Gamma and beta intra-operative imaging probes

https://doi.org/10.1016/S0168-9002(97)00247-7Get rights and content

Abstract

Small area (∼1.5 cm2) scintillation cameras for imaging gammas and betas using inter-changeable detector front ends were built and characterized. Components common to both emission imaging cameras include: (1) fiber optic bundles 2–3 m long, comprised of multi-clad fibers which connect the scintillation detector to (2) an MC-PMT; (3) parallel MC-PMT outputs feed a resistive positioning network and i-V converter/line driver network which produce balanced +X, X, +Y, and Y outputs; and (4) four ADCs and a Macintosh PC for system control and image display. The beta and gamma devices used distinct scintillation detectors which were characterized by both simulation and measurement.

The beta camera utilized a 0.5 mm by 1.25 cm φ CaF2(Eu) scintillation crystal coupled, through a diffusing light guide, to 19 2-mm φ optical fibers. These front-end fibers are in turn coupled by a more flexible fiber bundle to the MC-PMT. CaF2(Eu) has high light output, high beta sensitivity, and low gamma sensitivity. Image signals are histogrammed and displayed after Anger logic computations are performed on digitized signals. The beta camera has <0.6 mm FWHM intrinsic resolution.

The gamma camera concept was tested with matrices of discrete 1 × 1 mm2 and 2 × 2 mm2 CsI(Tl) and NaI(Tl) crystals of various lengths, and 3 mm thick continuous crystals. Configurations using 4 × 4 element matrices with one-to-one coupling between crystals and fiber channels, and light diffusers between each crystal matrix and fibers were evaluated. The continuous crystals were coupled directly to the fiber optics with signal and data processing analogous to the beta camera. Coupling of discrete crystals to fiber optics by both methods gave essentially perfect identification of the crystal of interaction, allowing spatial resolution to be defined by the crystal size and collimator. The continuous crystal gamma camera gave intrinsic resolution of ∼1.4 mm FWHM.

References (17)

  • B.E Patt

    Nucl. Instr. and Meth. A

    (1996)
  • Y Shao et al.

    IEEE Trans. Nucl. Sci.

    (1996)
  • L.R MacDonald

    IEEE Trans. Nucl. Sci.

    (1995)
  • M.P Tornai

    J. Nucl. Med.

    (1995)
  • L.R MacDonald
  • C.S Levin et al.

    IEEE Trans. Nucl. Sci.

    (1996)
  • M.P Tornai et al.

    IEEE Trans. Nucl. Sci.

    (1996)
  • M.P Tornai et al.M.P Tornai et al.

    IEEE Trans. Nucl. Sci.

    (1996)
There are more references available in the full text version of this article.

Cited by (32)

  • Nuclear probes and intraoperative gamma cameras

    2011, Seminars in Nuclear Medicine
    Citation Excerpt :

    Intraoperative gamma camera systems thus merit development and evaluation. A number of small field-of-view intraoperative gamma camera systems have been developed.47,48 The earliest systems were hand-held devices having fields of view of only 1.5-2.5 cm in diameter and using conventional NaI(Tl) or CsI(Tl) scintillation detectors.

  • Experimental study for an intraoperative probe for <sup>18</sup>F imaging with a silicon pixel detector

    2007, Nuclear Instruments and Methods in Physics Research, Section A: Accelerators, Spectrometers, Detectors and Associated Equipment
  • Application-Specific Small Field-of-View Nuclear Emission Imagers in Medicine

    2004, Emission Tomography: The Fundamentals of PET and SPECT
  • Instrumentation for intraoperative detection and imaging

    2022, Nuclear Oncology: From Pathophysiology to Clinical Applications
  • Limited-Angle TOF-PET for Intraoperative Surgical Applications: Latest Results

    2020, 2020 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2020
  • Limited-Angle TOF-PET for Intraoperative Surgical Application

    2019, 2019 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2019
View all citing articles on Scopus
View full text