Simultaneous acquisition of iodine-123 emission and technetium-99m transmission data for quantitative brain single-photon emission tomographic imaging

  • Original article
  • Published:
European Journal of Nuclear Medicine Aims and scope Submit manuscript

Abstract.

The aim of this study was to obtain quantitative iodine-123 brain single-photon emission tomographic (SPET) images with scatter and attenuation correction. We used a triple-headed SPET gamma camera system equipped with fan-beam collimators with a technetium-99m line transmission source placed at one of the focal lines of the fan-beam collimators. Four energy windows were employed for data acquisition: (a) 126–132 keV, (b) 132–143 keV, (c) 143–175 keV and (d) 175–186 keV. A simultaneous transmission-emission computed tomography scan (TCT-ECT) was carried out for a brain phantom containing 123I solution. The triple energy window scatter correction was applied to the 123I ECT data measured by means of the windows (b), (c) and (d) acquired by two detectors. Attenuation maps were reconstructed from 99mTc TCT data measured by means of the windows (a), (b) and (c) acquired by one detector. Chang’s iterative attenuation correction method using the attenuation maps was applied to the 123I ECT images. In the phantom study cross-calibrated SPET values obtained with the simultaneous mode were almost equal to those obtained with the sequential mode, and they were close to the true value, within an error range of 5.5%. In the human study corrected images showed a higher grey-to-white matter count ratio and relatively higher uptake in the cerebellum, basal ganglia and thalamus than uncorrected images. We conclude that this correction method provides improved quantification and quality of SPET images and that the method is clinically practical because it requires only a single scan with a 99mTc external source.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received 6 June and in revised form 27 July 1998

Rights and permissions

Reprints and permissions

About this article

Cite this article

Ogasawara, K., Hashimoto, J., Ogawa, K. et al. Simultaneous acquisition of iodine-123 emission and technetium-99m transmission data for quantitative brain single-photon emission tomographic imaging. Eur J Nucl Med 25, 1537–1544 (1998). https://doi.org/10.1007/s002590050333

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/s002590050333

Navigation