CommunicationFast and quiet MRI using a swept radiofrequency
Introduction
NMR signal is observed following one of three basic types of radiofrequency (RF) excitation: sequential, simultaneous, and random. Accordingly, three different types of NMR techniques have been developed: continuous wave (CW) [1], [2], pulsed [3], [4], [5], and stochastic [6], [7]. Historically, the first high resolution NMR spectra of liquid samples were acquired using the CW technique. NMR spectroscopy using the CW technique became an important tool in the field of chemistry until the early 1970’s. Then, shortly after realizing the efficiency of pulsed Fourier transform (FT) spectroscopy [8], pulsed FT supplanted CW as the main spectroscopic technique. Nowadays, pulsed FT spectroscopy continues to dominate the field, while the CW and stochastic NMR techniques are mainly used only for niche applications.
Magnetic resonance imaging (MRI) has additional technical requirements over high-resolution NMR spectroscopy of liquids. Because the object to be imaged is usually much larger than a sample tube, inevitably the static and RF fields used in MRI are more inhomogeneous than those used in high resolution NMR. Furthermore, MRI requires high power RF amplifiers to attain a sufficiently intense RF field to excite all spins simultaneously. In a search for ways to circumvent such difficulties, some researchers have reconsidered old, almost forgotten NMR techniques. For this reason, the “sleeping beauty” [9], stochastic NMR, was reawakened [10] and found to have remarkable features; most notably, the peak RF power required for spin excitation is at least two orders of magnitude lower than that required by conventional pulsed schemes. Very recently, a CW MRI system using low peak RF power has also been developed and applied to study a variety of heterogeneous materials exhibiting extremely fast spin–spin relaxation [11]. Another example is the extensive application in modern MRI of adiabatic sweep pulses, which evolved from the traditional CW technique. An advantage of the adiabatic sweep pulse is its ability to accomplish uniform rotation over a broad band of resonance frequencies [12].
This article presents a new MRI method that can be considered as a combination of all three basic NMR techniques. As in CW NMR, this new method uses swept RF excitation, but the sweep rate far exceeds the CW sweep rate even during a rapid-scan [13], [14]. Unlike the CW method in which the signal is acquired in the frequency domain, here the signal is treated as a function of time, as in the pulsed FT method. Finally, the method uses correlation, identical to that used in stochastic NMR, to extract the signal arising from the spin system. This novel method is dubbed SWIFT, for sweep imaging with Fourier transformation. The concept of using swept RF excitation instead of a monochromatic RF pulse or stochastic excitation and then reconstructing the NMR spectrum using the correlation method, was mentioned more than three decades ago [13], but was never put into practice.
The main advantage of SWIFT originates in its nearly simultaneous excitation and acquisition scheme. In conventional MRI, excitation and acquisition events are separated by the length of time known as the echo time (TE), which is typically >1 ms. This length of time is too long to allow detection of slowly tumbling nuclei with short spin–spin relaxation time (T2). By comparison, SWIFT allows TE ≈ 0, because signal acquisition can begin within a few microseconds after excitation. Thus, SWIFT is a powerful tool for imaging objects having a broad distribution of relaxation times, including very short T2 values. In particular, the method is expected to find extensive application for imaging and spectroscopy of semi-solid objects such as bone, macromolecules, and quadrupolar nuclei such as sodium-23 [15], potassium-39 [16], and boron-11 [17]. Here, a description of the new technique is provided, and proof-of-principle is shown by simulated and experimental data.
Section snippets
Sweeping technique
The simplest realization of the SWIFT method is presented in Fig. 1a. The scheme employs a sequence of RF pulses, each having a duration Tp typically in the millisecond range. In the present implementation, the RF pulse is one of the family of adiabatic hyperbolic secant (HSn) pulses [12], [18], [19] which utilizes both amplitude and frequency modulation. The amplitude of the pulse is denoted by the variable ω1 (t) (= γB1 (t), where γ is the gyromagnetic ratio and B1 (t) is the time-dependent
Pulsed FT NMR
In the pulsed FT technique, the excitation pulse is generally assumed to be short (a delta function), so that the FT of this input function can be considered approximately equal to unity, and thus, the system spectrum (H (ω)) can be obtained directly via FT of the impulse response (h (t)). In MRI, the RF pulse can approach a delta function only by restricting the flip angle to ≪90°. As the pulse length increases, the pulsed FT spectrum suffers from increasing phase and baseline distortions.
Image contrast
In the implementation of SWIFT, the magnetic field variation created by the applied gradient G can readily exceed other potential contributions (e.g., magnetic susceptibility differences and an inhomogeneous B0), so their effects are not perceivable in images. SWIFT images are also essentially unaffected by transverse relaxation, provided T2 > 10dw (see Fig. 4). Under these conditions, the signal intensity depends only on T1 and spin density (M0) as [8], [28]For
Experiments
Although many new applications are anticipated for this novel MRI technique, here the utility of SWIFT is demonstrated for imaging spins with extremely short T2 values (Fig. 5, Fig. 6). The first test object is thermoplastic, which is invisible by conventional MRI because the T2 value is extremely short (∼0.3 ms). As can be seen from Fig. 5, SWIFT produced a highly resolved 3D image of the object. It is worth noting that these images were obtained using a standard MRI scanner (4.7 T) which is not
Conclusions
In conclusion, the SWIFT technique has many novel and beneficial properties for MRI:
(a) Fast. The method avoids not only delays associated with refocusing pulses or gradient inversion, but also time for an excitation pulse, which is combined with the acquisition period. Of course, like all fast imaging sequences, SWIFT is limited by existing imaging system hardware and chosen compromise between acquisition speed, spatial resolution and S/N.
(b) Sensitive to short T2. The method is sensitive to
Acknowledgments
This research was supported by NIH Grants RR008079 and CA92004. The authors thank Dr. Ivan Tkac for helping with reconstruction software implementation and Dr. Jutta Ellermann for assistance in conducting the experiments.
References (34)
Magnetic resonance with stochastic excitation
J. Magn. Reson.
(1970)Coherent spectrometry with noise signals
J. Magn. Reson.
(1970)- et al.
Hadamard NMR imaging with slice selection
Magn. Reson. Imaging
(1996) - et al.
Development of a 3-D, multi-nuclear continuous wave NMR imaging system
J. Magn. Reson.
(2005) - et al.
The return of the frequency sweep: designing adiabatic pulses for contemporary NMR
J. Magn. Reson.
(2001) - et al.
Correlation NMR spectroscopy
J. Magn. Reson.
(1974) - et al.
Rapid scan Fourier transform NMR spectroscopy
J. Magn. Reson.
(1974) - et al.
Improved performance of frequency-swept pulses using offset-independent adiabaticity
J. Magn. Reson. A
(1996) Sensitivity enhancement in magnetic resonance
Adv. Magn. Reson.
(1966)Fast imaging in liquids and solids with the back-projection low angle shot (BLAST) technique
Magn. Reson. Imaging
(1994)
Stochastic spectroscopic imaging
Resonance absorption by nuclear magnetic moments in a solid
Phys. Rev.
Nuclear induction
Phys. Rev.
The nuclear induction experiment
Phys. Rev.
Transient nutations in nuclear magnetic resonance
Phys. Rev.
Spin echoes
Phys. Rev.
Application of Fourier transform spectroscopy to magnetic resonance
Rev. Sci. Instrum.
Cited by (283)
“Store-bought is fine”: Sensitivity considerations using shaped pulses for DEER measurements on Cu(II) labels
2023, Journal of Magnetic ResonanceCharacterizing Off-center MRI with ZTE
2022, Zeitschrift fur Medizinische PhysikUsing high spatial resolution fMRI to understand representation in the auditory network
2021, Progress in NeurobiologyCitation Excerpt :Apart from improved sound delivery systems, the development of alternative functional imaging techniques that do not use rapidly switching gradients can be of relevance. The minimal gradient switching steps used in sweep imaging with Fourier transform (SWIFT), for example (Idiyatullin et al., 2015, 2006), results in “quiet” functional imaging with no susceptibility induced dropouts typical of GE-EPI. SWIFT has the potential to become popular among auditory neuroscientists, who have been advocating for such features (Talavage and Hall, 2012).