of interest. These layers have different optical properties and thus variably scatter and polarize the injected light.

We chose a canine model because of specimen availability and a bone size similar to human bone. We selected the tibia, a site that is clinically important and has relatively few overlying soft tissues. Measurements were made on the medial surface, where the only additional optical barrier is the crural extensor retinaculum ligament. The canine hind limb was harvested from an animal euthanized in an approved (UCUCA) University of Michigan study. The section of the limb distal to the knee was excised and scanned using in vivo micro-CT (eXplore Locus RS, GE Healthcare, Ontario, Canada). The tibia was scanned at 80kV and 450μA with an exposure time of 100ms using a 360-deg scan technique. The image was reconstructed at a 93-μm voxel resolution [Fig. 1a ].

Fig. 1

Development of mesh for tomographic reconstruction. (a) Micro-CT image of canine hind limb section. (b) Raman spectra of limb section. (c) Geometric image of tissues from micro-CT data. (d) Volumetric mesh created from (c). Input coordinates for reconstruction: tomographic projections and their associated scores. Illumination line shown in red, and collection fibers shown in green. (Color online only.)

020506_1_014802jbo1.jpg

The Raman system (785-nm excitation wavelength) and preprocessing and data analysis protocols for the noninvasive recovery of bone spectra have been described previously.5 Briefly, hair was removed, and glycerol was applied for optical clearing. Sixty-second acquisitions were made at 10 different ring inner diameters between 6 and 16mm . Prior to tomographic reconstruction, a bone Raman factor was recovered transcutaneously in a backscattered geometry using data from three ring diameters that maximized bone tissue contributions to the Raman signal. After transcutaneous measurements were completed, overlying tissue was removed, and the exposed bone spectrum was obtained [Fig. 1b]. The cross-correlation coefficient between the transcutaneous bone factor and exposed bone spectrum was 0.97, indicating that factor recovery was successful.

The same laser, spectrograph, and detector were used for tomography. The specimen was mounted on a 360-deg rotation stage, and transmission (180-deg illumination/collection) measurements were taken transcutaneously. The laser ( 200mW at the specimen) was focused to a rectangle (approximately 8×1.5mm ) along the long axis of the tibia. Although the 1.67Wcm2 power density was above the ANSI Z163.1 threshold limit value, no thermal damage was observed. The 5×10 rectangular array of 100-μm core collection fibers were focused on the specimen 180 deg from the laser illumination point [Fig. 1d]. At the spectrograph, the collection fibers were arranged in a line. Thirteen projections were obtained at 8-deg intervals, spanning the range shown in Fig. 2d. Raman scatter was collected for 5min at each projection. The score (dot product) was calculated between the recovered bone factor and the Raman transmission signal collected by each fiber for all projections. These scores were used as inputs for tomographic reconstruction.

Fig. 2

Raman tomographic images of canine bone tissue. (a) Medial and anterior views of soft tissue mesh (white) and bone surface mesh of tibia and calcaneus (turquoise) overlaid with 50% contrast isosurface of the reconstructed three-dimensional (3-D) Raman image of bone (green). (b) Same view as (a) overlaid with 10% contrast isosurface of the reconstructed Raman image of bone (blue). (c) 3-D mesh of limb section (white), including bone (turquoise), illustrating location of the cross section (blue) containing the highest Raman scatter intensity. (d) Raman intensity at cross section in (c) in pseudo-color overlaid on the micro-CT image of the bone, showing range of illumination (red arrows) and collection (green dots) positions. (Color online only.)

020506_1_014802jbo2.jpg

For the reconstruction, a geometric model was created using the micro-CT data. Skin and bone tissues were segmented using thresholding and region-growing tools in the modeling software (Mimics, Materialise Inc., Ann Arbor, Michigan). The modeled outer surface of the canine limb section [Fig. 1c] was used to generate a volumetric grid for computation. The uniform 2-mm resolution volume mesh contained 8218 nodes and 40,969 tetrahedral elements [Fig. 1d]. In the mesh, the bone and background (assumed skin) regions based on segmentation from the CT images were tagged with the appropriate scattering and absorption coefficients. The excitation and emission optical properties were assumed to be identical.

For tomographic reconstruction, calibrations were performed using a tissue phantom (9-mm diameter Teflon sphere placed 10mm below the surface of a cylindrical phantom (diameter 59mm , height 44.5mm ) composed of 1% Intralipid/agarose gel. A data-scaling factor, computed as the average difference between experimental and model-generated measurements from the phantom, was subtracted from the canine tibia dataset to compensate for the model-data mismatch. This approach, which is commonly used in diffuse tomography to scale the dataset to the model and to avoid calibration errors and image artifacts, simply shifts the intensities up or down to match the model but does not affect the spatial recovery within the image.

The Raman reconstructions were based on NIRFAST (http://nir.thayer.dartmouth.edu/), a reconstruction software toolbox developed at Dartmouth for photon emission imaging.2, 11, 12 The toolbox uses a finite element model to solve a set of coupled diffusion equations, which predict the way the excitation fluence and the Raman emission fluence travel through the highly scattering media. This model, known to be approximate in small tissues, is sufficient for small animal studies with length scales similar to the dog leg. A modified iterative Newton’s method minimizing the least squares norm of model-data differences was used for reconstruction. The toolbox was modified to incorporate the line focus of the sources. Since diffuse optical image reconstruction is ill-posed and ill-conditioned, we used regularization to stabilize the inversion. The starting regularization value was 100, relative to the diagonal of the Hessian matrix to be inverted, and this value was based upon the preliminary phantom experiments, as has been done in our diffuse tomography imaging studies previously. However, the reconstruction of the Raman signal by itself did not assume any knowledge of location of inner tissues.

Raman diffuse tomography was used to reconstruct the canine hind limb section (Fig. 2 ). The medial-lateral limb thickness in the region of interest ranged between 24 and 45mm . The signal/noise ratio, measured using the PO43 v1 peak intensity, ranged from 0 to 5.7. Two reconstructed Raman isosurfaces were mapped onto the true location of the tibia [Figs. 2a and 2b]. Reconstructions appeared artificially spherical, possibly due to the diffuse propagation of light through tissue, which inherently results in diffuse tomography images with blurred features, and to Gaussian-shaped object recovery. The peaked intensity in the center of the bone is an artifact of this process. This problem was most apparent at high and low intensities. Isosurface thresholds between 10% [Fig. 2a] and 50% [Fig. 2b] preserved nonspherical shapes. Despite tissue thickness, the periosteal perimeter of the reconstructed image corresponded well to the true bone contour [Figs. 2c and 2d]. The highest Raman bone signal was found in the center of the bone [Fig. 2d], which is likely an artifact of the Gaussian-type blurring that occurs in diffuse tomography. Because light passed through both the tibia and the fibula in few projections, the fibula was not accurately reconstructed.

Without spatial priors in the reconstruction and with the limited mesh resolution, localization of the bone was imperfect. However, our Raman tomography approach provides the means to incorporate a more sophisticated reconstruction, such as with spatial priors. Although we used a diffusion model of light transport, an approximation commonly made for tissue studies, the light was likely not truly diffuse. Future work will focus on coupling more linear transport-forward models into our promising imaging application, along with the use of spatial priors to provide better localization.13

This proof-of-principle study provides evidence that Raman tomographic images can be obtained in vivo. We also show that the integration of Raman tomography measurements with CT (and by extension, MRI) is feasible. Future probe designs will focus on examining composition changes in the tissue. Improved probe geometry will allow us to reduce power density and impose spatial constraints on the recovery of composition properties for tissue imaged by the clinical modality. This dual imaging technique also provides a better estimate of the Raman signal, and the spatial mapping improves the template for accurate image recovery.11, 13

Acknowledgments

Supported through NIH Grant Nos. R01-AR055222 (M.D.M., S.A.G.), NRSA T90-DK070071-03 (J.H.C.), T32-GM008353 (K.A.D), P01-CA080139 (B.W.P., S.S), and R01-CA120368 (B.W.P.).

References

1. 

V. Ntziachristos, E. A. Schellenberger, J. Ripoli, D. Yessayan, E. Graves, A. Bogdanov Jr., L. Josephson, and R. Weissleder, “Visualization of antitumor treatment by means of a fluorescence molecular tomography with an annexin V-Cy5.5 conjugate,” Proc. Natl. Acad. Sci. U.S.A., 101 (33), 12294 –12299 (2004). https://doi.org/10.1073/pnas.0401137101 0027-8424 Google Scholar

2. 

S. C. Davis, B. W. Pogue, H. Dehghani, and K. D. Paulsen, “Contrast-detail analysis characterizing diffuse optical fluorescence tomography image reconstruction,” J. Biomed. Opt., 10 (5), 050501 (2005). https://doi.org/10.1117/1.2114727 1083-3668 Google Scholar

3. 

P. J. Treado and M. P. Nelson, “Raman Imaging,” Handbook of Raman Spectroscopy, 191 –249 Marcel Dekker, Inc., New York (2001). Google Scholar

4. 

M. V. Schulmerich, W. F. Finney, R. A. Fredericks, and M. D. Morris, “Subsurface Raman spectroscopy and mapping using a globally illuminated non-confocal fiber-optic array probe in the presence of Raman photon migration,” Appl. Spectrosc., 60 (2), 109 –114 (2006). https://doi.org/10.1366/000370206776023340 0003-7028 Google Scholar

5. 

M. V. Schulmerich, K. A. Dooley, T. M. Vanasse, S. A. Goldstein, and M. D. Morris, “Subsurface and transcutaneous Raman spectroscopy and mapping using concentric illumination rings and collection with a circular fiber optic array,” Appl. Spectrosc., 61 671 –678 (2007). 0003-7028 Google Scholar

6. 

P. Matousek, “Deep non-invasive Raman spectroscopy of living tissue and powders,” Chem. Soc. Rev., 36 (8), 1292 –1304 (2007). https://doi.org/10.1039/b614777c 0306-0012 Google Scholar

7. 

A. Carden and M. D. Morris, “Application of vibrational spectroscopy to the study of mineralized tissue (review),” J. Biomed. Opt., 5 (3), 259 –268 (2000). https://doi.org/10.1117/1.429994 1083-3668 Google Scholar

8. 

O. Akkus, F. Adar, and M. B. Schaffler, “Age-related changes in physicochemical properties of mineral crystals are related to impaired mechanical function of cortical bone,” Bone (N.Y.), 34 (3), 443 –453 (2004). https://doi.org/10.1016/j.bone.2003.11.003 8756-3282 Google Scholar

9. 

B. R. McCreadie, M. D. Morris, T.-C. Chen, D. Sudhaker Rao, W. F. Finney, E. Widjaja, and S. A. Goldstein, “Bone tissue compositional differences in women with and without osteoporotic fracture,” Bone (N.Y.), 39 (6), 1190 –1195 (2006). 8756-3282 Google Scholar

10. 

P. Mayer-Kuckuk and A. L. Boskey, “Molecular imaging promotes progress in orthopedic research,” Bone (N.Y.), 39 (5), 965 –977 (2006). 8756-3282 Google Scholar

11. 

C. M. Carpenter, B. W. Pogue, S. Jiang, H. Dehghani, X. Wang, K. D. Paulsen, W. A. Wells, J. Forero, C. Kogel, J. B. Weaver, S. P. Poplack, and P. A. Kaufman, “Image-guided optical spectroscopy provides molecular-specific information in vivo: MRI-guided spectroscopy of breast cancer hemoglobin, water, and scatterer size,” Opt. Lett., 32 (8), 933 –935 (2007). https://doi.org/10.1364/OL.32.000933 0146-9592 Google Scholar

12. 

S. C. Davis, H. Dehghani, J. Wang, S. Jiang, B. W. Pogue, and K. D. Paulsen, “Image-guided diffuse optical fluorescence tomography implemented with Laplacian-type regularization,” Opt. Express, 15 4066 –4082 (2007). https://doi.org/10.1364/OE.15.004066 1094-4087 Google Scholar

13. 

B. W. Pogue, S. C. Davis, X. Song, B. A. Brooksby, H. Dehghani, and K. D. Paulsen, “Image analysis methods for diffuse optical tomography,” J. Biomed. Opt., 11 (3), 033001 (2006). https://doi.org/10.1117/1.2209908 1083-3668 Google Scholar
©(2008) Society of Photo-Optical Instrumentation Engineers (SPIE)
Matthew V. Schulmerich, Jacqueline H. Cole, Kathryn A. Dooley, Michael D. Morris, Jaclynn Kreider, Steven A. Goldstein, Subhadra Srinivasan, and Brian W. Pogue "Noninvasive Raman tomographic imaging of canine bone tissue," Journal of Biomedical Optics 13(2), 020506 (1 March 2008). https://doi.org/10.1117/1.2904940
Published: 1 March 2008
Lens.org Logo
CITATIONS
Cited by 121 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Bone

Raman spectroscopy

Tomography

Tissues

Tissue optics

Image restoration

Data modeling

Back to Top