TABLE 3

Contrast, Background Variability, and Average Lung Residual for 4:1 Sphere-to-Background Ratio on Vision, Directly Compared with Published Results from mCT Flow

Contrast (%)Background variability (%)
Sphere size (mm)VisionmCT FlowVisionmCT Flow
1093.128.36.46.9
1373.547.95.06.8
1779.058.44.05.6
2287.070.83.14.7
2886.367.02.73.8
3789.476.92.22.9
Average lung residual (%)3.412.3
  • mCT Flow measurements include effect of low-resolution matrix and postreconstruction 3-mm gaussian filter (12).