TABLE 1

Criteria for Quantitative Analysis of PET and CT Results

ParameterData
PET IQ phantom
 Uniformity<15%
 RCs1
 SOR<0.20
 SUV<10% bias
CT air/water phantom
 Air0 HU
 Water−1,000 HU
CT TEM phantom
 Lung−700 HU
 Soft tissue (adipose/muscle)>0 HU
 Bone (soft/cortical)>200 HU
  • *HU for TEM originally defined based on literature (27).