TABLE 4

Interobserver Agreement Scores

Proportional agreement
Reconstruction methodComplete agreementOverallPositive agreementNegative agreement
4 mm14/24 (58.3%)78.5% (69.8%–85.7%)83.6% (73.2%–91.1%)68.7% (51.9%–82.5%)
4 mm+PSF16/24 (66.7%)82.6% (74.4%–89.1%)86.8% (77.0%–93.5%)74.7% (58.4%–87.2%)
2 mm16/24 (66.7%)82.6% (74.4%–89.1%)88.3% (79.4%–94.3%)66.7% (47.1%–82.8%)
2 mm+PSF18/24 (75.0%)84.7% (76.8%–90.8%)89.8% (81.4%–95.3%)69.4% (49.5%–85.2%)
  • Data are for equal interpretation (either positive or negative) of scan by all 4 readers. Data in parentheses are 95% CIs.