Skip to main content

Advertisement

Log in

Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET

  • Original Article
  • Published:
European Journal of Nuclear Medicine Aims and scope Submit manuscript

Abstract.

The first performance tests of a new fully programmable blood sampling device for monitoring blood radioactivity during positron emission tomography (PET) are described. Blood is withdrawn through 1-mm internal diameter tubing using an infusion pump which can be operated at rates varying from 0 to 600 ml/h. Activity in blood is measured by a 6-cm-thick bismuth germanate crystal connected to a photomultiplier tube and multichannel analyser (MCA) which are positioned within 6 cm lead shielding. Positioning of the tubing is an exact and simple procedure. The minimal readout time of the MCA is 1 s. Two independent energy windows can be set. Operation of the pump and MCA is fully controlled by a PC, i.e. sampling time, interval time and pump rate can be varied at any time during the PET scan by user-defined scripts. A number of characteristics of the new system were studied, such as sensitivity, dead time, linearity, effect of background radiation and pump rate as a function of input pressure. In addition, dispersion was measured as a function of pump rate. Finally, first clinical results were compared with manual samples. The sensitivity equalled 0.7 and 0.2 cps/Bq for 511- and 1022-keV 30% energy windows, respectively, and the system dead time was 500 ns. The system remained linear within 2% with activity concentrations up to 2.5 MBq/cc. Short-term reproducibility was better than 3% for a 1-h period. Long-term reproducibility was about 5% (1SD), which was mainly caused by variation in the diameter of the tubing. If the device was positioned in such a way that maximum shielding was directed towards the patient, the effects of background radiation from the patient on the measured activity concentration for clinically relevant conditions was minimal (<3%). Pump rate varied with input pressure, but remained constant for a given pressure. Dispersion constants smaller than 0.14 s–1 were observed for pump rates higher than 300 ml/h, indicating that the system dispersion is small. Clinical data showed an excellent agreement to within 3% (1SD) between the results obtained with the new system and manual samples. With the continuous blood sampler radioactivity in blood can be measured accurately during the entire course of the PET scan. Furthermore, the system is fully programmable allowing adjustment of all parameters during a single PET scan.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

Author information

Authors and Affiliations

Authors

Additional information

Received 19 June and in revised form 20 September 2000

Electronic Publication

Rights and permissions

Reprints and permissions

About this article

Cite this article

Boellaard, R., van Lingen, A., van Balen, S. et al. Characteristics of a new fully programmable blood sampling device for monitoring blood radioactivity during PET. Eur J Nucl Med 28, 81–89 (2001). https://doi.org/10.1007/s002590000405

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s002590000405

Navigation