The combination of clinical MRI and PET systems has received increased attention in recent years. In contrast to currently used PET/CT systems, PET/MRI offers not only improved soft-tissue contrast and reduced levels of ionizing radiation, but also a wealth of MRI-specific information such as functional, spectroscopic and diffusion tensor imaging. Combining PET and MRI, however, has proven to be very challenging, due to the detrimental cross-talk effects between the two systems.

Objective

Significant progress has been made in the recent years to overcome these difficulties, with several groups reporting PET/MRI prototypes for animal imaging and a clinical insert for neurological applications being demonstrated at the 2007 Annual Meeting of the Society of Nuclear Medicine.

Discussion

In this paper we review different architectures for clinical PET/MRI systems, and their possibilities, limitations and technological obstacles.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4

Similar content being viewed by others

References

  1. Beyer T, Townsend DW, Brun T, Kinahan PE, Charron M, Roddy R, et al. A combined PET/CT scanner for clinical oncology. J Nucl Med 2000;41:1369–79.

    PubMed  CAS  Google Scholar 

  2. European Commission. Referral guidelines for imaging radiation protection 118. Technical Report. ISBN: 92-828-9454-1; 2001.

  3. Zaidi H, Mawlawi O, Orton CG. Point/counterpoint. Simultaneous PET/MR will replace PET/CT as the molecular multimodality imaging platform of choice. Med Phys 2007;34:1525–8.

    Article  PubMed  Google Scholar 

  4. Christensen NL, Hammer BE, Heil BG, Fetterly K. Positron emission tomography within a magnetic field using photomultiplier tubes and lightguides. Phys Med Biol 1995;40:691–7.

    Article  PubMed  CAS  Google Scholar 

  5. Cumalat JP, Cheung HWK, Hassed J, Smith BD, Bross AD. Effects of magnetic fields on the light yield of scintillators. Nucl Instrum Methods Phys Res Sect A 1990;293:606–14.

    Article  Google Scholar 

  6. Yamamoto S, Kuroda K, Senda M. Scintillator selection for MR-compatible gamma detectors. IEEE Trans Nucl Sci 2003;50:1683–5.

    Article  CAS  Google Scholar 

  7. Lucas AJ, Hawkes RC, Ansorge RE, Williams GB, Nutt RE, Clark JC, et al. Development of a combined microPET-MR system. Technol Cancer Res Treat 2006;5:337–41.

    PubMed  CAS  Google Scholar 

  8. Marsden PK, Strul D, Keevil SF, Williams SCR, Cash D. Simultaneous PET and NMR. Br J Radiol 2002;75:53–9.

    Google Scholar 

  9. Raylman RR, Majewski S, Velan SS, Lemieux S, Kross B, Popov V, et al. Simultaneous acquisition of magnetic resonance spectroscopy (MRS) data and positron emission tomography (PET) images with a prototype MR-compatible, small animal PET imager. J Magn Reson 2007;186:305–10.

    Article  PubMed  CAS  Google Scholar 

  10. Shao Y, Cherry S, Farahani K, Slates R, Silverman R, Meadors K, et al. Development of a PET detector system compatible with MRI/NMR systems. IEEE Trans Nucl Sci 1997;44:1167–71.

    Article  CAS  Google Scholar 

  11. Pichler B, Lorenz E, Mirzoyan R, Pimpl W, Roder F, Schwaiger M, et al. Performance test of a LSO-APD PET module in a 9.4 Tesla magnet. Proc IEEE Nucl Sci Symp 1997;2:1237–9.

    CAS  Google Scholar 

  12. Catana C, Wu Y, Judenhofer MS, Qi J, Pichler BJ, Cherry SR. Simultaneous acquisition of multislice PET and MR images: initial results with a MR-compatible PET scanner. J Nucl Med 2006;47:1968–76.

    PubMed  Google Scholar 

  13. Catana C, Procissi D, Wu Y, Judenhofer MS, Qi J, Pichler BJ, et al. Simultaneous in vivo positron emission tomography and magnetic resonance imaging. Proc Natl Acad Sci U S A 2008;105:3705–10.

    Article  PubMed  CAS  Google Scholar 

  14. Grazioso R, Zhanga N, Corbeila J, Schmanda M, Ladebeck R, Vester M, et al. APD-based PET detector for simultaneous PET/MR imaging. Nucl Instrum Methods Phys Res Sect A 2006;569:301–5.

    Article  CAS  Google Scholar 

  15. Pichler BJ, Judenhofer MS, Catana C, Walton JH, Kneilling M, Nutt RE, et al. Performance test of an LSO-APD detector in a 7-T MRI scanner for simultaneous PET/MRI. J Nucl Med 2006;47:639–47.

    PubMed  Google Scholar 

  16. Woody C, Schlyer D, Vaska P, Tomasi D, Solis-Najera S, Rooney W. Preliminary studies of a simultaneous PET/MRI scanner based on the RatCAP small animal tomograph. Nucl Instrum Methods Phys Res Sect A 2007;571:102–5.

    Article  CAS  Google Scholar 

  17. Junnarkar S, Fried J, O’Connor P, Radeka V, Vaska P, Purschke M. MRI compatible G-link and PCI based data acquisition hardware for the RatCAP scanner. IEEE Nucl Sci Symp 2006;1:380–3.

    Google Scholar 

  18. Zang-Hee C, Young-Don S, Hang-Keun K, Kyoung-Nam K, Se-Hong O, Jae-Yong H, et al. A hybrid PET-MRI: an integrated molecular-genetic imaging system with HRRT-PET and 7.0-T MRI. Int J Imaging Sys Tech 2007;17:252–65.

    Article  Google Scholar 

  19. Shao Y, Cherry SR, Farahani K, Meadors K, Siegel S, Silverman RW, et al. Simultaneous PET and MR imaging. Phys Med Biol 1997;42:1965–70.

    Article  PubMed  CAS  Google Scholar 

  20. Slates RB, Farahani K, Shao Y, Marsden PK, Taylor J, Summers PE. A study of artefacts in simultaneous PET and MR imaging using a prototype MR compatible PET scanner. Phys Med Biol 1999;44:2015–27.

    Article  PubMed  CAS  Google Scholar 

  21. Pichler BJ, Swann BK, Rochelle J, Nutt RE, Cherry SR, Siegel SB. Lutetium oxyorthosilicate block detector readout by avalanche photodiode arrays for high resolution animal PET. Phys Med Biol 2004;49:4305–19.

    Article  PubMed  CAS  Google Scholar 

  22. Spanoudaki VC, McElroy DP, Torres-Espallardo I, Ziegler SI. Effect of temperature on the performance of proportional APD-based modules for gamma ray detection in positron emission tomography. IEEE Trans Nucl Sci 2008;55:469–80.

    Article  Google Scholar 

  23. Carson RE, Herscovitch P, Daube-Witherspoon ME, editors. Quantitative functional brain imaging with positron emission tomography. 1st ed. New York: Academic Press; 1998.

  24. Raylman RR, Majewski S, Lemieux SK, Velan SS, Kross B, Popov V. Simultaneous MRI and PET imaging of a rat brain. Phys Med Biol 2006;51:6371–9.

    Article  PubMed  Google Scholar 

  25. Schlyer D, Vaska P, Tomasi D, Woody C, Maramraju S-H, Southekal S. A simultaneous PET/MRI scanner based on RatCAP in small animals. IEEE Nucl Sci Symp 2007;5:3256–9.

    Google Scholar 

  26. Judenhofer MS, Wehrl HF, Newport DF, Catana C, Siegel SB, Becker M, et al. Simultaneous PET-MRI: a new approach for functional and morphological imaging. Nat Med 2008;14:459–65.

    Article  PubMed  CAS  Google Scholar 

  27. Schlemmer H, Pichler B, Wienhard K, Schmand M, Nahmias C, Townsend D, et al. Simultaneous MR/PET for brain imaging: first patient scans (abstract). J Nucl Med 2007;48(Suppl 2):45P.

    Google Scholar 

  28. Schlemmer H, Pichler BJ, Schmand M, Burbar Z, Michel C, Ladebeck R, et al. Simultaneous MR/PET imaging of the human brain: feasibility study. Radiology 2008;248:1028–35.

    Article  PubMed  Google Scholar 

  29. Handler W, Chronik B, Scholl T, Gilbert K, Peng H. Combining field-cycled magnetic resonance imaging with positron emission tomography (abstract). J Nucl Med Meeting Abstracts 2007;48(Suppl 2):89P.

    Google Scholar 

  30. Martinez M, Torres I, Ladebeck R, Nekolla S, Ziegler S, Schwaiger M. Whole-body MR-PET: characterization of PET performance by Monte Carlo simulations (abstract). J Nucl Med 2007;48(Suppl 2):46P.

    Google Scholar 

  31. Otte A, Barral J, Dolgoshein B, Hose J, Klemin S, Lorenz E, et al. A test of silicon photomultipliers as readout for PET. Nucl Instrum Methods Phys Res Sect A 2005;545:705–15.

    Article  CAS  Google Scholar 

  32. Spanoudaki VC, Mann AB, Otte AN, Konorov I, Torres-Espallardo I, Paul S, et al. Use of single photon counting detector arrays in combined PET/MR: characterization of LYSO-SiPM detector modules and comparison with a LSO-APD detector. J Instrum 2007;2:P12002.

    Article  Google Scholar 

  33. Strul D, Cash D, Keevil S, Halsted P, Williams S, Marsden P. Gamma shielding materials for MR-compatible PET. IEEE Trans Nucl Sci 2003;50:60–9.

    Article  CAS  Google Scholar 

  34. Lee H, Li Z, Chen K, Hsu A, Xu C, Xie J, et al. PET/MRI dual-modality tumor imaging using arginine-glycine-aspartic (RGD)-conjugated radiolabeled iron oxide nanoparticles. J Nucl Med 2008;49:1371–9.

    Article  PubMed  CAS  Google Scholar 

  35. Schwaiger M, Ziegler S, Nekolla S. MR-PET: Combining function, anatomy and more. Medical Solutions, Special edition 2005 - Molecular imagin, http://www.medical.siemens.com/siemens/en_US/rg_marcom_FBAs/files/brochures/brochures/magazin_special2005/P25–30_MRPET_Combining_e.pdf.

Download references

Acknowledgment

This work was supported by a research grant from Siemens AG.

Conflict of interest

None.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Gaspar Delso.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Delso, G., Ziegler, S. PET/MRI system design. Eur J Nucl Med Mol Imaging 36 (Suppl 1), 86–92 (2009). https://doi.org/10.1007/s00259-008-1008-6

Download citation

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00259-008-1008-6

Keywords

Navigation