Skip to main content

Advertisement

Log in

A newly developed intra-operative gamma camera: performance characteristics in a laboratory phantom study

  • Original Article
  • Published:
European Journal of Nuclear Medicine and Molecular Imaging Aims and scope Submit manuscript

Abstract

Purpose

Radioguided surgery depends on the intra-operative detection of radiolabelled tissues. This is currently accomplished with hand tools capable of providing a tone signal, depending on the proximity and direction of a radioactive source in relation to the probe. The advantages of visual images of radiolabelled tissues are well recognised, but satisfactory means of acquiring such images intra-operatively are not yet available. The goal of this study was to examine the performance of a newly developed intra-operative gamma camera, compact enough to be a hand tool and capable of yielding a visual image of the source field.

Methods

The study was performed in the laboratory with a phantom consisting of a water bath and small hollow spheres (1–2 cm in internal diameter) filled with 99mTc (1–5 μCi/cc), placed in different configurations within the bath. For comparison, studies were also performed using a standard intra-operative gamma probe, and others using a standard single-head high-resolution gamma camera.

Results

Compared with the gamma probe, the intra-operative camera was found to possess a superior ability to distinguish small, deep and weakly localised radioactivity sources from background. By acquiring images from different angles, it allowed a 3D understanding of multiple radioactive sources. It detected “cold” defects within a “hot” radiolabelled sphere. It discriminated a weak source located near a much “hotter” radioactivity source, similar to discrimination with the standard gamma camera, and discerned localised sources against a background of radioactivity.

Conclusion

It is anticipated that the high imaging potential of the camera tested in this study will offer clinical advantages.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7
Fig. 8
Fig. 9

Similar content being viewed by others

References

  1. Hoffman EJ, Tornai MP, Levin CS, MacDonald LR, Siegel S. Gamma and beta intra-operative imaging probes. Nucl Instr Methods Phys Res 1997;392:324–9

    CAS  Google Scholar 

  2. Menard L, Charon Y, Solal M, Ricard M, Lanièce P, Mastrippolito R, et al. Performance characterization and first clinical evaluation of a intra-operative compact gamma imager. IEEE Trans Nucl Sci 1998;46:2068–74

    Article  Google Scholar 

  3. MacDonald LR, Patt BE, Iwanczyk JS, Yamaguchi Y, McElroy D, Hoffman EJ, et al. High-resolution hand-held gamma camera. Proc SPIE Int Soc Opt Eng 2000; 4142:242–53

    Google Scholar 

  4. O’Connor MK, Blevis I. Evaluation of a small CZT detector for scintimammography. J Nucl Med 2001;42(Suppl):291P–2P

    Google Scholar 

  5. Parnham KB, Davies RK, Vydrin S, Ferraro F, Jimbo M, Ryou H. Development, design and performance of a CdZnTe-based nuclear medical imager. J Nucl Med 2002;43(Suppl):229P

    Google Scholar 

  6. Mariani G, Moresco L, Viale G, Villa G, Bagnasco M, Canavese G, et al. Radioguided sentinel lymph node biopsy in breast cancer surgery. J Nucl Med 2001;42:1198–215

    CAS  PubMed  Google Scholar 

  7. Keshtgar MRS, Ell PJ. Sentinel lymph node detection and imaging. Eur J Nucl Med 1999;26:57–67

    Article  CAS  PubMed  Google Scholar 

  8. Alazraki NP, Styblo T, Grant SF, Cohen C, Larsen T, Aarsvold JN. Sentinel node staging of early breast cancer using lymphoscintigraphy and the intraoperative gamma-detecting probe. Semin Nucl Med 2000;30:56–64

    CAS  PubMed  Google Scholar 

  9. Tiourina T, Arends B, Huysmans D, Rutten H, Lemaire B, Muller S. Evaluation of surgical gamma probes for radioguided sentinel node localization. Eur J Nucl Med 1998;25:1224–31

    Article  CAS  PubMed  Google Scholar 

  10. Britten AJ. A method to evaluate intra-operative gamma probes for sentinel lymph node localisation. Eur J Nucl Med 1999;26:76–83

    Article  CAS  PubMed  Google Scholar 

  11. Perkins AC, Britten AJ. Specification and performance of intra-operative gamma probes for sentinel node detection. Nucl Med Commun 1999;20:309–15

    CAS  PubMed  Google Scholar 

  12. Kitagawa Y, Fujii H, Mukai M, Kubota T, Otani Y, Kitajima M. Radio-guided sentinel node detection for gastric cancer. Br J Surg 2002;89:604–8

    Article  CAS  PubMed  Google Scholar 

  13. Cutini G, Gesuelli GC, Sartelli M, Brianzoni E, Musolino G, Nestori M, Scibe R, Berbellini A. The role of lymphoscintigraphy in rectal laparoscopic surgery: can sentinel node concept be applied to rectal carcinoma? Surg Endosc 2001;15:1440–43

    CAS  PubMed  Google Scholar 

  14. Merrie AE, van Rij AM, Phillips LV, Rossaak JI, Yun K, McCall JL. Diagnostic use of sentinel node in colon cancer. Dis Colon Rectum 2001;44:410–7

    CAS  PubMed  Google Scholar 

  15. Pelizzo MR, Boschin IM, Toniato A, Bernante P, Piotto A, Rinaldo A, Ferlito A. The sentinel node procedure with Patent Blue V dye in the surgical treatment of papillary thyroid carcinoma. Acta Otolaryngol 2001;121:421–4

    Article  CAS  PubMed  Google Scholar 

  16. Gitsch E, Philipp K, Pateisky N. Intraoperative lymph scintigraphy during radical surgery for cervical cancer. J Nucl Med 1984;25:486–9

    CAS  PubMed  Google Scholar 

  17. Lechner P, Lind P, Snyder M, Haushofer H. Probe-guided surgery for colorectal cancer. Recent Results Cancer Res 2000;157:273–80

    CAS  PubMed  Google Scholar 

Download references

Acknowledgements

This study was performed with the financial assistance of the MAGNET Program, Office of the Chief Scientist, Ministry of Industry and Trade, State of Israel.

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Moshe Hashmonai.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Kopelman, D., Blevis, I., Iosilevsky, G. et al. A newly developed intra-operative gamma camera: performance characteristics in a laboratory phantom study. Eur J Nucl Med Mol Imaging 32, 1217–1224 (2005). https://doi.org/10.1007/s00259-005-1823-y

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00259-005-1823-y

Keywords

Navigation