Design and performance from an integrated PET/MRI system for small animals

  • Original Article
  • Published:
Annals of Nuclear Medicine Aims and scope Submit manuscript

Abstract

Objective

Although simultaneous measurements of PET and magnetic resonance imaging (MRI) can provide interesting results in molecular imaging research, most of the combined systems are huge and animal handling in the system is not easy. To minimize these problems, we developed a compact integrated PET/MRI (iPET/MRI) system for small animals.

Methods

For the iPET/MRI system, a new MR-compatible PET and a permanent magnet open MRI were designed. In the MRI, a tunnel is opened at the yoke of the magnet. The position-sensitive photo-multiplier tubes (PSPMTs) of the MR-compatible PET are positioned at the back of the yoke where the magnetic field is sufficiently low. The scintillators for the PET system are positioned at the center of the MRI magnets, and the direction of the scintillation photons is changed by slanted light guides, and they are fed to the PSPMTs by 75 cm long optical fiber bundles. The PET detectors employed two types of LGSO crystals (1.9 mm × 2.2 mm × 6 mm and 7 mm) with different decay times (33 and 43 ns) for depth of interaction detection. Sixteen optical fiber-based block detectors are arranged in a 112 mm diameter ring.

Results

The transaxial field-of-view (FOV) of the PET system is ~80 mm, and the axial FOV is 21 mm which can be enlarged by the axial motion of the PET detector ring during MRI acquisition. The transaxial and axial resolutions at the center of the PET system was 2.9 and 2.4 mm FWHM, respectively. The absolute sensitivity was 1.5% at the center of the axial FOV. Phantom images revealed no artifact in either the PET or MRI images. We successfully obtained simultaneously measured small animal images using the iPET/MRI system.

Conclusion

The open geometry of the developed iPET/MRI facilitates easy accessibility to the subject. The iPET/MRI system appears to be a promising tool for molecular imaging research.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9
Fig. 10
Fig. 11
Fig. 12
Fig. 13
Fig. 14
Fig. 15
Fig. 16
Fig. 17
Fig. 18
Fig. 19

Similar content being viewed by others

References

  1. Shao Y, Cherry SR, Farahani K, Slates R, Silverman RW, Meadors K, et al. Development of a PET detector system compatible with MRI/NMR system. IEEE Trans Nucl Sci. 1997;44:1167–71.

    Article  CAS  Google Scholar 

  2. Shao Y, Cherry SR, Farahani K, Meadors K, Siegel S, Silverman RW, et al. Simultaneous PET and MR imaging. Phys Med Biol. 1997;42:1965–70.

    Article  CAS  PubMed  Google Scholar 

  3. Garlick PB, Marsden PK, Cave AC, Parkes HG, Slates R, Shao Y, et al. PET and NMR dual acquisition (PANDA): applications to isolated, perfused rat heart. NMR Biomed. 1997;10:138–42.

    Article  CAS  PubMed  Google Scholar 

  4. Slate R, Cherry SR, Boutefnouchet A, Shao Y, Dahlbom M, Farahani K. Design of a small animal MR compatible PET scanner. IEEE Trans Nucl Sci. 1999;46:565–70.

    Article  Google Scholar 

  5. Yamamoto S, Murayama H. A block detector for a multi-slice depth of interaction MR-compatible PET. IEEE Trans Nucl Sci. 2005;52:33–7.

    Article  Google Scholar 

  6. Imaizumi M, Yamamoto S, Kawakami M, Aoki M, Sugiyama E, Kanai Y, et al. Simultaneous imaging of magnetic resonance imaging and positron emission tomography by means of MRI-compatible optic fiber based PET: a validation study in ex vivo rat brain. Jpn J Radiol. 2009;27:252–6.

    Article  PubMed  Google Scholar 

  7. Raylman RR, Majewski S, Lemieux SK, Velan SS, Kross B, Popov V, et al. Simultaneous MRI and PET imaging of a rat brain. Phys Med Biol. 2006;51:6371–9.

    Article  PubMed  Google Scholar 

  8. Pichler BJ, Judenhofer MS, Catana C, Walton JH, Kneilling M, Nutt RE, et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med. 2006;47:639–47.

    PubMed  Google Scholar 

  9. Judenhofer MS, Wehrl HF, Newport DF, Catana C, Siegel SB, Becker M, et al. Simultaneous PET-MRI: a new approach for functional and morphological imaging. Nat Med. 2008;14:459–65.

    Article  CAS  PubMed  Google Scholar 

  10. Catana C, Wu Y, Judenhofer MS, Qi J, Pichler BJ, Cherry SR. Simultaneous acquisition of multi-slice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med. 2006;47:1968–76.

    PubMed  Google Scholar 

  11. Catana C, Procissi D, Wu Y, Judenhofer MS, Qi J, Pichler BJ, et al. Simultaneous in vivo positron emission tomography and magnetic resonance imaging. Proc Natl Acad Sci USA. 2008;105:3705–10.

    Article  CAS  PubMed  Google Scholar 

  12. Gilbert KM, Handler WB, Scholl TJ, Odegaard JW, Chronik BA. Design of field cycled magnetic resonance systems for small animal imaging. Phys Med Biol. 2006;51:2825–41.

    Article  CAS  PubMed  Google Scholar 

  13. Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR compatible gamma detectors. IEEE Trans Nucl Sci. 2003;50:1683–5.

    Article  CAS  Google Scholar 

  14. Yamamoto S. Optimization of the integration time of pulse shape analysis for dual layer detector with different amount of Ce. Nucl Instrum Methods A. 2008;587:319–23.

    Article  CAS  Google Scholar 

  15. Schlyer D, Vaska P, Tomasi D, Woody C, Solis-Najera S, Southekal S, et al. Preliminary studies of a simultaneous PET/MRI scanner based on the RatCAP small animal tomograph. Nucl Sci Symp Conf Rec IEEE. 2006;4:2340–4.

    Google Scholar 

Download references

Acknowledgments

This work was supported by the National Institute of Biomedical Innovation, Osaka, Japan.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Seiichi Yamamoto.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Yamamoto, S., Imaizumi, M., Kanai, Y. et al. Design and performance from an integrated PET/MRI system for small animals. Ann Nucl Med 24, 89–98 (2010). https://doi.org/10.1007/s12149-009-0333-6

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s12149-009-0333-6

Keywords

Navigation