Respiratory gating of cardiac PET data in list-mode acquisition

  • Original Article
  • Published:
European Journal of Nuclear Medicine and Molecular Imaging Aims and scope Submit manuscript

Abstract

Purpose

Respiratory motion has been identified as a source of artefacts in most medical imaging modalities. This paper reports on respiratory gating as a means to eliminate motion-related inaccuracies in PET imaging.

Methods

Respiratory gating was implemented in list mode with physiological signal recorded every millisecond together with the PET data. Respiration was monitored with an inductive respiration monitor using an elasticised belt around the patient’s chest. Simultaneous ECG gating can be maintained independently by encoding ECG trigger signal into the list-mode data. Respiratory gating is performed in an off-line workstation with gating parameters defined retrospectively. The technique was applied on a preliminary set of patient data with C15O.

Results

Motion was visually observed in the cine displays of the sagittal and coronal views of the reconstructed respiratory gated images. Significant changes in the cranial–caudal position of the heart could be observed. The centroid of the cardiac blood pool showed an excursion of 4.5–16.5 mm (mean 8.5±4.8 mm) in the cranial–caudal direction, with more limited excursion of 1.1–7.0 mm (mean 2.5±2.2 mm) in the horizontal direction and 1.3–3.7 mm (mean 2.4±0.9 mm) in the vertical direction.

Conclusion

These preliminary data show that the extent of motion involved in respiration is comparable to myocardial wall thickness, and respiratory gating may be considered in order to reduce this effect in the reconstructed images.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Fig. 1
Fig. 2
Fig. 3
Fig. 4
Fig. 5
Fig. 6
Fig. 7

Similar content being viewed by others

References

  1. Korin HW, Ehman RL, Riederer SJ, Felmlee JP, Grimm RC. Respiratory kinematics of the upper abdominal organs: a quantitative study. Magn Reson Med 1992;23(1):172–8

    Article  PubMed  CAS  Google Scholar 

  2. Taylor AM, Jhooti P, Wiesmann F, Keegan J, Firmin DN, Pennell DJ. MR navigator-echo monitoring of temporal changes in diaphragm position: implications for MR coronary angiography. J Magn Reson Imaging 1997;7(4):629–36

    Article  PubMed  CAS  Google Scholar 

  3. Hoffman EJ, Phelps ME, Wisenberg, G, Schelbert HR, Kuhl DE. Electrocardiographic gating in positron emission computed tomography. J Comput Assist Tomogr 1979;3(6):733–9

    Article  PubMed  CAS  Google Scholar 

  4. Ter-Pogossian MM, Bergmann SR, Sobel BE. Influence of cardiac and respiratory motion on tomographic reconstructions of the heart: implications for quantitative nuclear cardiology. J Comput Assist Tomogr 1982;6(6):1148–55

    Article  PubMed  CAS  Google Scholar 

  5. Susskind H, Alderson PO, Dzebolo NN, Bennett GW, Richards P, Rosen JM, Brill AB. Effect of respiratory motion on pulmonary activity determinations by positron tomography in dogs. Invest Radiol 1985;20(9):950–5

    Article  PubMed  CAS  Google Scholar 

  6. Klein GJ, Reutter BW, Ho MH, Reed JH, Huesman RH. Real-time system for respiratory-cardiac gating in positron tomography. IEEE Trans Nucl Sci 1998;45(4):2139–43

    Article  Google Scholar 

  7. Cho K, Kumiata S, Okada S, Kumazaki T. Development of respiratory gated myocardial SPECT system. J Nucl Cardiol. 1999;6(1):20–28

    Article  PubMed  CAS  Google Scholar 

  8. Goerres GW, Kamel E, Heidelberg TN, Schwitter MR, Burger C, von Schulthess GK. PET-CT image co-registration in the thorax: influence of respiration. Eur J Nucl Med Mol Imaging 2002;29(3):351–60

    Article  PubMed  CAS  Google Scholar 

  9. Nehmeh SA, Erdi YE, Ling CC, Rosenzweig KE, Squire OD, Braban LE, et al. Effect of respiratory gating on reducing lung motion artifacts in PET imaging of lung cancer. Med Phys 2002;29(3):366–71

    Article  PubMed  CAS  Google Scholar 

  10. Boucher L, Rodrigue S, Lecomte R, Benard F. Respiratory gating for 3-dimensional PET of the thorax: feasibility and initial results. J Nucl Med 2004;45(2):214–9

    PubMed  Google Scholar 

  11. Spinks TJ, Jones T, Bloomfield PM, Bailey DL, Miller M, Hogg D, et al. Physical characteristics of the ECAT EXACT3D positron tomograph. Phys Med Biol 2000;45(9):2601–18

    Article  PubMed  CAS  Google Scholar 

  12. Xu M, Luk WK, Cutler PD, Digby WM. Local threshold for segmented attenuation correction of PET imaging of the thorax. IEEE Trans Nucl Sci 1994;41(4):1532–37

    Article  Google Scholar 

  13. Bailey DL, Miller MP, Spinks TJ, Bloomfield PM, Livieratos L, Young HE, et al. Experience with fully 3D PET and implications for future high-resolution 3D tomographs. Phys Med Biol 1998;43(4):777–786

    Article  PubMed  CAS  Google Scholar 

  14. Bloomfield PM, Spinks TJ, Reed J, Schnorr L, Westrip AM, Livieratos L, et al. The design and implementation of a motion correction scheme for neurological PET. Phys Med Biol 2003;48(8):959–78

    Article  PubMed  Google Scholar 

  15. Kinahan PE, Rogers JG. Analytic 3D image reconstruction using all detected events. IEEE Trans Nucl Sci 1989;36(1):964–8

    Article  CAS  Google Scholar 

  16. Rajappan K, Livieratos L, Camici PG, Pennell DJ. Measurement of ventricular volumes and function: a comparison of gated PET and cardiovascular magnetic resonance. J Nucl Med 2002;43(6):806–10

    PubMed  Google Scholar 

  17. Klein GJ, Reutter BW, Huesman RH. Non-rigid summing of gated PET via optical flow. IEEE Trans Nucl Sci 1997;44(4):1509–12

    Article  Google Scholar 

  18. Livieratos L, Stegger L, Bloomfield PM, Schafers K, Bailey DL, Camici PG. Rigid-body transformation of list-mode projection data for respiratory motion correction in cardiac PET. Phys Med Biol 2005;50:3313–22

    Article  PubMed  CAS  Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Corresponding author

Correspondence to Lefteris Livieratos.

Rights and permissions

Reprints and permissions

About this article

Cite this article

Livieratos, L., Rajappan, K., Stegger, L. et al. Respiratory gating of cardiac PET data in list-mode acquisition. Eur J Nucl Med Mol Imaging 33, 584–588 (2006). https://doi.org/10.1007/s00259-005-0031-0

Download citation

  • Received:

  • Accepted:

  • Published:

  • Issue Date:

  • DOI: https://doi.org/10.1007/s00259-005-0031-0

Keywords

Navigation