Purpose

The aim of this study was to improve the uniformity of the axial spatial resolution and sensitivity in pinhole single-photon emission computed tomography (SPECT) and to extend the axial field of view, by using a dedicated three-pinhole collimator.

Methods

A rectangular tungsten plate with three pinhole apertures of 1.5 mm diameter was designed to image a cylindrical field of view of 55 mm diameter and 160 mm axial length using a clinical gamma camera. To evaluate the non-uniformity of spatial resolution and noise, a multiple-disk phantom was built. The phantom was filled with Tc-99m, and data were acquired using a circular orbit and reconstructed with a dedicated iterative reconstruction algorithm. The axial spatial resolution together with the noise was measured in each disk. These measurements were compared to a single-pinhole system using an identical acquisition geometry and reconstruction.

Results

At the central slice, a spatial resolution of 2.7 mm was observed for both the three-pinhole and single-pinhole geometries. At 17.5 mm from the central slice, the axial spatial resolution deteriorated to 10.3 mm when using a single pinhole, while the spatial resolution remained 2.7 mm for the three-pinhole system. In the central slice, 19% noise was observed for both geometries. At 31.5 mm from this central slice, the noise remained 19% for the three-pinhole geometry, while it increased to 32% using a single pinhole.

Conclusion

The presented three-pinhole collimator improves the uniformity of the axial spatial resolution and sensitivity in pinhole SPECT and consequently extends the axial field of view, a requirement for whole-body small-animal imaging.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Fig. 1.
Fig. 2.
Fig. 3.
Fig. 4.
Fig. 5.
Fig. 6.

Similar content being viewed by others

References

  1. Meikle SR, Kench P, Kassiou M, Banati RB. Small animal SPECT and its place in the matrix of molecular imaging technologies. Phys Med Biol. 2005;50(22):R45–61, Nov. 21.

    Article  PubMed  CAS  Google Scholar 

  2. Beekman FJ, van der Have F, Vastenhouw B, et al. U-SPECT-I: a novel system for submillimeter-resolution tomography with radiolabeled molecules in mice. J Nucl Med. 2005;46(7):1194–200, Jul.

    PubMed  Google Scholar 

  3. Furenlid LR, Wilson DW, Chen Y, et al. FastSPECT II: A second-generation high-resolution dynamic SPECT images. IEEE Trans Nucl Sci. 2004;51(3):631–5.

    Article  Google Scholar 

  4. Meikle SR, Kench P, Weisenberger AG, et al. A prototype coded aperture detector for small animal SPECT. IEEE Trans Nucl Sci. 2002;49(5):2167–71.

    Article  Google Scholar 

  5. Schramm NU, Ebel G, Engeland U, Schurrat T, Behe M, Behr TM. High-resolution SPECT using multipinhole collimation. IEEE Trans Nucl Sci. 2003;50(3):315–20.

    Article  Google Scholar 

  6. Beekman F, van der Have F. The pinhole: gateway to ultra-high-resolution three-dimensional radionuclide imaging. Eur J Nucl Med Mol Imaging. 2007;34(2):151–61, Feb.

    Article  PubMed  Google Scholar 

  7. Kudo H, Saito T. Feasible cone beam scanning methods for exact reconstruction in three-dimensional tomography. J Opt Soc Am. 1990;7(12):2169–83, Dec.

    CAS  Google Scholar 

  8. Toshiyuki A, Tsutomu Z, Hiroshi W, Hossain M, Tetsuya M, Hidehiro I. System design and development of a pinhole SPECT system for quantitative functional imaging of small animals. Ann Nucl Med. 2006;20(3):245–51.

    Article  Google Scholar 

  9. Zeniya T, Watabe H, Aoi T, et al. A new reconstruction strategy for image improvement in pinhole SPECT. Eur J Nucl Med Mol Imaging. 2004;31(8):1166–72, August.

    Article  PubMed  Google Scholar 

  10. Zeng GL, Gullberg GT. A cone-beam tomography algorithm for orthogonal circle-and-line orbit. Phys Med Biol. 1992;37(3):563–77, March.

    Article  PubMed  CAS  Google Scholar 

  11. Metzler SD, Accorsi R, Novak JR, Ayan AS, Jaszczak RJ. On-axis sensitivity and resolution of a slit-slat collimator. J Nucl Med. 2006;47(11):1884–90, Nov.

    PubMed  Google Scholar 

  12. Walrand S, Jamar F, de Jong M, Pauwels S. Evaluation of novel whole-body high-resolution rodent SPECT (Linoview) based on direct acquisition of linogram projections. J Nucl Med. 2005;46(11):1872–80, Nov.

    PubMed  Google Scholar 

  13. Vunckx K, Beque D, Defrise M, Nuyts J. Single and multipinhole collimator design evaluation method for small animal SPECT. Nuclear Science Symposium Conference Record, 2005 IEEE. 2005;4(23-29):2223–2227, Oct.

    Article  Google Scholar 

  14. Tuy HK. An inversion formula for cone-beam reconstruction. SIAM J Appl Math. 1983;43(3):546–52.

    Article  Google Scholar 

  15. Vanhove C, Defrise M, Franken PR, Everaert H, Deconinck F, Bossuyt A. Interest of the ordered subsets expectation maximization (OS-EM) algorithm in pinhole single-photon emission tomography reconstruction: a phantom study. Eur J Nucl Med. 2000;27(2):140–6, Feb.

    Article  PubMed  CAS  Google Scholar 

  16. Beque D, Nuyts J, Bormans G, Suetens P, Dupont P. Characterization of pinhole SPECT acquisition geometry. IEEE Trans Med Imag. 2003;22(5):599–612, May.

    Article  Google Scholar 

  17. Beque D, Nuyts J, Suetens P, Bormans G. Optimization of geometrical calibration in pinhole SPECT. IEEE Trans Med Imag. 2005;24(2):180–90, Feb.

    Article  Google Scholar 

  18. Gainkam Tchouate O, Huang L, Caveliers V, et al. Intra-individual comparison of 99mTc-anti-EGFR Nanobodies™ by in vivo imaging in mice tumor xenografts expressing EGFR. J Nucl Med. 2007:Abstract at the SNM 2007 meeting.

  19. Cao Z, Bal G, Accorsi R, Acton PD. Optimal number of pinholes in multi-pinhole SPECT for mouse brain imaging—a simulation study. Phys Med Biol. 2005;50(19):4609–24, Oct.

    Article  PubMed  Google Scholar 

  20. Vanhove C, Andreyev A, Defrise M, Nuyts J, Bossuyt A. Resolution recovery in pinhole SPECT based on multi-ray projections: a phantom study. Eur J Nucl Med Mol Imaging. 2007;34(2):170–80, Feb.

    Article  PubMed  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Christian Vanhove.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Vanhove, C., Defrise, M., Lahoutte, T. et al. Three-pinhole collimator to improve axial spatial resolution and sensitivity in pinhole SPECT. Eur J Nucl Med Mol Imaging 35, 407–415 (2008). https://doi.org/10.1007/s00259-007-0579-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00259-007-0579-y

Keywords

Navigation