TABLE 3

Quantification Errors for Multi-isotope Studies Using 1:1 Activity Ratios and MH Apertures

Isotope A (spill-out)
20% energy window10% energy window
Isotope B (spill-in)Peak99mTc111In123I177Lu201Tl99mTc111In123I177Lu201Tl
99mTc140.520.033.11.86.212.812.31.14.7
111In171.32.587.91.913.11.747.61.312.1
245.42.57.52.41.81.84.80.80.7
123I159.033.072.51.711.25.440.11.15.1
177Lu56.132.767.155.2105.829.359.147.882.5
112.969.7141.791.726.044.798.061.617.8
208.415.093.957.432.110.046.836.621.6
201Tl72.32.94.94.62.22.03.63.61.4
167.443.0749.4755.514.08.8735.2593.29.8
  • Peak is in keV; error is in %.