Abstract
Purpose
The aim of this study was to improve the uniformity of the axial spatial resolution and sensitivity in pinhole single-photon emission computed tomography (SPECT) and to extend the axial field of view, by using a dedicated three-pinhole collimator.
Methods
A rectangular tungsten plate with three pinhole apertures of 1.5 mm diameter was designed to image a cylindrical field of view of 55 mm diameter and 160 mm axial length using a clinical gamma camera. To evaluate the non-uniformity of spatial resolution and noise, a multiple-disk phantom was built. The phantom was filled with Tc-99m, and data were acquired using a circular orbit and reconstructed with a dedicated iterative reconstruction algorithm. The axial spatial resolution together with the noise was measured in each disk. These measurements were compared to a single-pinhole system using an identical acquisition geometry and reconstruction.
Results
At the central slice, a spatial resolution of 2.7 mm was observed for both the three-pinhole and single-pinhole geometries. At 17.5 mm from the central slice, the axial spatial resolution deteriorated to 10.3 mm when using a single pinhole, while the spatial resolution remained 2.7 mm for the three-pinhole system. In the central slice, 19% noise was observed for both geometries. At 31.5 mm from this central slice, the noise remained 19% for the three-pinhole geometry, while it increased to 32% using a single pinhole.
Conclusion
The presented three-pinhole collimator improves the uniformity of the axial spatial resolution and sensitivity in pinhole SPECT and consequently extends the axial field of view, a requirement for whole-body small-animal imaging.
Similar content being viewed by others
References
Meikle SR, Kench P, Kassiou M, Banati RB. Small animal SPECT and its place in the matrix of molecular imaging technologies. Phys Med Biol. 2005;50(22):R45–61, Nov. 21.
Beekman FJ, van der Have F, Vastenhouw B, et al. U-SPECT-I: a novel system for submillimeter-resolution tomography with radiolabeled molecules in mice. J Nucl Med. 2005;46(7):1194–200, Jul.
Furenlid LR, Wilson DW, Chen Y, et al. FastSPECT II: A second-generation high-resolution dynamic SPECT images. IEEE Trans Nucl Sci. 2004;51(3):631–5.
Meikle SR, Kench P, Weisenberger AG, et al. A prototype coded aperture detector for small animal SPECT. IEEE Trans Nucl Sci. 2002;49(5):2167–71.
Schramm NU, Ebel G, Engeland U, Schurrat T, Behe M, Behr TM. High-resolution SPECT using multipinhole collimation. IEEE Trans Nucl Sci. 2003;50(3):315–20.
Beekman F, van der Have F. The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging. Eur J Nucl Med Mol Imaging. 2007;34(2):151–61, Feb.
Kudo H, Saito T. Feasible cone beam scanning methods for exact reconstruction in three-dimensional tomography. J Opt Soc Am. 1990;7(12):2169–83, Dec.
Toshiyuki A, Tsutomu Z, Hiroshi W, Hossain M, Tetsuya M, Hidehiro I. System design and development of a pinhole SPECT system for quantitative functional imaging of small animals. Ann Nucl Med. 2006;20(3):245–51.
Zeniya T, Watabe H, Aoi T, et al. A new reconstruction strategy for image improvement in pinhole SPECT. Eur J Nucl Med Mol Imaging. 2004;31(8):1166–72, August.
Zeng GL, Gullberg GT. A cone-beam tomography algorithm for orthogonal circle-and-line orbit. Phys Med Biol. 1992;37(3):563–77, March.
Metzler SD, Accorsi R, Novak JR, Ayan AS, Jaszczak RJ. On-axis sensitivity and resolution of a slit-slat collimator. J Nucl Med. 2006;47(11):1884–90, Nov.
Walrand S, Jamar F, de Jong M, Pauwels S. Evaluation of novel whole-body high-resolution rodent SPECT (Linoview) based on direct acquisition of linogram projections. J Nucl Med. 2005;46(11):1872–80, Nov.
Vunckx K, Beque D, Defrise M, Nuyts J. Single and multipinhole collimator design evaluation method for small animal SPECT. Nuclear Science Symposium Conference Record, 2005 IEEE. 2005;4(23-29):2223–2227, Oct.
Tuy HK. An inversion formula for cone-beam reconstruction. SIAM J Appl Math. 1983;43(3):546–52.
Vanhove C, Defrise M, Franken PR, Everaert H, Deconinck F, Bossuyt A. Interest of the ordered subsets expectation maximization (OS-EM) algorithm in pinhole single-photon emission tomography reconstruction: a phantom study. Eur J Nucl Med. 2000;27(2):140–6, Feb.
Beque D, Nuyts J, Bormans G, Suetens P, Dupont P. Characterization of pinhole SPECT acquisition geometry. IEEE Trans Med Imag. 2003;22(5):599–612, May.
Beque D, Nuyts J, Suetens P, Bormans G. Optimization of geometrical calibration in pinhole SPECT. IEEE Trans Med Imag. 2005;24(2):180–90, Feb.
Gainkam Tchouate O, Huang L, Caveliers V, et al. Intra-individual comparison of 99mTc-anti-EGFR Nanobodies™ by in vivo imaging in mice tumor xenografts expressing EGFR. J Nucl Med. 2007:Abstract at the SNM 2007 meeting.
Cao Z, Bal G, Accorsi R, Acton PD. Optimal number of pinholes in multi-pinhole SPECT for mouse brain imaging—a simulation study. Phys Med Biol. 2005;50(19):4609–24, Oct.
Vanhove C, Andreyev A, Defrise M, Nuyts J, Bossuyt A. Resolution recovery in pinhole SPECT based on multi-ray projections: a phantom study. Eur J Nucl Med Mol Imaging. 2007;34(2):170–80, Feb.
Author information
Authors and Affiliations
Corresponding author
Rights and permissions
About this article
Cite this article
Vanhove, C., Defrise, M., Lahoutte, T. et al. Three-pinhole collimator to improve axial spatial resolution and sensitivity in pinhole SPECT. Eur J Nucl Med Mol Imaging 35, 407–415 (2008). https://doi.org/10.1007/s00259-007-0579-y
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00259-007-0579-y